Investigation of vibration effect on dynamic calibration of pressure sensors based on shock tube system

被引:21
|
作者
Diao, Kuan [1 ]
Yao, Zhenjian [1 ]
Wang, Zhongyu [2 ]
Liu, Xiaojun [1 ]
Wang, Chenchen [3 ]
Shang, Zhendong [4 ]
机构
[1] Huazhong Univ Sci & Technol, Sch Mech Sci & Engn, State Key Lab Digital Mfg Equipment & Technol, Wuhan 430074, Hubei, Peoples R China
[2] Beihang Univ, Sch Instrument Sci & Optoelect Engn, Beijing 100191, Peoples R China
[3] Beijing Changcheng Inst Metrol & Measurement, Beijing 100095, Peoples R China
[4] Henan Univ Sci & Technol, Sch Mechatron Engn, Luoyang 471003, Peoples R China
关键词
Pressure sensor; Vibration; Dynamic characteristic; Calibration parameters; Shock tube; MEASUREMENT UNCERTAINTY; WAVES;
D O I
10.1016/j.measurement.2019.107015
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper devotes to investigate the influence of vibration on dynamic calibration of pressure sensors based on shock tube system. Two sensors with different types are separately calibrated under the same step pressure to distinguish the individual sensor's dependent effect. To excavate how the sensor's dynamic characteristics fluctuate under different vibration cases, support mounts and buffer rings of different materials are used in the experimental shock tube system. Vibrations from two main sources are expounded in detail and detected by an acceleration sensor installed near the pressure sensors. Based on the outputs of sensors, the influences of vibration on the repeatability, stability and dynamic characteristics of pressure sensors are analyzed qualitatively and quantitatively. Furthermore, a weighted scoring method based on the relative standard deviation of the calibration parameters is proposed to comprehensively assess the vibration effect on dynamic calibration results of pressure sensors. The experimental results show that the influence of vibration from the rupture of diaphragm on dynamic calibration can be avoided suitably by using the mount with high damping material or a shock tube with long driven section. The vibration from the instantaneous impact of shock wave has slight influence on the repeatability, rise time and resonant frequency, and significant influence on the stability, settling time, overshoot and uncertainty in working band of pressure sensors. Based on the weighted scores under different vibration cases, the combination of steel mount and aluminum ring produces the most reliable calibration result. These findings are also verified by another group of calibration experiments with a smaller amplitude of step pressure. (C) 2019 Elsevier Ltd. All rights reserved.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Novel secondary standard for calibration of dynamic pressure sensors
    Sander, Christian
    Aspiala, Markus
    Hogstrom, Richard
    TM-TECHNISCHES MESSEN, 2022, 89 (10) : 721 - 728
  • [32] Dynamic calibration of low range silicon pressure sensors
    Zakrzewski, J
    Wróbel, K
    IMTC/2001: PROCEEDINGS OF THE 18TH IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-3: REDISCOVERING MEASUREMENT IN THE AGE OF INFORMATICS, 2001, : 493 - 496
  • [33] Analyzing Frequency Response of Differential Pressure Sensors by Dynamic Pressure Calibration
    Kober, Timo
    Sindlinger, Stefan
    Werthschuetzky, Roland
    TM-TECHNISCHES MESSEN, 2010, 77 (02) : 83 - 88
  • [34] Dynamic Pressure Calibration of Pressure Sensors Using Liquid Step Pressure Generator
    Wang, Shenghung
    Han, Leelong
    Tsung, Tsing Tshih
    MEASUREMENT TECHNOLOGY AND INTELLIGENT INSTRUMENTS IX, 2010, 437 : 8 - 12
  • [35] Dynamic modeling of liquid impulse pressure generator for calibration of pressure sensors
    Yang, Jun
    Fan, Shangchun
    Li, Bo
    Huang, Riheng
    Shi, Yusong
    Shi, Bo
    SENSORS AND ACTUATORS A-PHYSICAL, 2018, 279 : 120 - 131
  • [36] An improved reconstruction method of the reflected dynamic pressure in shock tube system based on inverse sensing model identification
    Yao, Zhenjian
    Li, Yongsheng
    Shi, Bo
    Yao, Lei
    Song, Jinlin
    AEROSPACE SCIENCE AND TECHNOLOGY, 2024, 145
  • [37] Test and calibration of Bragg grating based vibration sensors
    Andresen, Soren
    Licht, Torben R.
    Nielsen, Finn K.
    22ND INTERNATIONAL CONFERENCE ON OPTICAL FIBER SENSORS, PTS 1-3, 2012, 8421
  • [38] Application of shock tube in calibrating dynamic pressure transducers
    Yan, Laijun
    Chen, Yong
    Zhang, Lihu
    Zhang, Xu
    Yao, Xianghong
    Qi, Xinhua
    Zhang, Long
    2018 2ND INTERNATIONAL CONFERENCE ON FLUID MECHANICS AND INDUSTRIAL APPLICATIONS, 2018, 1064
  • [39] Characterization of a fiber-optic pressure sensor in a shock tube system for dynamic calibrations
    Zelan, Martin
    Arrhen, Fredrik
    Jarlemark, Per
    Mollmyr, Oscar
    Johansson, Hakan
    METROLOGIA, 2015, 52 (01) : 48 - 53
  • [40] Dynamic calibration of low-range silicon pressure sensors
    Zakrzewski, J
    Wróbel, K
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2002, 51 (06) : 1358 - 1362