Optimization of dual-energy imaging systems using generalized NEQ and imaging task

被引:104
|
作者
Richard, S.
Siewerdsen, J. H.
机构
[1] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
[2] Princess Margaret Hosp, Ontario Canc Inst, Toronto, ON M5G 2M9, Canada
[3] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
[4] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
关键词
DQE; MTF; NPS; dual-energy imaging; flat-panel detector; imaging performance; image noise; anatomical noise; cascaded systems analysis; optimization;
D O I
10.1118/1.2400620
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dual-energy (DE) imaging is a promising advanced application of flat-panel detectors (FPDs) with a potential host of applications ranging from thoracic and cardiac imaging to interventional procedures. The performance of FPD-based DE imaging systems is investigated in this work by incorporating the noise-power spectrum associated with overlying anatomical structures ("anatomical noise" modeled according to a I If characteristic) into descriptions of noise-equivalent quanta (NEQ) to yield the generalized NEQ (GNEQ). Signal and noise propagation in the DE imaging chain is modeled by cascaded systems analysis. A Fourier-based description of the imaging task is integrated with the GNEQ to yield a detectability index used as an objective function for optimizing DE image reconstruction, allocation of dose between low- and high-energy images, and selection of low- and hiah-kVp. Optimal reconstruction and acquisition parameters were found to depend on dose; for example, optimal kVp varied from [60/150] kVp at typical radiographic dose levels (similar to 0.5 mGy entrance surface dose, ESD) but increased to [90/150] kVp at high dose (ESD similar to 5.0 mGy). At very low dose (ESD similar to 0.05 mGy), detectability index indicates an optimal low-energy technique of 60 kVp but was largely insensitive to the choice of high-kVp in the range 120-150 kVp. Similarly, optimal dose allocation, defined as the ratio of low-energy ESD and the total ESD, varied from 0.2 to 0.4 over the range ESD=(0.05-5.0) mGy. Furthermore, two applications of the theoretical framework were explored: (i) the increase in detectability for DE imaging compared to conventional radiography; and (ii) the performance of single-shot vs double-shot DE imaging. wherein the latter is found to have a DQE approximately twice that of the former. Experimental and theoretical analysis of GNEQ,and task-based detectability index provides a fundamental understanding of the factors governing DE imaging performance and offers a framework for system design and optimization. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:127 / 139
页数:13
相关论文
共 50 条
  • [31] AN INVERSE MATRIX APPROACH TO DUAL-ENERGY IMAGING
    SHAW, CG
    PLEWES, DP
    MEDICAL PHYSICS, 1986, 13 (04) : 577 - 577
  • [32] Dual-Energy MDCT for Imaging the Renal Mass
    Mileto, Achille
    Nelson, Rendon C.
    Paulson, Erik K.
    Marin, Daniele
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (06) : W640 - W647
  • [33] Dual-Energy CT: Applications in Abdominal Imaging
    Bauer R.W.
    Fischer S.
    Current Radiology Reports, 3 (4)
  • [34] Numerical Simulation for Dual-energy CT Imaging
    Wang, Lixin
    Sun, Fengrong
    Babyn, Paul
    Zhong, Hai
    Zhou, Zhenzhen
    Huang, Hao
    Qin, Jiaojiao
    PROCEEDINGS OF THE 2015 2ND INTERNATIONAL CONFERENCE ON ELECTRICAL, COMPUTER ENGINEERING AND ELECTRONICS (ICECEE 2015), 2015, 24 : 316 - 320
  • [35] Dual-Energy Computed Tomography in Genitourinary Imaging
    Mileto, Achille
    Marin, Daniele
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2017, 55 (02) : 373 - +
  • [36] Dual-Energy Computed Tomography in Cardiac Imaging
    Boettcher, Benjamin
    Zsarnoczay, Emese
    Varga-Szemes, Akos
    Schoepf, Uwe Joseph
    Meinel, Felix G.
    van Assen, Marly
    De Cecco, Carlo N.
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 995 - 1009
  • [37] Dual-energy CT imaging of thoracic malignancies
    Zhang, Long Jiang
    Yang, Gui Fen
    Wu, Sheng Yong
    Xu, Jian
    Lu, Guang Ming
    Schoepf, U. Joseph
    CANCER IMAGING, 2013, 13 (01): : 81 - 91
  • [38] Dual-energy Computed Tomography Imaging of the Aorta
    Vlahos, Ioannis
    Godoy, Myrna C. B.
    Naidich, David P.
    JOURNAL OF THORACIC IMAGING, 2010, 25 (04) : 289 - 300
  • [39] Dual-Energy Computed Tomography in Body Imaging
    Toia, Giuseppe V.
    Kim, Sooah
    Dighe, Manjiri K.
    Mileto, Achille
    SEMINARS IN ROENTGENOLOGY, 2018, 53 (02) : 132 - 146
  • [40] THEORETICAL OPTIMIZATION OF DUAL-ENERGY X-RAY-IMAGING WITH APPLICATION TO MAMMOGRAPHY
    JOHNS, PC
    YAFFE, MJ
    MEDICAL PHYSICS, 1985, 12 (03) : 289 - 296