Transcatheter Aortic Valve Implantation in Patients of Small Body Size

被引:21
|
作者
Watanabe, Yusuke [1 ]
Hayashida, Kentaro [1 ]
Lefevre, Thierry [1 ]
Chevalier, Bernard [1 ]
Hovasse, Thomas [1 ]
Romano, Mauro [1 ]
Garot, Philippe [1 ]
Farge, Arnaud [1 ]
Donzeau-Gouge, Patrick [1 ]
Bouvier, Erik [1 ]
Cormier, Bertrand [1 ]
Morice, Marie-Claude [1 ]
机构
[1] Inst Cardiovasc Paris Sud, F-91300 Massy, France
关键词
transcatheter aortic valve implantation; small body size; vascular complication; aortic valve; PROSTHESIS; STENOSIS; RUPTURE; ECHOCARDIOGRAPHY; RECOMMENDATIONS; OUTCOMES; IMPACT;
D O I
10.1002/ccd.24970
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims: The aim of this study was to evaluate the efficacy and safety of transcatheter aortic valve implantation (TAVI) in patients of small body size (SB). Methods and Results: Between October 2006 and November 2011, 424 patients undergoing TAVI were evaluated. Patients with a small body surface area (BSA; <1.75 m(2)) were defined as the "SB group" (n = 215) and the remaining large BSA patients as the "LB group" (n = 209). In the SB group, mean annulus diameter measured by CT and minimal femoral artery diameter were significantly smaller (23.1 +/- 1.92 vs. 24.3 +/- 1.91 mm, P < 0.01; 7.59 +/- 1.06 vs. 8.29 +/- 1.34 mm, P < 0.01, respectively). The logistic EuroSCORE was similar between groups (22.5 +/- 10.4 vs. 23.4 +/- 12.0; P = 0.43). Equivalent device success (91.2% vs. 91.4%; P = 0.94) and 30-day survival rates (85.1% vs. 87.6%; P = 0.46) were achieved in both groups. Vascular complications and the 30-day combined safety end point rate were significantly higher in the SB group (13.0% vs. 4.3%, P < 0.01; 27.4% vs. 18.3%, P = 0.03, respectively). Similar post-procedural mean pressure gradient was achieved in both groups (10.2 +/- 4.3 vs. 11.0 +/- 4.6 mmHg; P = 0.12). Multivariate analysis showed that BSA was one of the independent predictors of 30-day combined safety end point (Odds ratio: 0.20; 95% confidence interval (CI): 0.05-0.88; P = 0.03). However, Kaplan-Meier analysis revealed no significant difference in mid-term survival rate between groups (log-rank P = 0.64). Conclusions: TAVI in SB patients resulted in similar clinical outcomes with effective post-procedural hemodynamics compared with LB patients. Care should be taken to avoid vascular complications in SB patients. (C) 2013 Wiley Periodicals, Inc.
引用
收藏
页码:272 / 280
页数:9
相关论文
共 50 条
  • [41] Transcatheter Aortic Valve Implantation
    Malaisrie, S. Chris
    Iddriss, Adam
    Flaherty, James D.
    Churyla, Andrei
    CURRENT ATHEROSCLEROSIS REPORTS, 2016, 18 (05)
  • [42] Transcatheter Aortic Valve Implantation
    Ye, Jian
    Lichtenstein, Samuel V.
    INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY, 2009, 4 (04) : 197 - 205
  • [43] Transcatheter Aortic Valve Implantation
    Krane, Markus
    CURRENT CARDIOLOGY REVIEWS, 2013, 9 (04) : 267 - 267
  • [44] Transcatheter Aortic Valve Implantation
    Nielsen, Hans Henrik Moller
    DANISH MEDICAL JOURNAL, 2012, 59 (12):
  • [45] Transcatheter Aortic Valve Implantation
    S. Chris Malaisrie
    Adam Iddriss
    James D. Flaherty
    Andrei Churyla
    Current Atherosclerosis Reports, 2016, 18
  • [46] Transcatheter Aortic Valve Implantation
    Moellmann, Helge
    Kempfert, Joerg
    Hamm, Christian W.
    Walther, Thomas
    HERZ, 2010, 35 (02) : 62 - 68
  • [47] Transcatheter Aortic Valve Implantation for Patients With Severe Bicuspid Aortic Valve Stenosis
    Hayashida, Kentaro
    Bouvier, Erik
    Lefevre, Thierry
    Chevalier, Bernard
    Hovasse, Thomas
    Romano, Mauro
    Garot, Philippe
    Watanabe, Yusuke
    Farge, Arnaud
    Donzeau-Gouge, Patrick
    Cormier, Bertrand
    Morice, Marie-Claude
    CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2013, 6 (03) : 284 - 291
  • [48] Transcatheter aortic valve implantation in patients with bicuspid aortic valve: a series of cases
    Kosek, Mikolaj
    Witkowski, Adam
    Dabrowski, Maciej
    Jastrzebski, Jan
    Michalowska, Ilona
    Chmielak, Zbigniew
    Demkow, Marcin
    Ksiezycka-Majczynska, Ewa
    Michalek, Piotr
    Szymanski, Piotr
    Hryniewiecki, Tomasz
    Stepinska, Janina
    Sitkowska-Rysiak, Ewa
    Ruzyllo, Witold
    KARDIOLOGIA POLSKA, 2015, 73 (08) : 627 - 636
  • [49] TRANSCATHETER AORTIC VALVE IMPLANTATION FOR PATIENTS WITH BICUSPID SEVERE AORTIC VALVE STENOSIS
    Hayashida, Kentaro
    Lefevre, Thierry
    Chevalier, Bernard
    Hovasse, Thomas
    Romano, Mauro
    Garot, Philippe
    Farge, Arnaud
    Donzeau-Gouge, Patrick
    Bouvier, Erik
    Cormier, Bertrand
    Morice, Marie-Claude
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 59 (13) : E4 - E4
  • [50] Transcatheter aortic valve implantation
    Yoshiki Sawa
    Surgery Today, 2015, 45 : 527 - 536