Feasibility of myocardial perfusion imaging studies in morbidly obese patients with a cadmium-zinc-telluride cardiac camera

被引:7
|
作者
Budzynska, Anna [1 ]
Osiecki, Sebastian [1 ]
Mazurek, Andrzej [1 ]
Piszczek, Stanistaw [1 ]
Dziuk, Mirostaw [1 ]
机构
[1] Mil Inst Med, Szaserow 128 St, PL-04141 Warsaw, Poland
关键词
obesity; BMI; myocardial perfusion imaging; CZT; cardiac camera; image artifacts; SPECT;
D O I
10.5603/NMR.2019.0003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BACKGROUND: A novel cardiac SPECT camera with cadmium-zinc-telluride (CZT) based technology has a fixed array of semiconductor detectors paired with pinhole collimators focused on the heart. Image acquisition in obese patients can be challenging because of much smaller detector field of view compared to conventional gamma cameras. The aim of this study was to evaluate the impact on high body mass on the feasibility of CZT myocardial perfusion imaging (MPI). The additional aim was to investigate the mechanism of the banana-shaped/obesity artifact, as referred to in literature, and to attempt at simulating it on a phantom study. MATERIAL AND METHODS: Study group consisted of 43 patients with morbid obesity (BMI >= 40 kg/m(2)). All these patients underwent myocardial perfusion imaging on both CZT cardiac camera and general purpose SPECT/CT gamma camera. Control group consisted of all patients who underwent myocardial perfusion imaging on CZT camera throughout one calendar year and whose BMI was lower than 40 kg/m(2). In this group, all repeated studies were re-analyzed for estimating the frequency of heart mispositioning in the camera field of view. The number of studies performed was 1180. A static cardiac phantom was used to simulate a banana-shaped artifact. A series of phantom acquisitions during which the phantom position was altered in the camera field of view was performed. RESULTS: In control group, 3.7% of all cardiac scintigrams required repetition, 18.9% of which were repeated due to wrong heart positioning; median BMI in this group of patients was 36.0. A banana-shaped artifact was observed in one female patient with BMI 36.0. In morbid obesity group, 32.6% of the studies were non-diagnostic with "truncation effect" on Scan Quality Control (QC). Median BMI in patients with diagnostic scans was 42.0, while in patients with not acceptable quality control test it was 45.0 (p < 0.05). Banana-shaped artifacts were observed in 5 non-diagnostic studies. In a phantom study an artifact of banana shape was obtained when gantry was distant from the phantom and target was on the edge of the camera field of view and was slightly truncated. CONCLUSIONS: Problem with heart mispositioning during imaging on the CZT camera affects less than 1% of all performed studies. Morbid obesity is not a contraindication to perform myocardial perfusion scintigraphy with the use of a CZT camera because over 2/3 of the studies of very obese patients is diagnostic.
引用
收藏
页码:18 / 22
页数:5
相关论文
共 50 条
  • [31] Myocardial perfusion imaging using single-photon emission computed tomography with cadmium-zinc-telluride technology
    Konsek-Komorowska, Sonia J.
    Peczkowska, Mariola
    Cwikla, Jaroslaw B.
    NUCLEAR MEDICINE REVIEW, 2022, 25 (02) : 119 - 126
  • [32] Relationships between myocardial perfusion abnormalities and poststress left ventricular functional impairment on cadmium-zinc-telluride imaging
    Alessia Gimelli
    Riccardo Liga
    Assuero Giorgetti
    Annette Kusch
    Emilio Maria Pasanisi
    Paolo Marzullo
    European Journal of Nuclear Medicine and Molecular Imaging, 2015, 42 : 994 - 1003
  • [33] Validation with thallium-201 of a new cadmium-zinc-telluride (CZT) cardiac camera
    Songy, Bernard
    Guernou, Mohamed
    Lussato, David
    Geronazzo, Ricardo
    Queneau, Mathieu
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [34] Preliminary results with a Cadmium-Zinc-Telluride gamma camera for scintimammography
    O'Connor, MK
    Rhodes, DJ
    Phillips, SW
    Collins, DA
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 : S243 - S244
  • [35] Relationships between myocardial perfusion abnormalities and poststress left ventricular functional impairment on cadmium-zinc-telluride imaging
    Gimelli, Alessia
    Liga, Riccardo
    Giorgetti, Assuero
    Kusch, Annette
    Pasanisi, Emilio Maria
    Marzullo, Paolo
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (07) : 994 - 1003
  • [36] REDUCING STUDY TIME FOR REST/STRESS 99MTC-TETROFOSMIN MYOCARDIAL PERFUSION IMAGING WITH MYOCARDIAL BLOOD FLOW MEASUREMENT USING CADMIUM-ZINC-TELLURIDE CAMERA IMAGING
    Kadoya, Yoshito
    Naji, Khalid
    Chow, Benjamin J. W.
    Small, Gary R.
    Wells, R. Glenn
    Ruddy, Terrence D.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2023, 81 (08) : 1448 - 1448
  • [37] First Experience DaTSCAN Imaging Using Cadmium-Zinc-Telluride Gamma Camera SPECT
    Farid, Karim
    Queneau, Mathieu
    Guernou, Mohamed
    Lussato, David
    Poullias, Xavier
    Petras, Slavomir
    Caillat-Vigneron, Nadine
    Songy, Bernard
    CLINICAL NUCLEAR MEDICINE, 2012, 37 (08) : E211 - E212
  • [38] Diagnostic performance of different cardiac stress protocols for myocardial perfusion imaging for the diagnosis of coronary artery disease using a cadmium-zinc-telluride camera with invasive coronary angiography correlation
    Ochoa-Figueroa, M.
    Frias-Rose, J.
    Good, E.
    Sanchez-Rodriguez, V.
    Davidsson, A.
    Pagonis, C.
    REVISTA ESPANOLA DE MEDICINA NUCLEAR E IMAGEN MOLECULAR, 2023, 42 (05): : 281 - 288
  • [39] Prognostic value of the cadmium-zinc-telluride camera: A comparison with a conventional (Anger) camera
    Jorge Daniel Oldan
    Linda K. Shaw
    Paul Hofmann
    Matthew Phelan
    Jeffrey Nelson
    Robert Pagnanelli
    Salvador Borges-Neto
    Journal of Nuclear Cardiology, 2016, 23 : 1280 - 1287
  • [40] Technical validation of myocardial flow reserve measurement using a dynamic cadmium-zinc-telluride camera
    Niimi, T.
    Unno, K.
    Yoshida, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (SUPPL 1) : S502 - S502