Evaluation of mouse red blood cell and platelet counting with an automated hematology analyzer

被引:23
|
作者
Fukuda, Teruko [1 ]
Asou, Eri [2 ]
Nogi, Kimiko [2 ]
Goto, Kazuo [1 ]
机构
[1] Teikyo Univ, Dept Clin Lab Med, 2-1-11 Kaga Itabashi, Tokyo 1738605, Japan
[2] Teikyo Univ Hosp, 2-1-11 Kaga, Tokyo 1730005, Japan
来源
JOURNAL OF VETERINARY MEDICAL SCIENCE | 2017年 / 79卷 / 10期
关键词
automated hematology analyzer; mouse; platelet; red blood cell; WHOLE-BLOOD; DOG; AGGREGOMETRY; PARAMETERS;
D O I
10.1292/jvms.17-0387
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
An evaluation of mouse red blood cell (RBC) and platelet (PLT) counting with an automated hematology analyzer was performed with three strains of mice, C57BL/6 (B6), BALB/c (BALB) and DBA/2 (D2). There were no significant differences in RBC and PLT counts between manual and automated optical methods in any of the samples, except for D2 mice. For D2, RBC counts obtained using the manual method were significantly lower than those obtained using the automated optical method (P<0.05), and PLT counts obtained using the manual method were higher than those obtained using the automated optical method (P<0.05). An automated hematology analyzer can be used for RBC and PLT counting; however, an appropriate method should be selected when D2 mice samples are used.
引用
收藏
页码:1707 / 1711
页数:5
相关论文
共 50 条
  • [31] Performance evaluation of the new platelet measurement channel on the BC-6800 Plus automated hematology analyzer
    Guo, Ping
    Cai, Qi
    Mao, Minjing
    Lin, Han
    Chen, Liting
    Wu, Fang
    Wang, Jianbiao
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2022, 44 (02) : 281 - 287
  • [32] MONITORING OF THE WHITE BLOOD CELLS FILTRATION BY AUTOMATED HEMATOLOGY ANALYZER
    Paiboonsukwong, Kittiphong
    Trisakulwattana, Krittanai
    Faknuam, Saruta
    Pattanapanyasat, Kovit
    Fucharoen, Suthat
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2019, 41 : 134 - 134
  • [33] Platelet counting using various hematology cell analyzers
    Dijkstra-Tiekstra, M. J.
    Kuipers, W.
    Setroikromo, A.
    de Wildt-Eggen, J.
    VOX SANGUINIS, 2006, 91 : 267 - 267
  • [34] Producing accurate platelet counts for platelet rich plasma: Validation of a hematology analyzer and preparation techniques for counting
    Woodell-May, JE
    Ridderman, DN
    Swift, MJ
    Higgins, J
    JOURNAL OF CRANIOFACIAL SURGERY, 2005, 16 (05) : 749 - 756
  • [35] AUTOMATED HEMATOLOGY - THE VALUE OF THE RED-BLOOD-CELL DISTRIBUTION WIDTH (RDW)
    FERNANDES, J
    JOURNAL OF THE AMERICAN OSTEOPATHIC ASSOCIATION, 1991, 91 (03): : 260 - 264
  • [36] IMPROVED AUTOMATED LEUKOCYTE COUNTING AND DIFFERENTIAL IN NEWBORNS ACHIEVED BY THE HEMATOLOGY ANALYZER CELL-DYN 3500(TM)
    DORNER, K
    SCHULZE, S
    REINHARDT, M
    SEEGER, H
    VANHOVE, L
    CLINICAL AND LABORATORY HAEMATOLOGY, 1995, 17 (01): : 23 - 30
  • [37] EVALUATION OF RETICULOCYTE COUNTING ON THE ROCHE COBAS M 511 HEMATOLOGY ANALYZER
    Harrison, Melody
    Zhang, Qian-Yun
    Bracco, Dan
    Allen, Tim
    Young, Steve
    George, Tracy
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 45 - 45
  • [38] Evaluation of IPF counting on Mindray BC-6800 hematology analyzer
    Pecoraro, A.
    Troia, A.
    Sacco, M.
    Maggio, A.
    Di Marzo, R.
    Gioia, M.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 (05) : E89 - E92
  • [39] EVALUATION OF THE TOA E-5000 AUTOMATED HEMATOLOGY ANALYZER
    PAYNE, BA
    PIERRE, RV
    LEE, WK
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1987, 88 (01) : 51 - 57
  • [40] EVALUATION OF THE AUTOMATED HEMATOLOGY ANALYZER SYSMEX NE-8000
    DEVREESE, K
    DELOGI, E
    FRANCART, C
    HEYNDRICKX, B
    PHILIPPE, J
    LEROUXROELS, G
    EUROPEAN JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1991, 29 (05): : 339 - 345