The MiniACcor:: Constructive redesign of an implantable total artificial heart, initial laboratory testing and further steps

被引:7
|
作者
Kwant, P. B.
Finocchiaro, T.
Foerster, F.
Reul, H.
Rau, G.
Morshuis, M.
El Banayosi, A.
Koerfer, R.
Schmitz-Rode, T.
Steinseifer, U.
机构
[1] Helmholtz Eye Dis Res Inst, Chair Appl Med Engn, D-52074 Aachen, Germany
[2] Heart & Diabet Ctr N Rhine Westphalia, Bad Oeynhausen, Germany
来源
关键词
ACcor; MiniACcor; total artificial heart; redesign; laboratory testing; compliance chamber;
D O I
10.1177/039139880703000411
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Aachen Total Artificial Heart (ACcor) has been under development at the Helmholtz Institute in Aachen over the last decade. It may serve as a bridge to transplant or as a long-term replacement of the natural heart. Based upon previous in vivo experiments with the ACcor total artificial heart, it was decided to optimize and redesign the pump unit. Smaller dimensions, passive filling and separability into three components were the three main design goals. The new design is called the MiniACcor, which is about 20% smaller than its predecessor, and weighs only 470 grams. Also its external driver/control unit was miniaturized and a new microcontroller was selected. To validate the design, it was extensively tested in laboratory mock loops. The MiniACcor was able to pump between 4.5 and 7 1/min at different pump rates against normal physiological pressures. Several requirements for the future compliance chamber and transcutaneous energy transmission (TET) system were also measured in the same mock loop. Further optimization and validation are being performed in cooperation with the Heart and Diabetes Centre North Rhine-Westphalia.
引用
收藏
页码:345 / 351
页数:7
相关论文
共 50 条
  • [1] Durability testing of a completely implantable electric total artificial heart
    Zapanta, CM
    Snyder, AJ
    Weiss, WJ
    Cleary, TJ
    Reibson, JD
    Rawhouser, MA
    Lewis, JP
    Pierce, WS
    Rosenberg, G
    ASAIO JOURNAL, 2005, 51 (03) : 214 - 223
  • [2] A NEW TOTAL IMPLANTABLE ARTIFICIAL-HEART
    MARTENS, J
    ARTIFICIAL ORGANS, 1990, 14 (04) : 284 - 289
  • [3] Initial In Vitro Testing of a Pediatric Continuous-Flow Total Artificial Heart
    Fukamachi, K.
    Horvath, D. J.
    Karimov, J. H.
    Sunagawa, G.
    Byram, N.
    Kuban, B. D.
    Gao, S.
    Dessoffy, R.
    Moazami, N.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2017, 36 (04): : S164 - S164
  • [4] Initial in vitro testing of a paediatric continuous-flow total artificial heart
    Fukamachi, Kiyotaka
    Karimov, Jamshid H.
    Horvath, David J.
    Sunagawa, Gengo
    Byram, Nicole A.
    Kuban, Barry D.
    Moazami, Nader
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2018, 26 (06) : 897 - 901
  • [5] The total artificial heart and implantable biventricular assist device
    Min B.G.
    Sun K.
    Journal of Artificial Organs, 2002, 5 (3) : 147 - 148
  • [6] Totally Implantable total artificial heart for clinical application
    Nosé, Y
    ARTIFICIAL ORGANS, 2002, 26 (03) : 214 - 215
  • [7] Prototype Development of an Implantable Compliance Chamber for a Total Artificial Heart
    Schmitz, Stephanie
    Unthan, Kristin
    Sedlaczek, Marc
    Wald, Felix
    Finocchiaro, Thomas
    Spiliopoulos, Sotirios
    Koerfer, Reiner
    Steinseifer, Ulrich
    ARTIFICIAL ORGANS, 2017, 41 (02) : 122 - 129
  • [8] Initial In Vivo Testing of a Self-Balancing Pulsatile Continuous Flow Total Artificial Heart
    Golding, L. A. R.
    Fukamachi, K.
    Horvath, D. J.
    Massiello, A. L.
    Fumoto, H.
    Arakawa, Y.
    Catanese, J.
    Chen, J. -F.
    Mielke, N.
    Horai, T.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2009, 28 (02): : S314 - S315
  • [9] Implantable total artificial heart: History and present status at the University of Tokyo
    Imachi K.
    Chinzei T.
    Abe Y.
    Isoyama T.
    Mochizuki S.
    Saito I.
    Ono T.
    Karita T.
    Matsuura H.
    Kouno A.
    Mabuchi K.
    Fujimasa I.
    Atsumi K.
    Journal of Artificial Organs, 1999, 2 (1) : 13 - 23
  • [10] PROGRESS IN THE DEVELOPMENT OF A COMPLETELY IMPLANTABLE TOTAL ARTIFICIAL-HEART (TAH)
    JACOBS, G
    YOZU, R
    STACY, G
    SMITH, W
    WATANABE, T
    SHIMOMITSU, T
    NOSE, Y
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1985, 32 (10) : 892 - 892