Validation of the angular measurements of a new inertial-measurement-unit based rehabilitation system: comparison with state-of-the-art gait analysis

被引:67
|
作者
Leardini, Alberto [1 ]
Lullini, Giada [1 ]
Giannini, Sandro [1 ]
Berti, Lisa [1 ]
Ortolani, Maurizio [1 ]
Caravaggi, Paolo [1 ]
机构
[1] Ist Ortoped Rizzoli, Movement Anal Lab, Bologna, Italy
关键词
Inertial measurement unit; Gait analysis; Rehabilitation; Knee; Hip; Thorax; Joint flexion; Audio-visual bio-feedback; Video-games; JOINT ANGLE MEASUREMENT; KINEMATICS; MOTION; SENSORS; BIOFEEDBACK; OUTCOMES; GAMES;
D O I
10.1186/1743-0003-11-136
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Several rehabilitation systems based on inertial measurement units (IMU) are entering the market for the control of exercises and to measure performance progression, particularly for recovery after lower limb orthopaedic treatments. IMU are easy to wear also by the patient alone, but the extent to which IMU's malpositioning in routine use can affect the accuracy of the measurements is not known. A new such system (Riablo (TM), CoRehab, Trento, Italy), using audio-visual biofeedback based on videogames, was assessed against state-of-the-art gait analysis as the gold standard. Methods: The sensitivity of the system to errors in the IMU's position and orientation was measured in 5 healthy subjects performing two hip joint motion exercises. Root mean square deviation was used to assess differences in the system's kinematic output between the erroneous and correct IMU position and orientation. In order to estimate the system's accuracy, thorax and knee joint motion of 17 healthy subjects were tracked during the execution of standard rehabilitation tasks and compared with the corresponding measurements obtained with an established gait protocol using stereophotogrammetry. Results: A maximum mean error of 3.1 +/- 1.8 deg and 1.9 +/- 0.8 deg from the angle trajectory with correct IMU position was recorded respectively in the medio-lateral malposition and frontal-plane misalignment tests. Across the standard rehabilitation tasks, the mean distance between the IMU and gait analysis systems was on average smaller than 5 degrees. Conclusions: These findings showed that the tested IMU based system has the necessary accuracy to be safely utilized in rehabilitation programs after orthopaedic treatments of the lower limb.
引用
收藏
页数:7
相关论文
共 49 条
  • [1] Validation of the angular measurements of a new inertial-measurement-unit based rehabilitation system: comparison with state-of-the-art gait analysis
    Alberto Leardini
    Giada Lullini
    Sandro Giannini
    Lisa Berti
    Maurizio Ortolani
    Paolo Caravaggi
    Journal of NeuroEngineering and Rehabilitation, 11
  • [2] Validity and repeatability of a new inertial measurement unit system for gait analysis on kinematic parameters: Comparison with an optoelectronic system
    Piche, Elodie
    Guilbot, Marine
    Chorin, Frederic
    Guerin, Olivier
    Zory, Raphael
    Gerus, Pauline
    MEASUREMENT, 2022, 198
  • [3] Inertial Measurement Units: A Brief State of the Art on Gait Analysis
    Ribeiro, Nuno Ferrete
    Santos, Cristina P.
    2017 IEEE 5TH PORTUGUESE MEETING ON BIOENGINEERING (ENBENG), 2017,
  • [4] Gait-force model and inertial measurement unit-based measurements: A new approach for gait analysis and balance monitoring
    Li, Xinan
    Xu, Hongyuan
    Cheung, Jeffrey T.
    JOURNAL OF EXERCISE SCIENCE & FITNESS, 2016, 14 (02) : 60 - 66
  • [5] Validation of a Rowing Oar Angle Measurement System based on an Inertial Measurement Unit
    Gravenhorst, Franz
    Troester, Gerhard
    Turner, Timothy
    Smith, Richard M.
    Draper, Conny
    2013 12TH IEEE INTERNATIONAL CONFERENCE ON TRUST, SECURITY AND PRIVACY IN COMPUTING AND COMMUNICATIONS (TRUSTCOM 2013), 2013, : 1412 - 1419
  • [6] Analysis of Human Gait Based on Smartphone Inertial Measurement Unit: A Feasibility Study
    Meigal, Alexander
    Reginya, Sergey
    Gerasimova-Meigal, Liudmila
    Prokhorov, Kirill
    Moschevikin, Alex
    PROCEEDINGS OF THE 2018 22ND CONFERENCE OF OPEN INNOVATIONS ASSOCIATION (FRUCT), 2018, : 151 - 158
  • [7] Surface roughness measurements of turned parts through a vision-based measurement system: uncertainty analysis and performance comparison with state-of-the-art instruments
    Baleani, Alessia
    Paone, Nicola
    Gladines, Jona
    Vanlanduit, Steve
    PROCEEDINGS OF 2022 IEEE INTERNATIONAL WORKSHOP ON METROLOGY FOR INDUSTRY 4.0 & IOT (IEEE METROIND4.0&IOT), 2022, : 17 - 22
  • [8] Validation of a portable system for spatial-temporal gait parameters based on a single inertial measurement unit and a mobile application
    Aqueveque, Pablo
    Gomez, Britam
    Saavedra, Francisco
    Canales, Cristian
    Contreras, Simon
    Ortega-Bastidas, Paulina
    Cano-de-la-Cuerda, Roberto
    EUROPEAN JOURNAL OF TRANSLATIONAL MYOLOGY, 2020, 30 (02) : 268 - U136
  • [9] Comparison of Camera based and Inertial Measurement Unit based Motion Analysis
    Jang, Seongho
    Park, Si-bog
    Moon, Sang-bog
    Kim, Jae Min
    Lee, Shi-uk
    PROCEEDINGS OF THE 7TH INTERNATIONAL CONFERENCE ON SENSOR NETWORKS (SENSORNETS), 2018, : 161 - 167
  • [10] Quantitative analysis of the bilateral coordination and gait asymmetry using inertial measurement unit-based gait analysis
    Han, Seung Hwan
    Kim, Chang Oh
    Kim, Kwang Joon
    Jeon, Jeanhong
    Chang, Hsienhao
    Kim, Eun Seo
    Park, Hoon
    PLOS ONE, 2019, 14 (10):