Test-Retest Reliability and Concurrent Validity of an fMRI-Compatible Pneumatic Vibrator to Stimulate Muscle Proprioceptors

被引:13
|
作者
Goossens, Nina [1 ]
Janssens, Lotte [1 ,2 ]
Pijnenburg, Madelon [1 ]
Caeyenberghs, Karen [3 ]
Van Rompuy, Charlotte [1 ]
Meugens, Paul [1 ]
Sunaert, Stefan [4 ]
Brumagne, Simon [1 ]
机构
[1] Univ Leuven, KU Leuven, Fac Kinesiol & Rehabil Sci, Dept Rehabil Sci, Tervuursevest 101,Box 1501, B-3001 Leuven, Belgium
[2] Hasselt Univ, Biomed Res Inst, Agoralaan A, B-3590 Diepenbeek, Belgium
[3] Australian Catholic Univ, Fac Hlth Sci, Sch Psychol, Melbourne Campus St Patrick,Locked Bag 4115, Fitzroy Mdc, Vic 3065, Australia
[4] Univ Leuven, KU Leuven, Dept Imaging & Pathol, UZ Herestr 49,Box 7003, B-3000 Leuven, Belgium
关键词
Reproducibility of results; proprioception; muscle spindle; postural balance; vibration; magnetic resonance imaging; LOW-BACK-PAIN; POSITION SENSE; POSTURAL SWAY; BALANCE; YOUNG; MOTOR; CORTEX; INSTABILITY; PARKINSONS; MOVEMENT;
D O I
10.1163/22134808-00002526
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Processing proprioceptive information in the brain is essential for optimal postural control and can be studied with proprioceptive stimulation, provided by muscle vibration, during functional magnetic resonance imaging (fMRI). Classic electromagnetic muscle vibrators, however, cannot be used in the high-strength magnetic field of the fMRI scanner. Pneumatic vibrators offer an fMRI-compatible alternative. However, whether these devices produce reliable and valid proprioceptive stimuli has not been investigated, although this is essential for these devices to be used in longitudinal research. Test-retest reliability and concurrent validity of the postural response to muscle vibration, provided by custom-made fMRI-compatible pneumatic vibrators, were assessed in a repeated-measures design. Mean center of pressure (CoP) displacements during, respectively, ankle muscle and back muscle vibration (45-60 Hz, 0.5 mm) provided by an electromagnetic and a pneumatic vibrator were measured in ten young healthy subjects. The test was repeated on the same day and again within one week. Intraclass correlation coefficients (ICC) were calculated to assess (a) intra-and interday reliability of the postural responses to, respectively, pneumatic and electromagnetic vibration, and (b) concurrent validity of the response to pneumatic compared to electromagnetic vibration. Test-retest reliability of mean CoP displacements during pneumatic vibration was good to excellent (ICCs = 0.64-0.90) and resembled that of responses to electromagnetic vibration (ICCs = 0.64-0.94). Concurrent validity of the postural effect of pneumatic vibration was good to excellent (ICCs = 0.63-0.95). In conclusion, the proposed fMRI-compatible pneumatic vibrator can be used with confidence to stimulate muscle spindles during fMRI to study central processing of proprioception.
引用
收藏
页码:465 / 492
页数:28
相关论文
共 50 条
  • [1] A New Vibrator to Stimulate Muscle Proprioceptors in fMRI
    Montant, Marie
    Romaiguere, Patricia
    Roll, Jean-Pierre
    HUMAN BRAIN MAPPING, 2009, 30 (03) : 990 - 997
  • [2] Test-retest reliability and concurrent validity of a digital manovacuometer
    Vieira, Danielle
    Pessoa, Isabela Sclauser
    Pereira, Hugo
    Aguiar, Larissa
    Tagliaferri, Thaysa
    Silva, Luisa
    Parreira, Veronica
    EUROPEAN RESPIRATORY JOURNAL, 2014, 44
  • [3] Concurrent validity and test-retest reliability of the information processing styles
    Baloglu, M
    Zelhart, PF
    PSYCHOLOGICAL REPORTS, 2000, 87 (03) : 973 - 979
  • [4] Muscle-Strengthening Exercise Questionnaire (MSEQ): an assessment of concurrent validity and test-retest reliability
    Shakespear-Druery, Jane
    De Cocker, Katrien
    Biddle, Stuart J. H.
    Bennie, Jason
    BMJ OPEN SPORT & EXERCISE MEDICINE, 2022, 8 (01):
  • [5] The concurrent validity and test-retest reliability of a submaximal exercise test in adolescents with autism
    Arkesteyn, Anke
    Cornelissen, Veronique
    Steyaert, Jean
    Claes, Jomme
    Michielsen, Matthijs
    Van Damme, Tine
    DISABILITY AND REHABILITATION, 2025, 47 (02) : 501 - 511
  • [6] The concurrent validity and test-retest reliability of a visuospatial working memory task
    Hockey, A
    Geffen, G
    INTELLIGENCE, 2004, 32 (06) : 591 - 605
  • [7] Test-retest reliability and concurrent validity of the expanded skills confidence inventory
    Robinson, CH
    Betz, NE
    JOURNAL OF CAREER ASSESSMENT, 2004, 12 (04) : 407 - 422
  • [8] Test-retest rehability concurrent validity of a movacuometer
    Vieira, Danielle
    Pessoa, Isabela Sclauser
    Pereira, Hugo
    Aguiar, Larissa
    Tagliaferri, Thaysa
    Silva, Luisa
    Parreira, Veronica
    EUROPEAN RESPIRATORY JOURNAL, 2014, 44
  • [9] The test-retest reliability and concurrent validity of 360° turn test in patients with knee osteoarthritis
    Yarar, Haci Ahmet
    Narin, Selnur
    Erduran, Mehmet
    Gurbanov, Iftikhar
    SOMATOSENSORY AND MOTOR RESEARCH, 2024, 41 (02): : 90 - 96
  • [10] Test-Retest Reliability and Concurrent Validity of the 30 second Sit to Stand Test in Adolescents
    Robinson, Katie J.
    Lubans, David R.
    Mavilidi, Myrto F.
    Ortega, Francisco B.
    Riley, Nicholas
    MEASUREMENT IN PHYSICAL EDUCATION AND EXERCISE SCIENCE, 2024, 28 (01) : 109 - 118