Comparison of high-field (1.5 T) and low-field (0.2 T) MR imaging for the reliability of meniscal diagnosis

被引:6
|
作者
Rand, T
Imhof, H
Breitenseher, M
Happel, B
Turetschek, K
Schneider, B
Trattnig, S
机构
[1] UNIV VIENNA,EINRICHTUNG MAGNETRESONANZTOMOG,A-1090 VIENNA,AUSTRIA
[2] UNIV VIENNA,LUDWIG BOLTZMANN INST RADIOL TUMOR DIAGNOST,INST MED STAT,A-1090 VIENNA,AUSTRIA
来源
RADIOLOGE | 1997年 / 37卷 / 10期
关键词
low field strength MR imaging; meniscus; diagnostic reliability;
D O I
10.1007/s001170050285
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to evaluate the reliability of a low field strength MRI unit (0.2 T; Esaote, Biomedica) and a high field strength MRI unit (1.5 T; Siemens) in detecting meniscal lesions. Twenty patients were investigated on a 0.2-T low-field MRI unit and a 1.5-T high-field MRI unit for meniscal lesions. Evaluation was performed of four meniscal compartments (anterior and posterior medial and lateral meniscus), for a total of 80 observations. The reliability of diagnosis was ranked qualitatively on a three-point scale. Investigations of the high-field MRI unit revealed a meniscal tear in 6 patients and ruled out a tear in 14 patients. Evaluation of individual meniscal compartments revealed diagnosis with high confidence in 58 observations and with moderate confidence in 22 observations. Investigations of the low-field MRI unit revealed a meniscal tear in 5 patients and ruled out tears in 15 patients. Evaluation of individual meniscal compartments revealed diagnosis with high confidence in 39 observations, with moderate confidence in 38 observations, and with low confidence in 3 observations. The diagnosis in high-and low-field images was identical in 98.7 % of cases; in one patient observations from the high-field unit changed the therapeutic management. Diagnostic reliability was significantly higher with highfield strength MR imaging. Low field strength MR imaging shows comparatively high diagnostic accuracy for meniscal lesions compared to high field strength imaging; however, we demonstrated a significantly higher confidence in diagnosis with high-field MR imaging.
引用
收藏
页码:802 / 806
页数:5
相关论文
共 50 条
  • [31] High-field-strength MR imaging of the liver at 3.0 T:: Intraindividual comparative study with MR imaging at 1.5 T
    von Falkenhausen, Marcus M.
    Lutterbey, Goetz
    Morakkabati-Spitz, Nuschin
    Walter, Oliver
    Gieseke, Juergen
    Bloemer, Renate
    Willinek, Winfried A.
    Schild, Hans H.
    Kuhl, Christiane K.
    RADIOLOGY, 2006, 241 (01) : 156 - 166
  • [32] Low-field and high-field characterization of THUNDER actuators
    Ounaies, Z
    Mossi, K
    Smith, R
    Bernd, J
    SMART STRUCTURES AND MATERIALS 2001: ACTIVE MATERIALS: BEHAVIOR AND MECHANICS, 2001, 4333 : 399 - 407
  • [33] Open whole body low-field MR imaging: Are we there yet? Results of a prospective multicentre study with 401 patients comparing low-field and high-field MRI imaging to goldstandard
    Meri, T
    Scholz, M
    Langer, M
    Weiss, HD
    Wolf, KJ
    RADIOLOGY, 1999, 213P : 297 - 297
  • [34] High-field and low-field transport in amorphous semiconductors
    Cleve, B.
    Movaghar, B.
    Schumacher, R.
    Thomas, P.
    Journal of Non-Crystalline Solids, 1991, 137-38 (pt 1) : 415 - 418
  • [35] MR imaging of the knee at 0.2 T and 1.5 T: Correlation with arthroscopy
    Cotten, A
    Lapegue, F
    Boukhelifa, M
    Guiot, S
    Fossati, P
    Gougeon, F
    RADIOLOGY, 1998, 209P : 192 - 192
  • [36] Interventional MRI at high-field (1.5 T): Needle artifacts
    Liu, HY
    Martin, AJ
    Truwit, CL
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01): : 214 - 219
  • [37] Magnetic resonance imaging of hindfoot involvement in patients with spondyloarthritides: Comparison of low-field and high-field strength units
    Eshed, Iris
    Althoff, Christian E.
    Feist, Eugen
    Minden, Kirsten
    Schink, Tania
    Hamm, Bernd
    Hermann, Kay-Geert A.
    EUROPEAN JOURNAL OF RADIOLOGY, 2008, 65 (01) : 140 - 147
  • [38] Bottomley, Paul A.: The Development of High-Field NMR Imaging: 0.12T to 1.5T
    Bottomley, Paul A.
    eMagRes, 2007, 2007
  • [39] Dynamic contrast-enhanced imaging of the wrist in rheumatoid arthritis: dedicated low-field (0.25-T) versus high-field (3.0-T) MRI
    Ryan K. L. Lee
    James F. Griffith
    D. F. Wang
    L. Shi
    David K. W. Yeung
    Edmund K. Li
    L. S. Tam
    Skeletal Radiology, 2015, 44 : 1095 - 1101
  • [40] Comparison of high-field and low-field magnetic resonance images of cadaver limbs of horses
    Murray, R. C.
    Mair, T. S.
    Sherlock, C. E.
    Blunden, A. S.
    VETERINARY RECORD, 2009, 165 (10) : 281 - 288