Induced magnetic moment in stainless steel components of orthodontic appliances in 1.5 T MRI scanners

被引:8
|
作者
Wang, Zhiyue J. [1 ,2 ]
Rollins, Nancy K. [1 ,2 ]
Liang, Hui [3 ]
Park, Yong Jong [4 ]
机构
[1] Childrens Med Ctr, Dept Radiol, Dallas, TX 75235 USA
[2] Univ Texas SW Med Ctr Dallas, Dept Radiol, Dallas, TX 75390 USA
[3] Texas A&M Univ, Baylor Coll Dent, Dallas, TX 75246 USA
[4] Childrens Med Ctr, Div Orthodont, Dallas, TX 75235 USA
关键词
MRI artifacts; orthodontic appliances; stainless steel; induced magnetic moment; FIELD INTERACTIONS; IMAGING ARTIFACTS; DENTAL MATERIALS; STROKE; IMAGES;
D O I
10.1118/1.4930796
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Most orthodontic appliances are made of stainless steel materials and induce severe magnetic susceptibility artifacts in brain MRI. In an effort for correcting these artifacts, it is important to know the value of induced magnetic moments in all parts of orthodontic appliances. In this study, the induced magnetic moment of stainless steel orthodontic brackets, molar bands, and arch-wires from several vendors is measured. Methods: Individual stainless steel brackets, molar bands, and short segments of arch-wire were positioned in the center of spherical flask filled with water through a thin plastic rod. The induced magnetic moment at 1.5 T was determined by fitting the B-0 map to the z-component of the magnetic dipole field using a computer routine. Results: The induced magnetic moment at 1.5 T was dominated by the longitudinal component m(z), with a small contribution from the transverse components. The m(z) was insensitive to the orientation of the metal parts. The orthodontic brackets collectively dominated the magnetic dipole moment in orthodontic appliances. In brackets from six vendors, the total induced m(z) from 20 brackets for nonmolar teeth ranged from 0.108 to 0.158 (median 0.122) A.m(2). The m(z) in eight molar bands with bracket attachment from two vendors ranged from 0.0004 to 0.0166 (median 0.0035) A.m(2). Several full length arch wires had induced magnetic moment in the range of 0.006-0.025 (median 0.015) A.m(2). Conclusions: Orthodontic brackets collectively contributed most to the total magnetic moment. Different types of brackets, molar bands, and arch wires all exhibit substantial variability in the induced magnetic moment. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:5871 / 5878
页数:8
相关论文
共 50 条
  • [31] Quantification of 3D Geometric Distortion for 1.5T and 3T MRI Scanners Used for Radiation Therapy
    Stowe, M.
    Gupta, N.
    Raterman, B.
    Lu, L.
    MEDICAL PHYSICS, 2016, 43 (06) : 3662 - 3662
  • [32] Measurement of the magnetic moment in a cold worked 304 stainless steel using HTS SQUID
    Park, D. G.
    Kim, D. W.
    Angani, C. S.
    Timofeev, V. P.
    Cheong, Y. M.
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2008, 320 (20) : E571 - E574
  • [33] Measurement of the weighted peak level for occupational exposure to gradient magnetic fields for 1.5 and 3 Tesla MRI body scanners
    Bonutti, F.
    Tecchio, M.
    Maieron, M.
    Trevisan, D.
    Negro, C.
    Calligaris, F.
    RADIATION PROTECTION DOSIMETRY, 2016, 168 (03) : 358 - 364
  • [34] Cytotoxicity of Metal Ions Released from NiTi and Stainless Steel Orthodontic Appliances, Part 1: Surface Morphology and Ion Release Variations
    Didovic, Mirna Petkovic
    Badovinac, Ivana Jelovica
    Fiket, Zeljka
    Zigon, Jure
    Mlinaric, Marijana Rincic
    Juresic, Gordana Canadi
    MATERIALS, 2023, 16 (11)
  • [35] Control of the RF magnetic field pattern by tunable metasurface in 1.5 T MRI
    Kretov, E.
    Shchelokova, A.
    Slobozhanyuk, A.
    METANANO 2019, 2020, 1461
  • [36] New proposal for the bending moment capacity of tubular T-joints in carbon steel and stainless steel structures
    Segura, G.
    Arrayago, I.
    Mirambell, E.
    THIN-WALLED STRUCTURES, 2023, 187
  • [37] Surface changes (scanning electron microscope) induced by artificial saliva in stainless steel orthodontic loops
    Castro, Saul
    Martins, Eugenio
    Ponces, Maria
    Vasconcelos, Mario
    Pollmann, Maria
    BIODENTAL ENGINEERING IV, 2018, : 73 - 74
  • [38] Cardiac T2 star mapping: standardized inline analysis of long and short axis at three identical 1.5 T MRI scanners
    Rafael Heiss
    Marco Wiesmueller
    Christoph Treutlein
    Hannes Seuss
    Michael Uder
    Matthias May
    Wolfgang Wuest
    The International Journal of Cardiovascular Imaging, 2019, 35 : 695 - 702
  • [39] Quantitative MRI using STrategically Acquired Gradient Echo (STAGE): optimization for 1.5 T scanners and T1 relaxation map validation
    Alice Pirastru
    Yongsheng Chen
    Laura Pelizzari
    Francesca Baglio
    Mario Clerici
    E. Mark Haacke
    Maria Marcella Laganà
    European Radiology, 2021, 31 : 4504 - 4513
  • [40] Assessment the Influencing Factors on MRS Signal Obtained from High Magnetic Field Strength (1.5T) MRI Scanners During the Application of Homemade Metabolite Phantom
    Saeedi-Moghadam, Mahdi
    Pouladian, Majid
    Lotfi, Mehrzad
    Faghihi, Reza
    JOURNAL OF MAGNETICS, 2019, 24 (01) : 90 - 98