The MiniACcor:: Constructive redesign of an implantable total artificial heart, initial laboratory testing and further steps

被引:7
|
作者
Kwant, P. B.
Finocchiaro, T.
Foerster, F.
Reul, H.
Rau, G.
Morshuis, M.
El Banayosi, A.
Koerfer, R.
Schmitz-Rode, T.
Steinseifer, U.
机构
[1] Helmholtz Eye Dis Res Inst, Chair Appl Med Engn, D-52074 Aachen, Germany
[2] Heart & Diabet Ctr N Rhine Westphalia, Bad Oeynhausen, Germany
来源
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS | 2007年 / 30卷 / 04期
关键词
ACcor; MiniACcor; total artificial heart; redesign; laboratory testing; compliance chamber;
D O I
10.1177/039139880703000411
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Aachen Total Artificial Heart (ACcor) has been under development at the Helmholtz Institute in Aachen over the last decade. It may serve as a bridge to transplant or as a long-term replacement of the natural heart. Based upon previous in vivo experiments with the ACcor total artificial heart, it was decided to optimize and redesign the pump unit. Smaller dimensions, passive filling and separability into three components were the three main design goals. The new design is called the MiniACcor, which is about 20% smaller than its predecessor, and weighs only 470 grams. Also its external driver/control unit was miniaturized and a new microcontroller was selected. To validate the design, it was extensively tested in laboratory mock loops. The MiniACcor was able to pump between 4.5 and 7 1/min at different pump rates against normal physiological pressures. Several requirements for the future compliance chamber and transcutaneous energy transmission (TET) system were also measured in the same mock loop. Further optimization and validation are being performed in cooperation with the Heart and Diabetes Centre North Rhine-Westphalia.
引用
收藏
页码:345 / 351
页数:7
相关论文
共 50 条
  • [21] Cycle testing of the MagScrew total artificial heart external battery pack
    Casas, Fernando
    Weber, Stephan
    Klatte, Ryan
    Luangphakdy, Viviane
    Smith, William A.
    ARTIFICIAL ORGANS, 2007, 31 (09) : 698 - 702
  • [22] Initial Fitting Study of a Pediatric Continuous-Flow Total Artificial Heart
    Miyagi, C.
    Karimov, J. H.
    Kado, Y.
    Ray, D.
    Polakowski, T.
    Ahmad, M.
    Karamlow, T.
    Najm, H. K.
    Fukamachi, K.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2021, 40 (04): : S173 - S174
  • [23] Design Aspects of Implantable Mechatronic Units for Systems of Auxiliary Blood Circulation and Total Artificial Heart
    Zhdanov, A.
    Morozov, V.
    INTERNATIONAL CONFERENCE ON AUTOMATION, MECHANICAL AND ELECTRICAL ENGINEERING (AMEE 2015), 2015, : 270 - 276
  • [24] Three-dimensional thoracic modeling for an anatomical compatibility study of the implantable total artificial heart
    Zhang, B
    Masuzawa, T
    Tatsumi, E
    Taenaka, Y
    Uyama, C
    Takano, H
    Takamiya, M
    ARTIFICIAL ORGANS, 1999, 23 (03) : 229 - 234
  • [26] Implantable control, telemetry, and solar energy system in the moving actuator type total artificial heart
    Ahn, JM
    Lee, JH
    Choi, SW
    Kim, WE
    Omn, KS
    Park, SK
    Kim, WG
    Roh, JR
    Min, BG
    ARTIFICIAL ORGANS, 1998, 22 (03) : 250 - 259
  • [27] Ultracompact, completely implantable permanent use electromechanical ventricular assist device and total artificial heart
    Honda, N
    Inamoto, T
    Nogawa, M
    Takatani, S
    ARTIFICIAL ORGANS, 1999, 23 (03) : 253 - 261
  • [28] Eddy-Current Linear-Rotary Position Sensor for an Implantable Total Artificial Heart
    Giuffrida, Rosario, V
    Kolar, Johann W.
    Bortis, Dominik
    2022 25TH INTERNATIONAL CONFERENCE ON ELECTRICAL MACHINES AND SYSTEMS (ICEMS 2022), 2022,
  • [29] Demikhov's "Mechanical Heart": The Circumstances Surrounding Creation of the World's First Implantable Total Artificial Heart in 1937
    Glyantsev, Sergey P.
    Tchantchaleishvili, Vakhtang
    Bockeria, Leo A.
    ASAIO JOURNAL, 2016, 62 (01) : 106 - 109
  • [30] Design and evaluation of a hybrid mock circulatory loop for total artificial heart testing
    Cuenca-Navalon, Elena
    Finocchiaro, Thomas
    Laumen, Marco
    Fritschi, Andreas
    Schmitz-Rode, Thomas
    Steinseifer, Ulrich
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2014, 37 (01): : 71 - 80