Use of a compact pixellated gamma camera for small animal pinhole SPECT imaging

被引:27
|
作者
Zeniya, Tsutomu
Watabe, Hiroshi
Aoi, Toshiyuki
Kim, Kyeong Min
Teramoto, Noboru
Takeno, Takeshi
Ohta, Yoichiro
Hayashi, Takuya
Mashino, Hiroyuki
Ota, Toshihiro
Yamamoto, Seiichi
Iida, Hidehiro
机构
[1] Natl Cardiovasc Ctr, Res Inst, Adv Med Engn Ctr, Dept Invest Radiol, Osaka 5658565, Japan
[2] Korea Inst Radiol & Med Sci, Radiol & Med Sci Res Ctr, Nucl Med Lab, Seoul, South Korea
[3] Kobe City Coll Technol, Dept Elect Engn, Kobe, Hyogo, Japan
关键词
SPECT; pinhole collimator; compact pixellated gamma camera; small animal;
D O I
10.1007/BF03027376
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: Pinhole SPECT which permits in vivo high resolution 3D imaging of physiological functions in small animals facilitates objective assessment of pharmaceutical development and regenerative therapy in pre-clinical trials. For handiness and mobility, the miniature size of the SPECT system is useful. We developed a small animal SPECT system based on a compact high-resolution gamma camera fitted to a pinhole collimator and an object-rotating unit. This study was aimed at evaluating the basic performance of the detection system and the feasibility of small animal SPECT imaging. Methods: The gamma camera consists of a 22 x 22 pixellated scintillator array of 1.8 mm x 1.8 mm x 5 mm NaI(Tl) crystals with 0.2-mm gap between the crystals coupled to a 2" flat panel position-sensitive photomultiplier tube (Hamamatsu H8500) with 64 channels. The active imaging region of the camera was 43.8 mm x 43.8 mm. Data acquisition is controlled by a personal computer (Microsoft Windows) through the camera controller. Projection data over 360 degrees for SPECT images are obtained by synchronizing with the rotating unit. The knife-edge pinhole collimators made of tungsten are attached on the camera and have 0.5-mm and 1.0-mm apertures. The basic performance of the detection system was evaluated with Tc-99m and Tl-201 solutions. Energy resolution, system spatial resolution and linearity of count rate were measured. Rat myocardial perfusion SPECT scans were sequentially performed following intravenous injection of 201 TlCl. Projection data were reconstructed using a previously validated pinhole 3D-OSEM method. Results: The energy resolution at 140 keV was 14.8% using a point source. The system spatial resolutions were 2.8-mm FWHM and 2.5-mm FWHM for Tc-99m and Tl-201 line sources, respectively, at 30-mm source distance (magnification factor of 1.3) using a 1.0-mm pinhole. The linearity between the activity and count rate was good up to 10 kcps. In a rat study, the left ventricular walls were clearly visible in all scans. Conclusions: We developed a compact SPECT system using compact gamma camera for small animals and evaluated basic physical performances. The present system may be of use for quantitation of biological functions such as myocardial blood flow in small animals.
引用
收藏
页码:409 / 416
页数:8
相关论文
共 50 条
  • [21] Multi-pinhole small animal SPECT
    Schramm, N
    Schurrat, T
    Béhé, M
    Lackas, C
    Hoppin, J
    Ebel, G
    Behr, TM
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 : S236 - S236
  • [22] Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography
    David, S.
    Georgiou, M.
    Fysikopoulos, E.
    Loudos, G.
    ENGINEERING TECHNOLOGY & APPLIED SCIENCE RESEARCH, 2011, 1 (01) : 17 - 22
  • [23] A geometric system model of finite aperture in small animal pinhole SPECT imaging
    Hsu, Ching-Han
    Huang, Po-Chia
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2006, 30 (03) : 181 - 185
  • [24] PINHOLE SPECT - ULTRA-HIGH-RESOLUTION IMAGING FOR SMALL ANIMAL STUDIES
    WEBER, DA
    IVANOVIC, M
    JOURNAL OF NUCLEAR MEDICINE, 1995, 36 (12) : 2287 - 2289
  • [25] Improving the quality of small animal brain pinhole SPECT imaging by Bayesian reconstruction
    Sohlberg, A
    Lensu, S
    Jolkkonen, J
    Tuomisto, L
    Ruotsalainen, U
    Kuikka, JT
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 (07) : 986 - 994
  • [26] Improving the quality of small animal brain pinhole SPECT imaging by Bayesian reconstruction
    Antti Sohlberg
    Sanna Lensu
    Jukka Jolkkonen
    Leena Tuomisto
    Ulla Ruotsalainen
    Jyrki T. Kuikka
    European Journal of Nuclear Medicine and Molecular Imaging, 2004, 31 : 986 - 994
  • [27] Phantom Experiments on a PSAPD-Based Compact Gamma Camera With Submillimeter Spatial Resolution for Small Animal SPECT
    Kim, Sangtaek
    McClish, Mickel
    Alhassen, Fares
    Seo, Youngho
    Shah, Kanai S.
    Gould, Robert G.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (05) : 2518 - 2523
  • [28] Compact CT/SPECT small-animal imaging system
    Kastis, GA
    Furenlid, LR
    Wilson, DW
    Peterson, TE
    Barber, HB
    Barrett, HH
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (01) : 63 - 67
  • [29] Compact CT/SPECT small-animal imaging system
    Kastis, GA
    Furenlid, LR
    Wilson, DW
    Peterson, TE
    Barber, HB
    Barrett, HH
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 797 - 801
  • [30] Two-level multi-pinhole collimator for SPECT imaging using a small-field-of-view gamma camera
    Jaekeon Bae
    Seungbin Bae
    Soo-young Lee
    Kisung Lee
    Yongkwon Kim
    Jinhun Joung
    MinHo Kim
    Kyeong Min Kim
    Journal of the Korean Physical Society, 2017, 70 : 192 - 200