Measurement of nasal nitric oxide by hand-held and stationary devices

被引:29
|
作者
Montella, Silvia [1 ]
Alving, Kjell [2 ]
Maniscalco, Mauro [3 ]
Sofia, Matteo [4 ]
De Stefano, Sara [1 ]
Raia, Valeria [1 ]
Santamaria, Francesca [1 ]
机构
[1] Univ Naples Federico II, Dept Pediat, I-80131 Naples, Italy
[2] Uppsala Univ, Dept Womens & Childrens Hlth, Uppsala, Sweden
[3] S Maria della Pieta Hosp, Sect Resp Med, Naples, Italy
[4] Univ Naples Federico II, Dept Resp Med, I-80131 Naples, Italy
关键词
Chemiluminescence; cystic fibrosis; electrochemical sensor; nasal nitric oxide; primary ciliary dyskinesia; PRIMARY CILIARY DYSKINESIA; CHILDREN;
D O I
10.1111/j.1365-2362.2011.02501.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Nasal nitric oxide (nNO) is assessed by nasal aspiration/insufflation via one nostril or by nasal silent exhalation through a facemask and is also measured during humming, a manoeuvre that results in increased nNO in the presence of a patent osteomeatal complex. Humming nNO peak is absent in primary ciliary dyskinesia (PCD) and in cystic fibrosis (CF). Hand-held devices are used successfully for exhaled or nNO analysis. No study compared nNO during silent and humming exhalation using hand-held and stationary analysers. Methods Thirty-eight subjects (14 PCD; 11 CF; 13 healthy individuals) measured nNO with a stationary and a hand-held analyser during silent and humming exhalations. Results No difference between nNO obtained from stationary or hand-held analyser during silent and humming exhalation was found (P > 0 05). Patients with PCD exhibited lower silent and humming nNO than CF or controls (P < 0 001). During both silent and humming exhalation, there was a significant correlation between nNO from the two analyzers both in the whole study population and within each group (r >= 0.7, P < 0 01). Bland-Altman plots confirmed this agreement. Using the hand-held device during humming, nNO values of 50, 81 and 21 ppb had sensitivity and specificity > 90% for discriminating PCD or CF from healthy subjects, and patients with PCD from patients with CF, respectively. Conclusions The hand-held device is as effective as the stationary analyzer for assessing nNO during silent and humming exhalation. Its wider use might result in an increased number of subjects suspected to have PCD.
引用
收藏
页码:1063 / 1070
页数:8
相关论文
共 50 条
  • [1] Usability of a hand-held device to determine nasal nitric oxide in clinical routine
    Jung, Andreas
    Bigger, Michael
    Bieli, Christian
    Moller, Alexander
    EUROPEAN RESPIRATORY JOURNAL, 2015, 46
  • [2] Nasal fractional exhaled nitric oxide analysis with a novel hand-held device
    Weschta, Michael
    Deutschle, Tom
    Riechelmann, Herbert
    RHINOLOGY, 2008, 46 (01) : 23 - 27
  • [3] Reliability of a new hand-held device for the measurement of exhaled nitric oxide
    Khalili, B.
    Boggs, P. B.
    Bahna, S. L.
    ALLERGY, 2007, 62 (10) : 1171 - 1174
  • [4] Validation study of nasal nitric oxide measurements using a hand-held electrochemical analyser
    Maniscalco, M.
    de Laurentiis, G.
    Weitzberg, E.
    Lundberg, J. O.
    Sofia, M.
    EUROPEAN JOURNAL OF CLINICAL INVESTIGATION, 2008, 38 (03) : 197 - 200
  • [5] Comparability of a hand-held nitric oxide analyser with online and offline chemiluminescence-based nitric oxide measurement
    Schiller, Barbara
    Hammer, Juerg
    Barben, Juerg
    Trachsel, Daniel
    PEDIATRIC ALLERGY AND IMMUNOLOGY, 2009, 20 (07) : 679 - 685
  • [6] Exhaled nitric oxide measurements: Correction equation to compare hand-held device to stationary analyzer
    Pizzimenti, Stefano
    Bugiani, Massimiliano
    Piccioni, Pavilio
    Heffler, Enrico
    Carosso, Aurelia
    Guida, Giuseppe
    Rolla, Giovanni
    RESPIRATORY MEDICINE, 2008, 102 (09) : 1272 - 1275
  • [7] Performance of a new hand-held device for exhaled nitric oxide measurement in adults and children
    Alving, K
    Janson, C
    Nordvall, L
    RESPIRATORY RESEARCH, 2006, 7 (1)
  • [8] Performance of a new hand-held device for exhaled nitric oxide measurement in adults and children
    K Alving
    C Janson
    L Nordvall
    Respiratory Research, 7
  • [9] Hand-held tidal breathing nasal nitric oxide measurement as a targeted case-finding tool for primary ciliary dyskinesia
    Marthin, June K.
    Nielsen, Kim G.
    EUROPEAN RESPIRATORY JOURNAL, 2012, 40
  • [10] Assessment of exhaled nitric oxide by a new hand-held device
    Antus, Balazs
    Horvath, Ildiko
    Barta, Imre
    RESPIRATORY MEDICINE, 2010, 104 (09) : 1377 - 1380