Measurement of Common Carotid Artery Intima-Media Thickness in Clinical Practice: Comparison of B-Mode and RF-Based Technique

被引:29
|
作者
Schreuder, F. H. [1 ]
Graf, M. [2 ]
Hameleers, J. M. [2 ]
Mess, W. H. [1 ]
Hoeks, A. P. [2 ]
机构
[1] Maastricht Univ, Med Ctr, Dept Clin Neurophysiol, NL-6202 AZ Maastricht, Netherlands
[2] Maastricht Univ, Dept Biophys, Cardiovasc Res Inst Maastricht, NL-6202 AZ Maastricht, Netherlands
来源
ULTRASCHALL IN DER MEDIZIN | 2009年 / 30卷 / 05期
关键词
intima-media thickness; carotid artery; B-mode ultrasonography; radiofrequency; REPRODUCIBILITY; ATHEROSCLEROSIS; DISTENSIBILITY; METAANALYSIS; ROTTERDAM; STIFFNESS; DISEASE; RISK;
D O I
10.1055/s-0028-1109187
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Purpose: The common carotid artery intima-media thickness (CCA-IMT) is usually measured using B-mode ultrasound images. A different approach for CCA-IMT detection is based on radio frequency (RF) multiple M-line analysis. Materials and Methods: The present study explores the relationship between B-mode and RF measurement of CCA-IMT, as well as the reproducibility of both methods in 136 patients recently diagnosed with cardiovascular disease. Within one session, repeated measurements were made in the distal CCA bilaterally, using the B-mode (averaged over 10 mm) and RF technique (avers aging 12 Wines over 14 mm). Results: The two methods correlate well (Pearson r = 0.765). The CCA-IMT values measured with B-mode and RF were 0.779 +/- 0.196 mm and 0.734 +/- 0.172 mm, respectively. B-mode CCA-IMT is significantly larger than RF CCA-IMT (mean difference of 0.045 mm, SEM 7.8 mu m; t = 5.82; p < 0.001). In the multivariate regression analysis, carotid artery stenosis, inhomogeneous IMT and diabetes mellitus were the main predictors of differences between B-mode and RF CCA-IMT. The intrapatient variation for B-mode and RF-based CCA-IMT is comparable (0.05 +/- 0.04mm and 0.07 +/- 0.05 mm, respectively). Conclusion: CCA-IMT values measured with RF and B-mode have similar reproducibility and exhibit acceptable correlation, but RF CCA-IMT is significantly smaller. The difference between both methods is mainly due to advanced atherosclerosis. Hence, both methods can be used reliably to measure CCA-IMT in clinical practice.
引用
收藏
页码:459 / 465
页数:7
相关论文
共 50 条
  • [1] Automated measurement of mean wall thickness in the common carotid artery by MRI: A comparison to intima-media thickness by B-mode ultrasound
    Underhill, Hunter R.
    Kerwin, William S.
    Hatsukami, Thomas S.
    Yuan, Chun
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (02) : 379 - 387
  • [2] Assessment of Common Carotid Artery Intima-Media Thickness in Hypertensive Individuals by B-Mode Ultrasonography
    Afzal, Maria
    Imran, Muhammad
    Malik, Ammara Saeed
    Rasheed, Noreen
    Ran, Fouzia
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2019, 13 (04): : 751 - 752
  • [3] Automatic intima-media thickness measurement of carotid artery wall in B-mode sonographic images
    Cheng, Da-Chuan
    Jiang, Xiaoyi
    Schmidt-Trucksaess, Amo
    Cheng, Kuo-Sheng
    2006 3RD IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1-3, 2006, : 912 - +
  • [4] Does B-mode common carotid artery intima-media thickness differ from M-mode?
    Van Bortel, LM
    Vanmolkot, FH
    van der Heijden-Spek, JJ
    Bregu, M
    Staessen, JA
    Hoeks, AP
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (10): : 1333 - 1336
  • [5] Individual measurement and significance of carotid intima, media, and intima-media thickness by B-mode ultrasonographic image processing
    Bae, Jang-Ho
    Kim, Wuon-Shik
    Rihal, Charanjit S.
    Lerman, Amir
    ARTERIOSCLEROSIS THROMBOSIS AND VASCULAR BIOLOGY, 2006, 26 (10) : 2380 - 2385
  • [6] Detections of Intima-Media Thickness in B-Mode Carotid Artery Images Using Segmentation Methods
    Savithri, V.
    Purushothaman, S.
    PROCEEDINGS OF INTERNATIONAL CONFERENCE ON INTERNET COMPUTING AND INFORMATION COMMUNICATIONS (ICICIC GLOBAL 2012), 2014, 216 : 471 - 482
  • [7] Simple clinical method of average intima-media thickness measurement in the common carotid artery
    Csanyi, A
    Egervari, A
    VASA-JOURNAL OF VASCULAR DISEASES, 1996, 25 (03): : 242 - 248
  • [8] Usefulness of carotid intima-media thickness measurement and peripheral B-mode ultrasound scan in the clinical screening of patients with coronary artery disease
    Balbarini, A
    Buttitta, F
    Limbruno, U
    Petronio, AS
    Baglini, R
    Strata, G
    Mariotti, R
    Ciccone, M
    Mariani, M
    ANGIOLOGY, 2000, 51 (04) : 269 - 279
  • [9] Quality control of B-mode ultrasonic measurement of carotid artery intima-media thickness: the European Lacidipine Study on Atherosclerosis
    Tang, R
    Hennig, M
    Bond, MG
    Hollweck, R
    Manica, G
    Zanchetti, A
    JOURNAL OF HYPERTENSION, 2005, 23 (05) : 1047 - 1054
  • [10] Automated RF versus manual B-mode common carotid intima-media thickness measurements in routine clinical practice: Direct comparison of risk factor relations and relations with future events
    Dogan, Soner
    Dijk, Joke M.
    van der Graaf, Yolanda
    Grobbee, Diederick E.
    Bots, Michiel L.
    CIRCULATION, 2007, 115 (08) : E268 - E268