Investigation of vibration effect on dynamic calibration of pressure sensors based on shock tube system

被引:21
|
作者
Diao, Kuan [1 ]
Yao, Zhenjian [1 ]
Wang, Zhongyu [2 ]
Liu, Xiaojun [1 ]
Wang, Chenchen [3 ]
Shang, Zhendong [4 ]
机构
[1] Huazhong Univ Sci & Technol, Sch Mech Sci & Engn, State Key Lab Digital Mfg Equipment & Technol, Wuhan 430074, Hubei, Peoples R China
[2] Beihang Univ, Sch Instrument Sci & Optoelect Engn, Beijing 100191, Peoples R China
[3] Beijing Changcheng Inst Metrol & Measurement, Beijing 100095, Peoples R China
[4] Henan Univ Sci & Technol, Sch Mechatron Engn, Luoyang 471003, Peoples R China
关键词
Pressure sensor; Vibration; Dynamic characteristic; Calibration parameters; Shock tube; MEASUREMENT UNCERTAINTY; WAVES;
D O I
10.1016/j.measurement.2019.107015
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper devotes to investigate the influence of vibration on dynamic calibration of pressure sensors based on shock tube system. Two sensors with different types are separately calibrated under the same step pressure to distinguish the individual sensor's dependent effect. To excavate how the sensor's dynamic characteristics fluctuate under different vibration cases, support mounts and buffer rings of different materials are used in the experimental shock tube system. Vibrations from two main sources are expounded in detail and detected by an acceleration sensor installed near the pressure sensors. Based on the outputs of sensors, the influences of vibration on the repeatability, stability and dynamic characteristics of pressure sensors are analyzed qualitatively and quantitatively. Furthermore, a weighted scoring method based on the relative standard deviation of the calibration parameters is proposed to comprehensively assess the vibration effect on dynamic calibration results of pressure sensors. The experimental results show that the influence of vibration from the rupture of diaphragm on dynamic calibration can be avoided suitably by using the mount with high damping material or a shock tube with long driven section. The vibration from the instantaneous impact of shock wave has slight influence on the repeatability, rise time and resonant frequency, and significant influence on the stability, settling time, overshoot and uncertainty in working band of pressure sensors. Based on the weighted scores under different vibration cases, the combination of steel mount and aluminum ring produces the most reliable calibration result. These findings are also verified by another group of calibration experiments with a smaller amplitude of step pressure. (C) 2019 Elsevier Ltd. All rights reserved.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Towards a shock tube method for the dynamic calibration of pressure sensors
    Downes, Stephen
    Knott, Andy
    Robinson, Ian
    PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2014, 372 (2023):
  • [2] Dynamic calibration of pressure transducers with an improved shock tube system
    Wisniewiski, David
    SENSORS AND SMART STRUCTURES TECHNOLOGIES FOR CIVIL, MECHANICAL, AND AEROSPACE SYSTEMS 2013, 2013, 8692
  • [3] Improved shock tube method for dynamic calibration of the sensitivity characteristic of piezoresistive pressure sensors
    Yao, Zhenjian
    Li, Yongsheng
    Ding, Yifan
    Wang, Chenchen
    Yao, Lei
    Song, Jinlin
    MEASUREMENT, 2022, 196
  • [4] Water shock tube for high pressure dynamic calibration
    Ming-Wu, Zhu
    Xiao-Jian, Wen
    Conference Record - IEEE Instrumentation and Measurement Technology Conference, 1994, 2 : 468 - 470
  • [5] Dynamic Calibration Method of Blast Pressure Pencil Probes Based on Adjustable Shock Tube
    Shi, Yucheng
    Kong, Deren
    Ma, Xuejiao
    Zhang, Chunyan
    IEEE SENSORS JOURNAL, 2023, 23 (11) : 11704 - 11712
  • [6] Shock tube-based calibration installation for dynamic pressure transducers and performance testing
    Yan, Laijun
    Zhang, Xu
    Zhang, Lihu
    Yao, Xianghong
    Qi, Xinhua
    Chen, Yong
    Zhang, Long
    JOURNAL OF ENGINEERING-JOE, 2019, 2019 (23): : 8577 - 8582
  • [7] The study on water shock tube for high pressure dynamic calibration ISTM conference
    Gong, CK
    Li, YX
    ISTM/2005: 6th International Symposium on Test and Measurement, Vols 1-9, Conference Proceedings, 2005, : 4589 - 4592
  • [8] A method for correcting the high-frequency mechanical vibration effects in the dynamic calibration of pressure measurement systems using a shock tube
    Svete, A.
    Amer, E.
    Jonsson, G.
    Kutin, J.
    Arrhen, F.
    MECHANICAL SYSTEMS AND SIGNAL PROCESSING, 2023, 193
  • [9] SHOCK-TUBE FOR CALIBRATION OF PRESSURE DETECTORS
    MAKAREVICH, GA
    LISENKOVA, GS
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1980, 23 (03) : 811 - 812
  • [10] Evaluation of Shock Tube Retrofitted with Fast-Opening Valve for Dynamic Pressure Calibration
    Amer, Eynas
    Wozniak, Mikolaj
    Jonsson, Gustav
    Arrhen, Fredrik
    SENSORS, 2021, 21 (13)