Long-term dosimetric stability of multiple TomoTherapy delivery systems

被引:12
|
作者
Smilowitz, Jennifer B. [1 ,2 ]
Dunkerley, David [2 ]
Hill, Patrick M. [1 ]
Yadav, Poonam [1 ]
Geurts, Mark W. [1 ]
机构
[1] Univ Wisconsin Madison, Dept Human Oncol, Madison, WI 53706 USA
[2] Univ Wisconsin Madison, Dept Med Phys, Madison, WI USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2017年 / 18卷 / 03期
关键词
dosimetric stability; tomotherapy; Radixact; MATCH-LINE WEDGE; CRANIOSPINAL IRRADIATION; NECK-CANCER; IMRT; JUNCTION; FIELDS; COLLIMATOR; HEAD;
D O I
10.1002/acm2.12085
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dosimetric stability of six TomoTherapy units was analyzed to investigate changes in performance over time and with system upgrades. Energy and output were tracked using monitor chamber signal, onboard megavoltage computed tomography (MVCT) detector profile, and external ion chamber measurements. The systems (and monitoring periods) include three Hi-Art (67, 61, and 65 mos.), two TomoHDA (31 and 26 mos.), and one Radixact unit (11 mos.), representing approximately 10 years of clinical use. The four newest systems use the Dose Control Stability (DCS) system and Fixed Target Linear Accelerator (linac) (FTL). The output stability is reported as deviation from reference monitor chamber signal for all systems and/ or from an external chamber signal. The energy stability was monitored using relative (center versus off-axis) MVCT detector signal (beam profile) and/ or the ratio of chamber measurements at 2 depths. The clinical TomoHDA data were used to benchmark the Radixact stability, which has the same FTL but runs at a higher dose rate. The output based on monitor chamber data of all systems is very stable. The standard deviation of daily output on the non-DCS systems was 0.94-1.52%. As expected, the DCS systems had improved standard deviation: 0.004-0.06%. The beam energy was also very stable for all units. The standard deviation in profile flatness was 0.23-0.62% for rotating target systems and 0.04-0.09% for FTL. Ion chamber output and PDD ratios supported these results. The output stability on the Radixact system during extended treatment delivery (20, 30, and 40 min) was comparable to a clinical TomoHDA system. For each system, results are consistent between different measurement tools and techniques, proving not only the dosimetric stability, but also these quality parameters can be confirmed with various metrics. The replacement history over extended time periods of the major dosimetric components of the different delivery systems (target, linac, and magnetron) is also reported.
引用
收藏
页码:137 / 143
页数:7
相关论文
共 50 条
  • [31] Dosimetric Evaluation of a Delivery Verification and Dose Reconstruction Method for Helical Tomotherapy
    Hesston, R.
    Gibbons, J.
    Hogstrom, K.
    Chen, Q.
    Chapman, T.
    Olivera, G.
    Ruchala, K.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [32] LONG-TERM INTRAPERITONEAL INSULIN DELIVERY
    PITT, HA
    SAUDEK, CD
    ZACUR, HA
    ANNALS OF SURGERY, 1992, 216 (04) : 483 - 492
  • [33] Long-term clinical experience with helical tomotherapy for head and neck cancer
    Chen, Allen M.
    Daly, Megan E.
    Cui, Jing
    Perks, Julian
    Purdy, James A.
    Benedict, Stanley
    JOURNAL OF RADIATION ONCOLOGY, 2014, 3 (04) : 355 - 361
  • [34] Long-term Stability of Planetary Systems Formed from a Transitional Disk
    Bowens, Rory
    Shannon, Andrew
    Dawson, Rebekah
    Dong, Jiayin
    ASTROPHYSICAL JOURNAL, 2023, 944 (02):
  • [35] Cryogenic performance and long-term stability of metal optics and optical systems
    Robichaud, JL
    Wang, D
    Mastandrea, A
    INFRARED ASTRONOMICAL INSTRUMENTATION, PTS 1-2, 1998, 3354 : 178 - 186
  • [36] Long-term stability of guided wave electromagnetic acoustic transducer systems
    Herdovics, Balint
    Cegla, Frederic
    STRUCTURAL HEALTH MONITORING-AN INTERNATIONAL JOURNAL, 2020, 19 (01): : 3 - 11
  • [37] Long-term Voltage Stability Analysis and Network Topology in Power Systems
    Rincon, R.
    Pavas, A.
    Mojica-Nava, E.
    2016 IEEE PES INNOVATIVE SMART GRID TECHNOLOGIES CONFERENCE EUROPE (ISGT-EUROPE), 2016,
  • [38] DOSIMETRIC COMPLEX FOR LONG-TERM MANNED SPACE-FLIGHTS
    AKATOV, Y
    KOVALEV, EE
    SAKOVICH, VA
    DEME, S
    FEHER, I
    NGUYEN, VD
    NUCLEAR TRACKS AND RADIATION MEASUREMENTS, 1992, 20 (01): : 7 - 11
  • [39] DOSIMETRIC PREDICTORS OF LONG-TERM DYSPHAGIA FOLLOWING OROPHARYNGEAL IMRT
    Barringer, Denise
    Hutcheson, K.
    Tucker, S.
    Dong, L.
    Schwartz, D.
    Lewin, J.
    DYSPHAGIA, 2009, 24 (04) : 450 - 450
  • [40] Dosimetric predictors of long-term dysphagia following oropharyngeal IMRT
    Lewin, J. S.
    Hutcheson, K.
    Barringer, D.
    Tucker, S. L.
    Rosenthal, D. I.
    Chao, C.
    Morrison, W.
    Garden, A. S.
    Dong, L.
    Schwartz, D. L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S34 - S34