Measurement of Common Carotid Artery Intima-Media Thickness in Clinical Practice: Comparison of B-Mode and RF-Based Technique

被引:29
|
作者
Schreuder, F. H. [1 ]
Graf, M. [2 ]
Hameleers, J. M. [2 ]
Mess, W. H. [1 ]
Hoeks, A. P. [2 ]
机构
[1] Maastricht Univ, Med Ctr, Dept Clin Neurophysiol, NL-6202 AZ Maastricht, Netherlands
[2] Maastricht Univ, Dept Biophys, Cardiovasc Res Inst Maastricht, NL-6202 AZ Maastricht, Netherlands
来源
ULTRASCHALL IN DER MEDIZIN | 2009年 / 30卷 / 05期
关键词
intima-media thickness; carotid artery; B-mode ultrasonography; radiofrequency; REPRODUCIBILITY; ATHEROSCLEROSIS; DISTENSIBILITY; METAANALYSIS; ROTTERDAM; STIFFNESS; DISEASE; RISK;
D O I
10.1055/s-0028-1109187
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Purpose: The common carotid artery intima-media thickness (CCA-IMT) is usually measured using B-mode ultrasound images. A different approach for CCA-IMT detection is based on radio frequency (RF) multiple M-line analysis. Materials and Methods: The present study explores the relationship between B-mode and RF measurement of CCA-IMT, as well as the reproducibility of both methods in 136 patients recently diagnosed with cardiovascular disease. Within one session, repeated measurements were made in the distal CCA bilaterally, using the B-mode (averaged over 10 mm) and RF technique (avers aging 12 Wines over 14 mm). Results: The two methods correlate well (Pearson r = 0.765). The CCA-IMT values measured with B-mode and RF were 0.779 +/- 0.196 mm and 0.734 +/- 0.172 mm, respectively. B-mode CCA-IMT is significantly larger than RF CCA-IMT (mean difference of 0.045 mm, SEM 7.8 mu m; t = 5.82; p < 0.001). In the multivariate regression analysis, carotid artery stenosis, inhomogeneous IMT and diabetes mellitus were the main predictors of differences between B-mode and RF CCA-IMT. The intrapatient variation for B-mode and RF-based CCA-IMT is comparable (0.05 +/- 0.04mm and 0.07 +/- 0.05 mm, respectively). Conclusion: CCA-IMT values measured with RF and B-mode have similar reproducibility and exhibit acceptable correlation, but RF CCA-IMT is significantly smaller. The difference between both methods is mainly due to advanced atherosclerosis. Hence, both methods can be used reliably to measure CCA-IMT in clinical practice.
引用
收藏
页码:459 / 465
页数:7
相关论文
共 50 条
  • [41] B-mode ultrasound common carotid artery intima-media thickness and external diameter: cross-sectional and longitudinal associations with carotid atherosclerosis in a large population sample
    Marsha L Eigenbrodt
    Zoran Bursac
    Richard E Tracy
    Jawahar L Mehta
    Kathryn M Rose
    David J Couper
    Cardiovascular Ultrasound, 6
  • [42] The role of measurement of intima-media thickness by B-mode ultrasound for epidemiological and pharmacological trials in atherosclerosis
    Faggiotto, A
    Baldassarre, D
    Paoletti, R
    XXI CONGRESS OF THE EUROPEAN SOCIETY OF CARDIOLOGY, 1999, : 195 - 201
  • [43] Carotid wall stress calculated with continuous intima-media thickness assessment using B-mode ultrasound
    Pascaner, A. F.
    Craiem, D.
    Casciaro, M. E.
    Danielo, R.
    Graf, S.
    Guevara, E.
    20TH ARGENTINEAN BIOENGINEERING SOCIETY CONGRESS (XX ARGENTINE BIOENGINEERING CONGRESS AND IX CONFERENCE OF CLINICAL ENGINEERING), (SABI 2015), 2016, 705
  • [44] Measurement of carotid artery intima-media thickness.: Analysis of reproducibility
    Baguet, JP
    Moreau-Gaudry, A
    Siché, JP
    Tremel, F
    Cinquin, P
    Mallion, JM
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 2000, 93 (09): : 1139 - 1144
  • [45] Carotid Artery Segmentation in Ultrasound Images and Measurement of Intima-Media Thickness
    Naik, Vaishali
    Gamad, R. S.
    Bansod, P. P.
    BIOMED RESEARCH INTERNATIONAL, 2013, 2013
  • [46] Manual B-Mode Versus Automated Radio-Frequency Carotid Intima-Media Thickness Measurements
    Dogan, Soner
    Plantinga, Yvonne
    Dijk, Joke M.
    van der Graaf, Yolanda
    Grobbee, Diederick E.
    Bots, Michiel L.
    JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2009, 22 (10) : 1137 - 1144
  • [47] Anisotropic edge-preserving smoothing in carotid B-mode ultrasound for improved segmentation and intima-media thickness (IMT) measurement
    Chan, RC
    Kaufhold, J
    Hemphill, LC
    Lees, RS
    Karl, WC
    COMPUTERS IN CARDIOLOGY 2000, VOL 27, 2000, 27 : 37 - 40
  • [48] Comparison between automated and manual measurements of carotid intima-media thickness in clinical practice
    Vilas Freire, Claudia Maria
    Pinho Ribeiro, Antonio Luiz
    Lima Barbosa, Felipe Batista
    Nogueira, Anelise Impelisiere
    Costa de Almeida, Maria Cristina
    Barbosa, Marcia Melo
    Quintao Lana, Angela Maria
    Simoes e Silva, Ana Cristina
    Ribeiro-Oliveira, Antonio, Jr.
    VASCULAR HEALTH AND RISK MANAGEMENT, 2009, 5 : 811 - 817
  • [49] Increased intima-media thickness of the common carotid artery in primary aldosteronism in comparison with essential hypertension
    Holaj, Robert
    Zelinka, Tomas
    Wichterle, Dan
    Petrak, Ondrej
    Strauch, Branislav
    Widimsky, Jiri, Jr.
    JOURNAL OF HYPERTENSION, 2007, 25 (07) : 1451 - 1457
  • [50] Comparison of Intima-media Thickness of the Common Carotid Artery in Dialysis and Kidney Transplant Recipient Patients
    Saedi, Daryoush
    Narooinejad, Minoo
    Saffari, Samira
    Khak, Mohammad
    EXPERIMENTAL AND CLINICAL TRANSPLANTATION, 2011, 9 (01) : 26 - 31