A model-based algorithm to correct for the loss of backscatter in superficial X-ray radiation therapy

被引:2
|
作者
Harwood, Jeffrey R. [1 ]
Nelli, Flavio E. [1 ]
机构
[1] Univ Hosp Geelong, Andrew Love Canc Ctr, 70 Swanston St, Geelong, Vic 3220, Australia
关键词
Backscatter; Superficial X-ray therapy; Gafchromic film; Monte Carlo; DOSE REDUCTION; MM AL; RADIOTHERAPY; DOSIMETRY;
D O I
10.1016/j.ejmp.2019.08.018
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dosimetry protocols for superficial X-rays prescribe the determination of kerma on the surface of a phantom through the use of a backscatter factor (B-w) that accounts for the effect of phantom scatter. B-w values corresponding to full-scatter phantoms are provided by these protocols. In practice, clinical situations arise wherein there is insufficient scattering material downstream, resulting in published B-w values that overestimate the amount of occurring scatter. To provide an accurate dose calculation the backscatter values need to be corrected for any reduction in scattered radiation. Estimating the change of B-w in situations with incomplete backscatter has previously been achieved by direct measurements or Monte Carlo modelling. For increasing the accuracy of clinical dosimetries, we developed a physical model to deduce an algorithm for calculating backscatter factors in situations with reduced downstream scattering medium. The predictions of the model were validated by comparison with published data, Monte Carlo simulations and film-based measurements for beams with a half-value layer of 0.8, 2 and 4 mm Al. Our algorithm accurately predicts the effect of partial scatter conditions with suitable precision. Its reliability, combined with the simplicity of calculation, makes this methodology suitable to be incorporated into routine clinical dosimetry. The algorithm's underlying physical model provides an intuitive understanding of the effects of field size and beam energy on backscatter reduction, permitting a rational management of this effect.
引用
收藏
页码:157 / 166
页数:10
相关论文
共 50 条
  • [1] Dose reduction from loss of backscatter in superficial x-ray radiation therapy with the Pantak SXT 150 unit
    B. J. Healy
    S. Sylvander
    K. N. Nitschke
    Australasian Physics & Engineering Sciences in Medicine, 2008, 31 : 49 - 55
  • [2] Dose reduction from loss of backscatter in superficial x-ray radiation therapy with the Pantak SXT 150 unit
    Healy, B. J.
    Sylvander, S.
    Nitschke, K. N.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2008, 31 (01) : 49 - 55
  • [3] A NEW TECHNIQUE FOR RADIATION SHIELDING IN SUPERFICIAL X-RAY THERAPY
    BAILY, B
    COE, MA
    HEARNDEN, TM
    BRITISH JOURNAL OF RADIOLOGY, 1981, 54 (645): : 805 - 807
  • [4] Radiochromic film for verification of superficial x-ray backscatter factors
    M. J. Butson
    T. Cheung
    P. K. N. Yu
    Australasian Physics & Engineering Sciences in Medicine, 2007, 30 : 269 - 273
  • [5] Radiochromic film for verification of superficial x-ray backscatter factors
    Butson, M. J.
    Cheung, T.
    Yu, P. K. N.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2007, 30 (04) : 269 - 273
  • [6] SUPERFICIAL X-RAY THERAPY IN CHILDREN
    不详
    BRITISH MEDICAL JOURNAL, 1960, 2 (JUL2): : 84 - 84
  • [7] Hidden Markov model-based extraction of target objects in X-ray image sequence for lung radiation therapy
    Shindo M.
    Ichiji K.
    Homma N.
    Zhang X.
    Okuda S.
    Sugita N.
    Yamaki S.
    Takai Y.
    Yoshizawa M.
    IEEJ Transactions on Electronics, Information and Systems, 2020, 140 (01) : 49 - 60
  • [8] AN EVALUATION OF X-RAY LOSS DUE TO ELECTRON BACKSCATTER
    MYKLEBUST, RL
    JOURNAL DE PHYSIQUE, 1984, 45 (NC-2): : 41 - 42
  • [9] X-RAY THERAPY FOR RADIATION
    GLADSTEIN, AH
    JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1980, 75 (04) : 371 - 371
  • [10] SUPERFICIAL X-RAY THERAPY IN DERMATOLOGY - 1978
    JOLLY, HW
    INTERNATIONAL JOURNAL OF DERMATOLOGY, 1978, 17 (09) : 691 - 697