EVALUATION OF THE RETINAL HAZARD WITH 3D DIGITALLY ASSISTED VISUALIZATION SYSTEM AND CONVENTIONAL MICROSCOPE IN MACULAR SURGERIES

被引:2
|
作者
Horigome, Yuka [1 ]
Iwashita, Yume [1 ]
Hirono, Kazushi [1 ]
Inoue, Tatsuya [1 ]
Konno, Ami [2 ]
Kadonosono, Kazuaki [1 ]
Yanagi, Yasuo [1 ,3 ]
机构
[1] Yokohama City Univ, Dept Ophthalmol & Microtechnol, Sch Med, Yokohama, Kanagawa, Japan
[2] Asahikawa Med Univ, Dept Ophthalmol, Asahikawa, Hokkaido, Japan
[3] Yokohama City Univ, Dept Ophthalmol & Microtechnol, Sch Med, Yokohama, Kanagawa 2320024, Japan
关键词
vitrectomy; light hazard; epiretinal membrane; heads-up surgery; OPERATING MICROSCOPE; PHOTOTOXICITY; VITRECTOMY; LIGHT; ENDOILLUMINATION; ILLUMINATION; DAMAGE;
D O I
10.1097/IAE.0000000000003621
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:To investigate the retinal light hazard during macular surgery using a digital three-dimensional visualization system (3D) and a conventional microscope (CM).Design:Experimental study and retrospective evaluation of a case-control study.Subjects:A total of 20 and 10 patients who underwent pars plana vitrectomy for epiretinal membrane using 3D and CM, respectively.Methods:Spectral irradiances of endoilluminators were measured for representative settings used during core vitrectomy and macular manipulations with 3D and CM. From the medical record of the patients, time needed for core vitrectomy and macular manipulations was extracted. The total retinal light hazard index and the macular hazard index were calculated based on the spectral irradiances weighted by the standard functions. Total retinal light hazard index, macular hazard index, and the number of cases that exceeded the maximum permissible radiant power exposure were compared between the two groups.Results:The spectral irradiance were 1.6 and 3.9 mW/cm(2) for core vitrectomy and 3.4 and 8.1 mW/cm(2) for macular manipulations using typical settings for 3D and CM groups, respectively. The total retinal light hazard index ranged from 4.31 kJ/m(2) to 17.37 kJ/m(2) for 3D and 11.09 kJ/m(2) to 27.70 kJ/m(2) for CM groups, respectively, whereas the macular hazard index ranged from 2.93 kJ/m(2) to 14.58 kJ/m(2) for 3D and from 6.84 kJ/m(2) to 23.55 kJ/m(2) for CM, respectively (P < 0.001). One (5%) and 6 (60%) pars plana vitrectomy cases exceeded the threshold limits with 3D and CM groups, respectively (P < 0.05, chi-square test).Conclusion:The 3D digitally assisted visualization system offers significantly safer macular surgery compared with the CM, considering the potential retinal hazard.
引用
收藏
页码:2301 / 2306
页数:6
相关论文
共 50 条
  • [1] Comparison of 3D Digitally Assisted Visualization System with Current Standard Visualization for the Removal of Vitreous in a Preclinical Model
    Mura, Marco
    Martin, Wendy
    Williams, K. Keven
    AbuIon, Dina Joy K.
    CLINICAL OPHTHALMOLOGY, 2021, 15 : 4499 - 4505
  • [2] Comparison of clinical outcomes between "heads-up" 3D viewing system and conventional microscope in macular hole surgeries: A pilot study
    Kumar, Atul
    Hasan, Nasiq
    Kakkar, Prateek
    Mutha, Vineet
    Karthikeya, R.
    Sundar, Dheepak
    Ravani, Raghav
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2018, 66 (12) : 1816 - 1819
  • [3] 3D Visualization System in Descemet Membrane Endothelial Keratoplasty (DMEK): A Six-Month Comparison with Conventional Microscope
    Morelli, Alberto
    Ferrandina, Rosangela
    Favuzza, Eleonora
    Cennamo, Michela
    Mencucci, Rita
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (15)
  • [4] 3D Visualization System-Assisted Vitrectomy for Rhegmatogenous Retinal Detachment: Leave Out the Perfluorocarbon Liquid
    Guan, Lina
    Chen, Jiayu
    Tang, Yu
    Lu, Zhaolin
    Zhang, Zhengpei
    Ji, Sujuan
    Li, Meili
    Liu, Yalu
    Li, Suyan
    Liu, Haiyang
    OPHTHALMOLOGY AND THERAPY, 2023, 12 (03) : 1611 - 1619
  • [5] 3D Visualization System-Assisted Vitrectomy for Rhegmatogenous Retinal Detachment: Leave Out the Perfluorocarbon Liquid
    Lina Guan
    Jiayu Chen
    Yu Tang
    Zhaolin Lu
    Zhengpei Zhang
    Sujuan Ji
    Meili Li
    Yalu Liu
    Suyan Li
    Haiyang Liu
    Ophthalmology and Therapy, 2023, 12 : 1611 - 1619
  • [6] Automated 3D Visualization of Electron Microscope Tomograms
    Yamashita, Chikara
    Nishio, Koji
    Kobori, Ken-ichi
    WSCG 2011: COMMUNICATION PAPERS PROCEEDINGS, 2011, : 157 - 162
  • [7] Framework and prototype for B/S based 3D visualization of a mine and hazard system
    Ma, Bin
    Li, Zhong-Xue
    Hao, Xiu-Qiang
    Beijing Keji Daxue Xuebao/Journal of University of Science and Technology Beijing, 2009, 31 (01): : 10 - 14
  • [8] Use of Color Channel Optimization in 3D Heads-Up Vitrectomy vs Standard Operating Microscope for Macular Surgeries
    Agarwal, Aniruddha
    Ghazi, Nicola
    El Ghrably, Ibraheem
    Iovino, Claudio
    Peiretti, Enrico
    Alcibahy, Yasmine
    Menia, Nitin Kumar
    Chhablani, Jay
    Pichi, Francesco
    Grewal, Dilraj
    JOURNAL OF VITREORETINAL DISEASES, 2024, 8 (05) : 517 - 523
  • [9] 3D visualization system and standard operating microscope for ease of visualization and surgeon comfort during phacoemulsification surgery
    Shetty, Naren
    Aishwarya
    Nayak, Rohitha P.
    Kaweri, Luci
    Kanaradi, Ravi Krishna
    Pisharody, Anchana A.
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2024, 72 (01) : 56 - 62
  • [10] 3D digital image microscope system-assisted vasovasostomy and vasoepididymostomy in rats
    Li, Peng
    Liu, Na-Chuan
    Zhi, Er-Lei
    Yao, Chen-Cheng
    Zhao, Zhi-Liang
    Yu, Zhi-Yong
    Li, Qi-Meng
    Huang, Yu-Hua
    Ju, Jie-Chang
    Huang, Wen-Bin
    Rozi, Husanjan
    Ji, Zhi-Yong
    Guo, San-Wei
    Tian, Ru-Hui
    Li, Zheng
    ASIAN JOURNAL OF ANDROLOGY, 2021, 23 (04) : 396 - +