Femoral head material loss at the head-neck junction in total hip arthroplasty: the effect of head size, stem material and stem offset

被引:8
|
作者
Valente, Giuseppe [1 ]
Lanting, Brent [1 ]
MacDonald, Steven [1 ]
Teeter, Matthew G. [2 ,3 ]
Van Citters, Douglas [4 ]
Howard, James [1 ]
机构
[1] Western Univ, London Hlth Sci Ctr, Schulich Sch Med & Dent, Div Orthopaed Surg,Dept Surg,Univ Hosp, London, ON, Canada
[2] Western Univ, Schulich Sch Med & Dent, Dept Med Biophys, London, ON, Canada
[3] Lawson Hlth Res Inst, Surg Innovat Program, London, ON, Canada
[4] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
关键词
Corrosion; metal-on-polyethylene; taper; total hip arthroplasty; trunnion; trunnionosis; TAPER JUNCTION; CORROSION; METAL; TRUNNIONOSIS; COMPONENTS; INTERFACE;
D O I
10.1177/1120700018814473
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: Material loss at the head-neck junction in total hip arthroplasty may cause adverse clinical symptoms and implant failure. The purpose of this study was to quantitatively examine the effects of head size, stem material and stem offset on material loss of the head-neck taper interface of a single trunnion design in retrieval implants of metal on polyethylene bearing surfaces. Methods: A retrieval study was performed to identify all 28-mm and 32-mm femoral heads from a single implant/taper design implanted for >2 years. This included n = 56 of the 28-mm heads and n = 23 of the 32-mm heads. The 28-mm femoral heads were matched to 32-mm femoral heads based on time in vivo and head length. A coordinate measuring machine was used to determine maximum linear corrosion depth (MLD). Differences in MLD for head diameter, stem material, and stem offset were determined. Results: There were no differences between groups for age, gender, BMI, or implantation time. There was no difference in MLD between 28 mm and 32 mm matched paired head diameters (p = 0.59). There was also no difference in MLD between titanium or cobalt-chromium stems (p = 0.79), and regular or high-offset stems (p = 0.95). Conclusion: There is no statistical difference in femoral head MLD at the head-neck junction in THA between 28-mm and 32-mm matched paired femoral heads, similar or mixed alloy coupled femoral head stem constructs, and regular or high offset stems.
引用
收藏
页码:647 / 651
页数:5
相关论文
共 50 条
  • [41] Femoral head-neck offset in the Indian population: A CT based study
    Malhotra, Rajesh
    Kannan, Arun
    Kancherla, Ramprasad
    Khatri, Dharmesh
    Kumar, Vijay
    INDIAN JOURNAL OF ORTHOPAEDICS, 2012, 46 (02) : 212 - 215
  • [42] Femoral head-neck offset in the Indian population: A CT based study
    Rajesh Malhotra
    Arun Kannan
    Ramprasad Kancherla
    Dharmesh Khatri
    Vijay Kumar
    Indian Journal of Orthopaedics, 2012, 46 : 212 - 215
  • [43] Mathematical evaluation of jumping distance in total hip arthroplasty Influence of abduction angle, femoral head offset, and head diameter
    Sariali, Elhadi
    Lazennec, Jean Yves
    Khiami, Frederic
    Catonne, Yves
    ACTA ORTHOPAEDICA, 2009, 80 (03) : 277 - 282
  • [44] Contribution of the BioballTM head–neck adapter to the restoration of femoral offset in hip revision arthroplasty with retention of a well-fixed cup and stem
    Clara Chimeno
    Jenaro Ángel Fernández-Valencia
    Alfonso Alías
    Adrià Serra
    Yury Postnikov
    Andrés Combalia
    Ernesto Muñoz-Mahamud
    International Orthopaedics, 2023, 47 : 2245 - 2251
  • [45] Assessing Taper Geometry, Head Size, Head Material, and Their Interactions in Taper Fretting Corrosion of Retrieved Total Hip Arthroplasty Implants
    El-Zein, Zein S.
    Gehrke, Corinn K.
    Croley, J. Sawyer
    Siljander, Matthew P.
    Mallow, Murphy A.
    Flierl, Michael A.
    Verner, James J.
    Baker, Erin A.
    JOURNAL OF ARTHROPLASTY, 2021, 36 (07): : S386 - +
  • [46] Influence of the taper on the fretting wear of the femoral stem-femoral head taper junction in total hip prosthesis
    Capitanu, Lucian
    Badita, Liliana-Laura
    Tiganesteanu, Constantin
    Florescu, Virgil
    INDUSTRIAL LUBRICATION AND TRIBOLOGY, 2022, 74 (04) : 385 - 391
  • [47] Understanding the role of head size and neck length in micromotion generation at the taper junction in total hip arthroplasty
    Bologna, Federico A.
    Putame, Giovanni
    Audenino, Alberto L.
    Terzini, Mara
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [48] Ceramic femoral head fractures in total hip arthroplasty
    Fritsch, EW
    Gleitz, M
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (328) : 129 - 136
  • [49] Alternatives to total hip arthroplasty in osteonecrosis of the femoral head
    Thornhill, TS
    ORTHOPEDICS, 2001, 24 (09) : 861 - 863
  • [50] Femoral head fractures: hemiarthroplasty or total hip arthroplasty?
    Ullmark, Gosta
    HIP INTERNATIONAL, 2014, 24 : S12 - S14