An Integrated System for Real-Time Image Guided Cardiac Catheter Ablation

被引:0
|
作者
Rettmann, M. E. [1 ]
Holmes, D. R., III [1 ]
Su, Y. [1 ]
Cameron, B. M. [1 ]
Camp, J. J. [1 ]
Packer, D. L. [1 ]
Robb, R. A. [1 ]
机构
[1] Mayo Clin, Coll Med, Rochester, MN 55905 USA
关键词
cardiac ablation; electrophysiologic mapping; EP mapping;
D O I
暂无
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Minimally invasive cardiac catheter ablation procedures require effective visualization of the relevant heart anatomy and electrophysiology (EP). In a typical ablation procedure, the visualization tools available to the cardiologist include bi-plane fluoroscopy, real-time ultrasound, and a coarse 3D model which gives a rough representation of cardiac anatomy and electrical activity. Recently, there has been increased interest in incorporating detailed, patient specific anatomical data into the cardiac ablation procedure. We are currently developing a prototype system which both integrates a patient specific, preoperative data model into the procedure as well as fuses the various visualization modalities (i.e. fluoroscopy, ultrasound, EP) into a single display. In this paper, we focus on two aspects of the prototype system. First, we describe the framework for integrating the various system components, including an efficient communication protocol. Second, using a simple two-chamber phantom of the heart, we demonstrate the ability to integrate preoperative data into the ablation procedure. This involves the registration and visualization of tracked catheter points within the cardiac chambers of the preoperative model.
引用
收藏
页码:455 / 460
页数:6
相关论文
共 50 条
  • [21] Characterization of a real-time surface image-guided stereotactic positioning system
    Peng, Jean L.
    Kahler, Darren
    Li, Jonathan G.
    Samant, Sanjiv
    Yan, Guanghua
    Amdur, Robert
    Liu, Chihray
    MEDICAL PHYSICS, 2010, 37 (10) : 5421 - 5433
  • [22] Interactive real-time mapping and catheter ablation of the cavotricuspid isthmus guided by magnetic resonance imaging in a porcine model
    Hoffmann, Boris A.
    Koops, Andreas
    Rostock, Thomas
    Muellerleile, Kai
    Steven, Daniel
    Karst, Roman
    Steinke, Mark U.
    Drewitz, Imke
    Lund, Gunnar
    Koops, Susan
    Adam, Gerhard
    Willems, Stephan
    EUROPEAN HEART JOURNAL, 2010, 31 (04) : 450 - 456
  • [23] Radiofrequency ablation guided by real-time cardiovascular magnetic resonance
    Fusco, Armando
    De Santis, Domenico
    De Ruvo, Ermenegildo
    Calo, Leonardo
    Borrelli, Alessio
    Caruso, Damiano
    Tremamunno, Giuseppe
    Laghi, Andrea
    Simonetti, Giovanni
    Stefanini, Matteo
    RADIOLOGIA MEDICA, 2024, 129 (12): : 1823 - 1829
  • [24] Fast, automated, real-time 3D passive balloon catheter tracking during MRI-guided cardiac catheterization using orthogonal projection imaging and real-time image-based catheter detection
    Kowalik, Grzegorz T.
    Kerfoot, Eric
    Kunze, Karl
    Neji, Radhouene
    Moon, Tracy
    Mellor, Nina
    Razavi, Reza
    Pushparajah, Kuberan
    Roujol, Sebastien
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (01) : 311 - 320
  • [25] Real-time intra-cardiac ultrasound imaging of the ablation catheter's relation to the esophagus during left atrial ablation
    Ferguson, JD
    Helms, AS
    Hummel, JP
    Mounsey, JP
    DiMarco, JP
    Mangrum, JM
    CIRCULATION, 2005, 112 (17) : U617 - U617
  • [26] Novel use of an irrigated ablation catheter to monitor real-time hemodynamics during ablation
    Bhatia, Neal K.
    Iravanian, Shahriar
    Ravi, Namita
    Kiani, Soroosh
    Lloyd, Michael S.
    Westerman, Stacy B.
    Merchant, Faisal M.
    El-Chami, Mikhael F.
    Hoque, Azizul
    Shah, Anand D.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 (05) : 1111 - 1118
  • [27] A REAL-TIME INTEGRATED BACKSCATTER MEASUREMENT SYSTEM FOR QUANTITATIVE CARDIAC TISSUE CHARACTERIZATION
    THOMAS, LJ
    WICKLINE, SA
    PEREZ, JE
    SOBEL, BE
    MILLER, JG
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1986, 33 (01) : 27 - 32
  • [28] Accuracy of a novel image integration technique for real-time guided ablation procedures on computed tomographic images
    Dong, J
    Solomon, SB
    Lardo, A
    Brem, E
    Preiss, A
    Berger, RD
    Halperin, H
    Dickfeld, T
    CIRCULATION, 2005, 112 (17) : U438 - U438
  • [29] A piecewise function-to-structure registration algorithm for image guided cardiac catheter ablation
    Su, Yi
    Holmes, David R., III
    Rettmann, Maryam E.
    Robb, Richard A.
    MEDICAL IMAGING 2006: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2006, 6141
  • [30] Multiscale Registration of Real-Time and Prior MRI Data for Image-Guided Cardiac Interventions
    Xu, Robert
    Athavale, Prashant
    Nachman, Adrian
    Wright, Graham A.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (10) : 2621 - 2632