Statistical reconstruction for x-ray CT systems with non-continuous detectors

被引:18
|
作者
Zbijewski, Wojciech [1 ]
Defrise, Michel
Viergever, Max A.
Beekman, Freek J.
机构
[1] UMC Utrecht, Image Sci Inst, Dept Nucl Med, NL-3584 CG Utrecht, Netherlands
[2] UMC Utrecht, Rudolf Magnus Inst Neurosci, NL-3584 CG Utrecht, Netherlands
[3] AZ Vrije Univ Brussel, Univ Hosp, Dept Nucl Med, B-1090 Brussels, Belgium
[4] UMC Utrecht, Image Sci Inst, NL-3584 CX Utrecht, Netherlands
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2007年 / 52卷 / 02期
关键词
D O I
10.1088/0031-9155/52/2/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We analyse the performance of statistical reconstruction (SR) methods when applied to non-continuous x-ray detectors. Robustness to projection gaps is required in x-ray CT systems with multiple detector modules or with defective detector pixels. In such situations, the advantage of statistical reconstruction is that it is able to ignore missing or faulty pixels and that it makes optimal use of the remaining line integrals. This potentially obviates the need to fill the sinogram discontinuities by interpolation or any other approximative preprocessing techniques. In this paper, we apply SR to cone beam projections of (i) a hypothetical modular detector micro-CT scanner and of (ii) a system with randomly located defective detector elements. For the modular-detector system, SR produces reconstruction volumes free of noticeable gap-induced artefacts as long as the location of detector gaps and selection of the scanning range provide complete object sampling in the central imaging plane. When applied to randomly located faulty detector elements, SR produces images free of substantial ring artefacts even for cases where defective pixels cover as much as 3% of the detector area.
引用
收藏
页码:403 / 418
页数:16
相关论文
共 50 条
  • [21] FBP initialization for transition artifacts reduction in statistical X-ray CT reconstruction.
    Zbijewski, W
    Beekman, FJ
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2970 - 2972
  • [22] A Splitting-Based Iterative Algorithm for Accelerated Statistical X-Ray CT Reconstruction
    Ramani, Sathish
    Fessler, Jeffrey A.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (03) : 677 - 688
  • [23] Sparse reconstruction methods in X-ray CT
    Abascal, J. F. P. J.
    Abella, M.
    Mory, C.
    Ducros, N.
    de Molina, C.
    Marinetto, E.
    Peyrin, F.
    Desco, M.
    DEVELOPMENTS IN X-RAY TOMOGRAPHY XI, 2017, 10391
  • [24] Ordered subset reconstruction for x-ray CT
    Beekman, FJ
    Kamphuis, C
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (07): : 1835 - 1844
  • [25] Iterative reconstruction methods in X-ray CT
    Beister, Marcel
    Kolditz, Daniel
    Kalender, Willi A.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02): : 94 - 108
  • [26] Non-Gaussian noise in x-ray and γ-ray detectors
    Chen, LY
    Barrett, HH
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 366 - 376
  • [27] Energy Reconstruction for Events Detected in TES X-ray Detectors
    Ceballos, M. T.
    Cardiel, N.
    Cobo, B.
    ASTRONOMICAL DATA ANALYSIS SOFTWARE AND SYSTEMS: XXIV, 2015, 495 : 331 - 334
  • [28] Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction
    Nien, Hung
    Fessler, Jeffrey A.
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [29] Improving noise-resolution tradeoffs in X-ray CT imaging by statistical image reconstruction
    Williamson, J.
    Politte, D.
    O'Sullivan, J.
    Keesing, D.
    Whiting, B.
    Lazos, D.
    Evans, J.
    MEDICAL PHYSICS, 2007, 34 (06) : 2346 - 2346
  • [30] XEUS - the X-ray mirrors and X-ray detectors
    Willingale, R
    X-RAY ASTRONOMY 2000, 2001, 234 : 589 - 596