High-Pitch Electrocardiogram-Triggered Computed Tomography of the Chest Initial Results

被引:67
|
作者
Lell, Michael [1 ]
Hinkmann, Fabian [1 ]
Anders, Katharina [1 ]
Deak, Paul [2 ]
Kalender, Willi A. [2 ]
Uder, Michael [1 ]
Achenbach, Stephan [3 ]
机构
[1] Univ Erlangen Nurnberg, Dept Radiol, D-91054 Erlangen, Germany
[2] Univ Erlangen Nurnberg, Inst Med Phys, D-91054 Erlangen, Germany
[3] Univ Erlangen Nurnberg, Dept Cardiol, D-91054 Erlangen, Germany
关键词
computed tomography; pulmonary embolism; aortic dissection; coronary CT angiography; dual source CT; radiation exposure; CORONARY CT ANGIOGRAPHY; TRIPLE RULE-OUT; DUAL-SOURCE CT; IMAGE QUALITY; NONCARDIAC FINDINGS; RADIATION-EXPOSURE; MULTISLICE CT; HEART-RATES; RECONSTRUCTION; PREVALENCE;
D O I
10.1097/RLI.0b013e3181b9df7e
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: Chest pain is one of the most frequent symptoms in the emergency department. A variety of different diseases, some of them acutely life threatening, can be the underlying cause. Electrocardiogram (ECG)gated computed tomography angiography of the thorax has been proposed as a cost and time effective imaging technique for these patients. We describe a new high-pitch scan mode, which has been developed specifically for low-dose ECG-triggered computed tomography angiography using dual source computed tomography (CT). Material and Methods: Twenty-four patients were examined with this technique on a second generation dual source CT system. The scan mode uses a pitch of 3.2 to acquire a spiral CT data set of the complete thorax in less than I second with a temporal resolution of 75 ms (scan parameters: 128 X 0.6 mm collimation, 0.28 seconds gantry rotation time, 370 mAs at 100 kV [15 patients] and 320 mAs at 120 kV [9 patients], reconstructed slice thickness 0.6 mm, increment 0.4 mm). Data acquisition was prospectively triggered at 50% to 60% of the RR interval to cover the range over the heart in diastole. A triple phase contrast injection protocol (total volume: 80 mL) was used to optimize enhancement of the pulmonary and systemic arterial vessels. Image quality was evaluated using a 4-point scale (1 = absence of motion artifacts; 2 = slight motion artifacts, fully evaluable; 3 = motion artifacts, but evaluable; 4 = unevaluable) on a per-segment basis. Results: The patients had an average heart rate of 68 +/- 15 bpm (range: 43-111 bpm) during data acquisition. Motion artifact free visualization of the aorta and pulmonary vessels was possible in each case, of 344 coronary artery segments, 242 (70%) had an image quality score of 1, 60 segments (17%) a score of 2, 28 segments (8%) a score of 3, and 14 segments (4%) were rated as "unevaluable." In 17 patients (10 patients with a heart rate <= 60 bpm) all segments were evaluable. The average dose length product was 113 +/- 11 mGy x cm per scan (mean effective dose 1.6 +/- 0.2 mSv) at 100 kV and 229 +/- 31 mGy x cm per scan (mean effective dose 3.2 +/- 0.4 mSv) at 120 kV. Conclusion: Our initial results indicate that this high-pitch scan mode allows motion artifact free and accurate visualization of the thoracic vessels, and diagnostic image quality of the coronary arteries in patients with low and stable heart rates at a very low radiation exposure.
引用
收藏
页码:728 / 733
页数:6
相关论文
共 50 条
  • [21] Diagnostic Value of Prospective Electrocardiogram-triggered Dual-source Computed Tomography Angiography for Infants and Children with Interrupted Aortic Arch
    Li, Hai-Ou
    Wang, Xi-Ming
    Nie, Pei
    Ji, Xiao-Peng
    Cheng, Zhao-Ping
    Chen, Jiu-Hong
    Xu, Zhuo-Dong
    CHINESE MEDICAL JOURNAL, 2015, 128 (09) : 1184 - 1189
  • [22] High-pitch prospective ECG-triggered helical coronary computed tomography angiography in clinical practice: image quality and radiation dose
    Bogaard, Kjell
    van der Zant, Friso M.
    Knol, Remco J. J.
    Reinders, Sjoerd
    Krul, Marije M. G.
    van Rossum, Albert C.
    Knaapen, Paul
    Cornel, Jan H.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2015, 31 (01): : 125 - 133
  • [23] Prospectively ECG-triggered High-pitch Spiral Acquisition for Cardiac CT Angiography in Routine Clinical Practice Initial Results
    Kroepil, Patric
    Rojas, Carlos A.
    Ghoshhajra, Brian
    Lanzman, Rotem S.
    Miese, Falk R.
    Scherer, Axel
    Kalra, Mannudeep
    Abbara, Suhny
    JOURNAL OF THORACIC IMAGING, 2012, 27 (03) : 194 - 201
  • [24] Diagnostic Value of Prospective Electrocardiogram-triggered Dual-source Computed Tomography Angiography for Infants and Children with Interrupted Aortic Arch
    Li Hai-Ou
    Wang Xi-Ming
    Nie Pei
    Ji Xiao-Peng
    Cheng Zhao-Ping
    Chen Jiu-Hong
    Xu Zhuo-Dong
    中华医学杂志英文版, 2015, 128 (09) : 1184 - 1189
  • [25] Development of A deep Learning-based algorithm for High-Pitch helical computed tomography imaging
    Duan, Xiaoman
    Ding, Xiao Fan
    Khoz, Samira
    Chen, Xiongbiao
    Zhu, Ning
    EXPERT SYSTEMS WITH APPLICATIONS, 2025, 262
  • [26] High-Pitch Coronary Computed Tomographic Angiography Using the Third-Generation Dual-Source Computed Tomography: Initial Experience in Patients With High Heart Rate
    Feng, Ruiqi
    Mao, Jingsong
    Liu, Xiaofei
    Zhao, Yu
    Tong, Jiajie
    Zhang, Liang
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (02) : 248 - 255
  • [27] High-pitch Dual-source Computed Tomography Pulmonary Angiography in Freely Breathing Patients
    Bauer, Ralf W.
    Schell, Boris
    Beeres, Martin
    Wichmann, Julian L.
    Bodelle, Boris
    Vogl, Thomas J.
    Kerl, Josef Matthias
    JOURNAL OF THORACIC IMAGING, 2012, 27 (06) : 376 - 381
  • [28] Measuring coronary artery calcium with high-pitch computed tomography acquisition: is it as consistent as prospective acquisition
    Truong, L-L Lan-Linh
    Erthal, F.
    Clarkin, O.
    Hossain, A.
    Yam, Y.
    Inacio, J.
    Chow, B. J. W.
    EUROPEAN JOURNAL OF HEART FAILURE, 2018, 20 : 158 - 158
  • [29] Full Field Image Reconstruction Is Suitable for High-Pitch Dual-Source Computed Tomography
    Mahnken, Andreas H.
    Allmendinger, Thomas
    Sedlmair, Martin
    Tamm, Miriam
    Reinartz, Sebastian D.
    Flohr, Thomas
    INVESTIGATIVE RADIOLOGY, 2012, 47 (11) : 642 - 648
  • [30] Radiation dose considerations by intra-individual Monte Carlo simulations in dual source spiral coronary computed tomography angiography with electrocardiogram-triggered tube current modulation and adaptive pitch
    May, Matthias S.
    Deak, Paul
    Kuettner, Axel
    Lell, Michael M.
    Wuest, Wolfgang
    Scharf, Michael
    Keller, Andrea K.
    Haeberle, Lothar
    Achenbach, Stephan
    Seltmann, Martin
    Uder, Michael
    Kalender, Willi A.
    EUROPEAN RADIOLOGY, 2012, 22 (03) : 569 - 578