Voltammetric determination of catechol based on a glassy carbon electrode modified with a composite consisting of graphene oxide and polymelamine

被引:30
|
作者
Palanisamy, Selvakumar [1 ]
Ramaraj, Sayee Kannan [2 ]
Chen, Shen-Ming [1 ]
Velusamy, Vijayalakshmi [3 ]
Yang, Thomas C. K. [4 ]
Chen, Tse-Wei [1 ]
机构
[1] Natl Taipei Univ Technol, Electroanal & Bioelectrochem Lab, Dept Chem Engn & Biotechnol, 1,Sect 3,Chung Hsiao East Rd, Taipei 106, Taiwan
[2] Thiagarajar Coll, PG & Res Dept Chem, Madurai 09, Tamil Nadu, India
[3] Manchester Metropolitan Univ, Sch Engn, Div Elect & Elect Engn, Manchester M1 5GD, Lancs, England
[4] Natl Taipei Univ Technol, Dept Chem Engn & Biotechnol, 1,Sect 3,Chung Hsiao East Rd, Taipei 106, Taiwan
关键词
Electrochemical sensor; Differential pulse voltammetry; Conductive polymer: nanocomposite; FESEM; FTIR; Water analysis; SIMULTANEOUS ELECTROCHEMICAL DETERMINATION; DIHYDROXYBENZENE ISOMERS; HYDROQUINONE; POLY(MELAMINE); NANOCOMPOSITE; RESORCINOL;
D O I
10.1007/s00604-017-2073-y
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The authors describe an voltammetric catechol (CC) assay based on the use of a glassy carbon electrode (GCE) modified with a composite consisting of graphene oxide and polymelamine (GO/PM). The modified GCE was characterized by field emission scanning electron microscopy, elemental analysis, Raman spectroscopy and FTIR. Cyclic voltammetry reveals a well-defined response to CC, with an oxidation peak current that is distinctly enhanced compared to electrodes modified with GO or PM only. The combined synergetic activity of GO and PM in the composite also results in a lower oxidation potential. Differential pulse voltammetry (DPV) shows a response that is linear in the 0.03 to 138 mu M CC concentration range. The detection limit is 8 nM, and the sensitivity is 0.537 mu Aai...mu M(-1)ai...cm(-2). The sensor is selective for CC even in the presence of potentially interfering compounds including hydroquinone, resorcinol and dopamine. The modified GCE is highly reproducible, stable, sensitive, and shows an excellent practicability for detection of CC in water samples.
引用
收藏
页码:1051 / 1057
页数:7
相关论文
共 50 条
  • [41] Electrochemical Behavior and Voltammetric Determination of Theophylline at a Glassy Carbon Electrode Modified with Graphene/Nafion
    Yamin Li
    Shufang Wu
    Peili Luo
    Jie Liu
    Ge Song
    Kai Zhang
    Baoxian Ye
    Analytical Sciences, 2012, 28 : 497 - 502
  • [42] Voltammetric Determination of Cefuroxime Axetil in Pharmaceuticals Using Graphene Oxide Modified Glassy Carbon Electrode and Hanging Mercury Drop Electrode
    Kablan, Sevilay Erdogan
    Ozaltin, Nuran
    RUSSIAN JOURNAL OF ELECTROCHEMISTRY, 2019, 55 (04) : 305 - 313
  • [43] Voltammetric Determination of Cefuroxime Axetil in Pharmaceuticals Using Graphene Oxide Modified Glassy Carbon Electrode and Hanging Mercury Drop Electrode
    Sevilay Erdoğan Kablan
    Nuran Özaltın
    Russian Journal of Electrochemistry, 2019, 55 : 305 - 313
  • [44] Impedimetric Salmonella aptasensor using a glassy carbon electrode modified with an electrodeposited composite consisting of reduced graphene oxide and carbon nanotubes
    Fei Jia
    Nuo Duan
    Shijia Wu
    Ruitong Dai
    Zhouping Wang
    Xingmin Li
    Microchimica Acta, 2016, 183 : 337 - 344
  • [45] Impedimetric Salmonella aptasensor using a glassy carbon electrode modified with an electrodeposited composite consisting of reduced graphene oxide and carbon nanotubes
    Jia, Fei
    Duan, Nuo
    Wu, Shijia
    Dai, Ruitong
    Wang, Zhouping
    Li, Xingmin
    MICROCHIMICA ACTA, 2016, 183 (01) : 337 - 344
  • [46] A glassy carbon electrode modified with a composite consisting of gold nanoparticle, reduced graphene oxide and poly(L-arginine) for simultaneous voltammetric determination of dopamine, serotonin and L-tryptophan
    Md. Zaved H. Khan
    Xiaoqiang Liu
    Yunfei Tang
    Jinhua Zhu
    Weiping Hu
    Xiuhua Liu
    Microchimica Acta, 2018, 185
  • [47] A glassy carbon electrode modified with a composite consisting of gold nanoparticle, reduced graphene oxide and poly(L-arginine) for simultaneous voltammetric determination of dopamine, serotonin and L-tryptophan
    Khan, Md. Zaved H.
    Liu, Xiaoqiang
    Tang, Yunfei
    Zhu, Jinhua
    Hu, Weiping
    Liu, Xiuhua
    MICROCHIMICA ACTA, 2018, 185 (09)
  • [48] A graphene oxide-mesoporous MnO2 nanocomposite modified glassy carbon electrode as a novel and efficient voltammetric sensor for simultaneous determination of hydroquinone and catechol
    Gan, Tian
    Sun, Junyong
    Huang, Kejing
    Song, Li
    Li, Youmei
    SENSORS AND ACTUATORS B-CHEMICAL, 2013, 177 : 412 - 418
  • [49] A voltammetric sensor based on graphene-modified electrode for simultaneous determination of catechol and hydroquinone
    Du, Haijun
    Ye, Jianshan
    Zhang, Jiaqi
    Huang, Xiaodan
    Yu, Chengzhong
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2011, 650 (02) : 209 - 213
  • [50] Voltammetric determination of nonylphenol using a glassy carbon electrode modified with a nanocomposite consisting of CTAB, Fe3O4 nanoparticles and reduced graphene oxide
    Zou, Jing
    Guo, Manli
    Feng, Yanlong
    Yang, Mei
    Cao, Yujuan
    Zhu, Debin
    Yu, Ying
    MICROCHIMICA ACTA, 2017, 184 (02) : 533 - 540