Multi-Energy CT with Triple X-ray Beams and Photon-Counting-Detector CT for Simultaneous Imaging of Two Contrast Agents: An Experimental Comparison

被引:0
|
作者
Ren, Liqiang [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55902 USA
基金
美国国家卫生研究院;
关键词
Multi-energy CT (MECT); Photon-counting-detector CT (PCD-CT); Dual-energy CT (DECT); Material Decomposition; Multiple Contrast Agents; DUAL-ENERGY CT; IODINE;
D O I
10.1117/12.2513460
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Multi-energy CT (MECT) enabled by energy-resolved photon-counting-detector CT (PCD-CT) is promising for material specific imaging with multiple contrast agents. However, non-idealities of the PCD such as pulse pileup, K-edge escape, and charge sharing may degrade the spectral performance. To perform MECT, an alternative approach was proposed by extending a "Twin Beam" design to a dual-source CT scanner with energy-integrating-detector (EID) by operating one or both sources in the "Twin Beam" mode to acquire three (triple beam configuration) or four (quadruple-beam configuration) distinct X-ray beam measurements. Previous computer simulation studies demonstrated that the image quality and dose efficiency of the triple-beam configuration were comparable to that in PCD-CT for a three-material decomposition task involving iodine, bismuth, and water. The purpose of this work is to experimentally validate the proposed triple-beam MECT technique in comparison with PCD-CT. To mimic the dual-source triple-beam acquisition, two separate scans, one at 80 kV and the other at 120 kV operated in the "Twin Beam" mode, were performed on a single-source CT scanner. Two potential clinical applications of MECT for multiple contrast agents were investigated: iodine/gadolinium for biphasic liver imaging and iodine/bismuth for small bowel imaging. The results indicate that the imaging performance of the EID-based MECT may be comparable to that on the current PCD-CT platform for both the iodine/gadolinium and the iodine/bismuth material decomposition tasks.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Photon counting energy dispersive detector arrays for x-ray imaging
    Iwanczyk, J. S.
    Nygard, E.
    Meirav, O.
    Arenson, J.
    Barber, W. C.
    Hartsough, N. E.
    Malakhov, N.
    Wessel, J. C.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2741 - +
  • [32] Nanoceria as Safe Contrast Agents for X-ray CT Imaging
    Garcia, Ana
    Camara, Juan Antonio
    Boullosa, Ana Maria
    Gusta, Muriel F.
    Mondragon, Laura
    Schwartz Jr, Simo
    Casals, Eudald
    Abasolo, Ibane
    Bastus, Neus G.
    Puntes, Victor
    NANOMATERIALS, 2023, 13 (15)
  • [33] Photon Counting X-Ray Detectors and DR and CT Imaging Methods
    Taguchi, K.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [34] Imaging Characterization of a Multi-Energy CT with Quasi-Monochromatic X-ray source
    Rossi, Pier Luca
    Andreani, Lucia
    Bollini, Dante
    Bontempi, Marco
    Cappelli, Stefano
    Zuffa, Mirco
    Margotti, Anselmo
    Labanti, Claudio
    Baldazzi, Giuseppe
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2832 - 2835
  • [35] The Effect of Photon Energy Weighting on X-ray Imaging Based on a Photon Counting Detector
    Choi, Yu-Na
    Lee, Seung-Wan
    Cho, Hyo-Min
    Ryu, Hyun-Ju
    Lee, Young-Jin
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2011, 59 (05) : 3114 - 3119
  • [36] Solid-state photon-counting hybrid detector array for high-resolution multi-energy X-ray imaging
    Sia, R.
    Kleinfelder, S.
    Nagarkar, V. V.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 652 (01): : 470 - 473
  • [37] Modeling the performance of a photon counting x-ray detector for CT: Energy response and pulse pileup effects
    Taguchi, Katsuyuki
    Zhang, Mengxi
    Frey, Eric C.
    Wang, Xiaolan
    Iwanczyk, Jan S.
    Nygard, Einar
    Hartsough, Neal E.
    Tsui, Benjamin M. W.
    Barber, William C.
    MEDICAL PHYSICS, 2011, 38 (02) : 1089 - 1102
  • [38] Simultaneous X-ray fluorescence and K-edge CT imaging with photon-counting detectors
    Li, Liang
    Li, Ruizhe
    Zhang, Siyuan
    Chen, Zhiqiang
    DEVELOPMENTS IN X-RAY TOMOGRAPHY X, 2016, 9967
  • [39] Multi-contrast CT imaging using a high energy resolution CdTe detector and a CZT photon-counting detector
    Clements, N.
    Richtsmeier, D.
    Hart, A.
    Bazalova-Carter, M.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (01)
  • [40] Photon-counting CT for simultaneous imaging of multiple contrast agents in the abdomen: An in vivo study
    Symons, Rolf
    Krauss, Bernhard
    Sahbaee, Pooyan
    Cork, Tyler E.
    Lakshmanan, Manu N.
    Bluemke, David A.
    Pourmorteza, Amir
    MEDICAL PHYSICS, 2017, 44 (10) : 5120 - 5127