Technical accuracy of ten self-monitoring blood glucose devices commonly used in Dhaka City of Bangladesh

被引:1
|
作者
Nayeem, Jannatul [1 ]
Kamaluddin, S. M. [1 ]
Chowdhury, Hasina Akhter [2 ]
Ali, Liaquat [1 ,3 ]
机构
[1] BUHS, Dept Appl Lab Sci, Dhaka, Bangladesh
[2] BUHS, Dept Biostat, Dhaka, Bangladesh
[3] BUHS, Dept Biochem & Cell Biol, 125-1 Darussalam,Mirpur 1, Dhaka 1216, Bangladesh
关键词
Self-monitoring blood glucose devices; Glucometer; Glycemic control tools; Technical accuracy;
D O I
10.1007/s13410-019-00712-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Due to inadequate regulatory mechanisms, the accuracy of the self-monitoring blood glucose (SMBG) devices is not ensured leading to potentially serious clinical consequences in Bangladesh. The present study was undertaken to evaluate the technical accuracy of ten most commonly used SMBG devices marketed in Dhaka City of Bangladesh. Top ten SMBG devices sold in Dhaka City were studied on a group of 100 type 2 diabetes mellitus subjects. Blood glucose values estimated (at fasting and 2 h after breakfast) by SMBG devices, using a blinding technique, were compared with the corresponding laboratory values by enzymatic method using the Dimension RXLMax automated chemistry analyzer. Hematocrit was measured using the Sysmex XT 2000 hematology autoanalyzer. The mean absolute relative error (MARE, %) was used as an indicator of accuracy and precision together. A highly significant correlation was observed between the device and laboratory values. However, none of the devices showed an acceptable accuracy at 5-10% deviation from the corresponding laboratory values either at fasting or postprandial states. On pooling together data from two prandial states, even at 15% deviation limits, 70% of the devices failed to show accurate results. On calculation of MARE, 60% devices were found to be beyond 15% error limit at 95% accuracy level. Corresponding analysis with 90% accuracy level showed 30% beyond the limit. Blood glucose results from around 30% of the top ten SMBG devices sold in Dhaka City do not have even minimum level of technical accuracy, and many others are not optimum in their accuracy levels.
引用
收藏
页码:579 / 584
页数:6
相关论文
共 50 条
  • [21] Self-monitoring of blood glucose
    Garg, S.
    Hirsch, I. B.
    Bode, B. W.
    Battelino, T.
    Pickup, J.
    Dassau, E.
    Atlas, E.
    Phillip, M.
    Danne, T.
    Bolinder, J.
    Heinemann, L.
    Kaufman, N.
    Hod, M.
    Jovanovic, L.
    Spector, D.
    Shikora, S.
    Skyler, J. S.
    Zisser, H.
    Riddell, M. C.
    INTERNATIONAL JOURNAL OF CLINICAL PRACTICE, 2010, 64 : 1 - 69
  • [22] Self-monitoring of blood glucose
    不详
    DIABETES CARE, 1996, 19 : S62 - S66
  • [23] EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE
    CLARKE, WL
    COX, D
    GONDERFREDERICK, LA
    CARTER, W
    POHL, SL
    DIABETES CARE, 1987, 10 (05) : 622 - 628
  • [24] BEHAVIORAL SKILLS IN SELF-MONITORING OF BLOOD-GLUCOSE - RELATIONSHIP TO ACCURACY
    WING, RR
    KOESKE, R
    NEW, A
    LAMPARSKI, D
    BECKER, D
    DIABETES CARE, 1986, 9 (04) : 330 - 333
  • [25] Accuracy, reliability and human factors associated with self-monitoring of blood glucose
    Davis, E
    Latman, NS
    FASEB JOURNAL, 2005, 19 (05): : A1530 - A1531
  • [26] 10/10 Accuracy Analysis of a Self-Monitoring Blood Glucose System
    Katz, Laurence B.
    Smith, Antony
    Cameron, Hilary
    Setford, Steven
    Grady, Mike
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2021, 15 (03): : 717 - 718
  • [27] Accuracy and Precision Evaluation of Seven Self-Monitoring Blood Glucose Systems
    Kuo, Chih-Yi
    Hsu, Cheng-Teng
    Ho, Cheng-Shiao
    Su, Ting-En
    Wu, Ming-Hsun
    Wang, Chau-Jong
    DIABETES TECHNOLOGY & THERAPEUTICS, 2011, 13 (05) : 596 - 600
  • [28] Evaluation of five devices for self-monitoring of blood glucose in the normoglycaemic range
    Glasmacher, AG
    Brennemann, W
    Hahn, C
    Marko, C
    Wehnert, S
    Löhr, U
    Kutschkow, R
    EXPERIMENTAL AND CLINICAL ENDOCRINOLOGY & DIABETES, 1998, 106 (04) : 360 - 364
  • [29] SELF-MONITORING OF BLOOD-GLUCOSE
    不详
    DIABETES CARE, 1991, 14 : 57 - 62
  • [30] BLOOD-GLUCOSE SELF-MONITORING
    KAVELMAN, DA
    DIABETES CARE, 1979, 2 (06) : 531 - 532