Long-term Results of Stent-graft Placement to Treat Central Venous Stenosis and Occlusion in Hemodialysis Patients with Arteriovenous Fistulas

被引:79
|
作者
Jones, Robert G. [1 ]
Willis, Andrew P. [1 ]
Jones, Catherine [1 ]
McCafferty, Ian J. [1 ]
Riley, Peter L. [1 ]
机构
[1] Univ Hosp Birmingham NHS Trust, Queen Elizabeth Hosp Birmingham, Dept Radiol, Birmingham B15 2TH, W Midlands, England
关键词
ANGIOPLASTY; MANAGEMENT;
D O I
10.1016/j.jvir.2011.06.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To determine the effectiveness of stent-grafts for the treatment of central venous disease in hemodialysis patients with functioning arteriovenous (AV) fistulas. Materials and Methods: Between October 2004 and March 2010, 42 VIABAHN stent-grafts were deployed in central veins of 30 patients (16 men, 14 women; mean age 60 y) with functioning AV fistulas and central venous disease that did not respond to percutaneous transluminal angioplasty (PTA). Eighteen patients had central vein stenosis and 12 had occlusion. Previous PTA and/or bare metal stent placement had been performed in 23 patients (77%). Surveillance was carried out at 3, 6, 9, 12, 18, and 24 months with diagnostic fistulography. The mean follow-up was 705 days (range, 66-1,645 d). Statistical analysis included Kaplan-Meier and log-rank studies. Results: Technical success rate was 100%. Primary patency rates were 97%, 81%, 67%, and 45% at 3, 6, 12, and 24 months, respectively. Primary assisted patency rates were 100%, 100%, 80%, and 75% at 3, 6, 12, and 24 months, respectively. Patients without previous procedures had significantly shorter times to repeat intervention (P = .018) than those who had undergone PTA or bare metal stent placement previously. Patients with occlusive lesions had a significantly shorter primary patency interval (P = .05) than patients with stenoses. Occluded veins were more likely to require further stent-grafts (P = .02). Twelve patients required further stent-grafts to maintain patency. There was one minor complication. Conclusions: Stent-graft placement to treat central venous disease in hemodialysis patients with autogenous AV fistulas is safe and effective if PTA fails to maintain luminal patency.
引用
收藏
页码:1240 / 1245
页数:6
相关论文
共 50 条
  • [41] Primary endovascular stent placement for focal infrarenal aortic stenosis: Long-term results
    Stoeckelhuber, B. M.
    Stoeckelhuber, M.
    Gellissen, J.
    Helmberger, T.
    Kueffer, G.
    ATHEROSCLEROSIS SUPPLEMENTS, 2007, 8 (01) : 101 - 101
  • [42] Treatment of hemodialysis-related central venous stenosis or occlusion: Results of primary Wallstent placement and follow-up in 50 patients
    Haage, P
    Vorwerk, D
    Piroth, W
    Schuermann, K
    Guenther, RW
    RADIOLOGY, 1999, 212 (01) : 175 - 180
  • [43] Primary endovascular stent placement for focal infrarenal aortic stenosis: Long-term results
    Stoeckelhuber, Beate Maria
    Stoeckelhuber, Mechthild
    Gellissen, Joerg
    Kueffer, Georg
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (07) : 1105 - 1109
  • [44] Endovascular treatment of central thoracic venous stenosis and occlusion in hemodialysis patients
    Birgi, Erdem
    Ergun, Onur
    Durmaz, Hasanali
    Hekimoglu, Azad
    Conkbayir, Isik
    Hekimoglu, Baki
    ANNALS OF CLINICAL AND ANALYTICAL MEDICINE, 2021, 12 (12): : 1406 - 1410
  • [45] Stent Graft Placement for Treatment of Prosthetic Graft Outflow Stenosis, a Viable Long-term Solution: A Case Report
    Pawar, Pranay
    Sharma, Achintya
    Ayyappan, M. K.
    Mathur, Kapil
    Jagan, Jithin
    Raju, Radhakrishnan
    INDIAN JOURNAL OF NEPHROLOGY, 2019, 29 (03) : 211 - 213
  • [46] Percutaneous transluminal angioplasty for central venous stenosis or occlusion in hemodialysis patients
    Horita, Yuki
    JOURNAL OF VASCULAR ACCESS, 2019, 20 : 87 - 92
  • [47] Endovascular stent-graft applications in abdominal aortic aneurysms: mid- and long-term results
    Yavuz, Sadan
    Ozbudak, Ersan
    Gumustas, Sevtap
    Kanko, Muhip
    Ciftci, Ercument
    Berki, Turan
    TURK GOGUS KALP DAMAR CERRAHISI DERGISI-TURKISH JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 21 (02): : 333 - 340
  • [48] Endovascular Stent-Graft Repair of Type B Aortic Dissection: Intermediate to Long-Term Results
    Kim, Ung
    Hong, Seong-Jin
    Kim, Jung-Sun
    Ko, Young-Guk
    Choi, Dong-hoon
    Jang, Yang-Soo
    Shim, Won-Heum
    Cho, Seung-Yun
    CIRCULATION, 2008, 118 (18) : S1078 - S1078
  • [49] Long-term results of infrainguinal venous bypass surgery: risk factors for stenosis and occlusion
    Prevost, G. A.
    Heller, G.
    Odermatt, M.
    Furrer, M.
    BRITISH JOURNAL OF SURGERY, 2010, 97 : 26 - 26
  • [50] Long-term Results of Interwoven Nitinol Stents to Treat the Radiocephalic Anastomotic Arteriovenous Fistula Stenosis
    Thomas, Shannon D.
    Peden, Samantha
    Katib, Nedal
    Crowe, Phillip
    Barber, Tracie
    Varcoe, Ramon L.
    JOURNAL OF ENDOVASCULAR THERAPY, 2023, 30 (02) : 176 - 184