VALIDATION OF AN AUTOMATIC CONTOUR PROPAGATION METHOD FOR LUNG CANCER 4D ADAPTIVE RADIATION THERAPY

被引:4
|
作者
Peroni, M. [1 ]
Spadea, M. F. [1 ,2 ]
Riboldi, M. [1 ]
Baroni, G. [1 ]
Chen, G. T. Y. [3 ,4 ]
Sharp, G. C. [3 ,4 ]
机构
[1] Politecn Milano Univ, Dept Bioengn, Milan, Italy
[2] Magna Graecia Univ Catanzaro, Dept Expt & Clin Med, Catanzaro, Italy
[3] Massachusetts Gen Hosp, Dept Radiat Oncol, Boston, MA 02114 USA
[4] Harvard Med Sch, Boston, MA USA
来源
2009 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, VOLS 1 AND 2 | 2009年
关键词
Contour Propagation; Deformable Registration; Adaptive Radiation Therapy; 4D Treatment Planning; Re-planning; REGISTRATION;
D O I
10.1109/ISBI.2009.5193241
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this work is to validate a fast and reliable 4D-based adaptive treatment planning procedure for the Radiation Oncology clinic. A non-rigid B-Spline based registration is per-formed between CT scans at different respiratory phases. Radiotherapy contours drawn on a reference phase are than propagated to the target phase. We present results from two novel validation methods: a validation study using a 4D anthropomorphic computational phantom, and a patient data analysis comparing auto-rater versus inter-rater variability. For the phantom study, the overlap was greater than 90%, while the diameter variation was 5.1% at exhale and 3.4% at inhale. For the patient dataset, pairwise auto-rater Dice coefficient values exceeded inter-rater values for 9 of 15 structures. Though subject to inaccuracies caused by residual motion artifacts, and manual contouring variations, the reliability of the method suggests it could be soon introduced in the clinical practice.
引用
收藏
页码:1071 / +
页数:2
相关论文
共 50 条
  • [31] Evaluation of Elekta 4D cone beam CT-based automatic image registration for radiation treatment of lung cancer
    Li, Jun
    Harrison, Amy
    Yu, Yan
    Xiao, Ying
    Werner-Wasik, Maria
    Lu, Bo
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1053):
  • [32] Predictive adaptive radiation therapy for lung cancer
    Ramsey, C. R.
    Seibert, R.
    Hines, W.
    Meeks, S.
    Langen, K.
    Kupelian, P.
    Li, X.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S24 - S24
  • [33] A Method to Determine Individualized Prescription for Biologically Adaptive Radiation Therapy of Lung Cancer
    Tai, A.
    Li, X. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S690 - S690
  • [34] Comparison of 3D and 4D Plans in the Radiation Therapy for Lung Cancer Using Real Time Tumor Tracking System
    Okawa, K.
    Inoue, M.
    Taro, M.
    Koshi, T.
    Ota, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S891 - S892
  • [35] Biological modeling indices for 4D radiation therapy
    Anthony, J.
    Luxton, G.
    Leek, L.
    Chao, M.
    Carlson, D.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S629 - S629
  • [36] 4D CT lung ventilation images are affected by the 4D CT sorting method
    Yamamoto, Tokihiro
    Kabus, Sven
    Lorenz, Cristian
    Johnston, Eric
    Maxim, Peter G.
    Diehn, Maximilian
    Eclov, Neville
    Barquero, Cristian
    Loo, Billy W., Jr.
    Keall, Paul J.
    MEDICAL PHYSICS, 2013, 40 (10)
  • [37] 4D PET/CT VALUTATION ON INTEGRATION AND IMPACT IN RADIATION THERAPY PLANNING OF LUNG TUMOR
    D'Abbiero, N.
    Paiusco, M.
    Versari, A.
    Fioroni, F.
    Iori, M.
    Iotti, C.
    RADIOTHERAPY AND ONCOLOGY, 2010, 94 : S33 - S33
  • [38] 4D VMAT, gated VMAT, and 3D VMAT for stereotactic body radiation therapy in lung
    Chin, E.
    Loewen, S. K.
    Nichol, A.
    Otto, K.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (04): : 749 - 770
  • [39] Adaptive Re-Planning for Proton Treatment of Lung Cancer Using 4D Planning
    Zhao, L.
    Singh, H.
    Prabhu, K.
    Gary, L.
    Zheng, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S923 - S923
  • [40] Evaluation of 4D Cone Beam Computed Tomography for Target Localization in Stereotactic Body Radiation Therapy for Early Lung Cancer
    Glick, D.
    Karan, T.
    Le, K.
    Allibhai, Z.
    Taremi, M. M.
    Moseley, D. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S143 - S143