Long-term dosimetric stability of multiple TomoTherapy delivery systems

被引:12
|
作者
Smilowitz, Jennifer B. [1 ,2 ]
Dunkerley, David [2 ]
Hill, Patrick M. [1 ]
Yadav, Poonam [1 ]
Geurts, Mark W. [1 ]
机构
[1] Univ Wisconsin Madison, Dept Human Oncol, Madison, WI 53706 USA
[2] Univ Wisconsin Madison, Dept Med Phys, Madison, WI USA
来源
关键词
dosimetric stability; tomotherapy; Radixact; MATCH-LINE WEDGE; CRANIOSPINAL IRRADIATION; NECK-CANCER; IMRT; JUNCTION; FIELDS; COLLIMATOR; HEAD;
D O I
10.1002/acm2.12085
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dosimetric stability of six TomoTherapy units was analyzed to investigate changes in performance over time and with system upgrades. Energy and output were tracked using monitor chamber signal, onboard megavoltage computed tomography (MVCT) detector profile, and external ion chamber measurements. The systems (and monitoring periods) include three Hi-Art (67, 61, and 65 mos.), two TomoHDA (31 and 26 mos.), and one Radixact unit (11 mos.), representing approximately 10 years of clinical use. The four newest systems use the Dose Control Stability (DCS) system and Fixed Target Linear Accelerator (linac) (FTL). The output stability is reported as deviation from reference monitor chamber signal for all systems and/ or from an external chamber signal. The energy stability was monitored using relative (center versus off-axis) MVCT detector signal (beam profile) and/ or the ratio of chamber measurements at 2 depths. The clinical TomoHDA data were used to benchmark the Radixact stability, which has the same FTL but runs at a higher dose rate. The output based on monitor chamber data of all systems is very stable. The standard deviation of daily output on the non-DCS systems was 0.94-1.52%. As expected, the DCS systems had improved standard deviation: 0.004-0.06%. The beam energy was also very stable for all units. The standard deviation in profile flatness was 0.23-0.62% for rotating target systems and 0.04-0.09% for FTL. Ion chamber output and PDD ratios supported these results. The output stability on the Radixact system during extended treatment delivery (20, 30, and 40 min) was comparable to a clinical TomoHDA system. For each system, results are consistent between different measurement tools and techniques, proving not only the dosimetric stability, but also these quality parameters can be confirmed with various metrics. The replacement history over extended time periods of the major dosimetric components of the different delivery systems (target, linac, and magnetron) is also reported.
引用
收藏
页码:137 / 143
页数:7
相关论文
共 50 条
  • [1] Long Term Dosimetric Stability of 6 TomoTherapy Systems
    Smilowitz, J.
    Dunkerley, D.
    Geurts, M.
    Hill, P.
    Yadav, P.
    MEDICAL PHYSICS, 2016, 43 (06) : 3360 - 3360
  • [2] Dosimetric Impact of Dose Servo Control on Long Helical Tomotherapy Treatment Delivery
    Lam, W.
    Geng, H.
    Kong, C.
    Yang, B.
    Chiu, T.
    Cheung, K.
    Yu, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S734 - S734
  • [3] Long-term vaginal antibody delivery: Delivery systems and biodistribution
    Saltzman, WM
    Sherwood, JK
    Adams, DR
    Haller, P
    BIOTECHNOLOGY AND BIOENGINEERING, 2000, 67 (03) : 253 - 264
  • [4] Dosimetric validation of clinical serial tomotherapy delivery
    Sanford, R
    Al-Ghazi, M
    Chung, H
    Yakoob, R
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 2067 - 2070
  • [5] Long-term stability of planets in binary systems
    Holman, MJ
    Wiegert, PA
    ASTRONOMICAL JOURNAL, 1999, 117 (01): : 621 - 628
  • [6] Dosimetric Analysis, Acute Toxicity and Long-Term Outcome of Craniospinal Irradiation Using Helical Tomotherapy in Children and Adults
    Coassin, E.
    Drigo, A.
    Chiovati, P.
    Polesel, J.
    Gigante, M.
    Cinalli, G.
    Masotto, B.
    Passone, E.
    Kiren, V.
    Sartor, G.
    Franchin, G.
    Brandes, A. A.
    Mascarin, M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E624 - E624
  • [7] CERAMIC SYSTEMS FOR LONG-TERM DELIVERY OF CHEMICALS AND BIOLOGICALS
    BAJPAI, PK
    BENGHUZZI, HA
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (12): : 1245 - 1266
  • [8] LONG-TERM STABILITY OF WASTE LIGNINS IN AQUATIC SYSTEMS
    BOUVENG, HO
    SOLYOM, P
    SVENSK PAPPERSTIDNING-NORDISK CELLULOSA, 1973, 76 (01): : 26 - 29
  • [9] Predicting the long-term stability of compact multiplanet systems
    Tamayo, Daniel
    Cranmer, Miles
    Hadden, Samuel
    Rein, Hanno
    Battaglia, Peter
    Obertas, Alysa
    Armitage, Philip J.
    Ho, Shirley
    Spergel, David N.
    Gilbertson, Christian
    Hussain, Naireen
    Silburt, Ari
    Jontof-Hutter, Daniel
    Menou, Kristen
    PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2020, 117 (31) : 18194 - 18205