Validation of an automated home blood pressure measurement device in oldest-old populations

被引:8
|
作者
Godai, Kayo [1 ]
Kabayama, Mai [1 ]
Saito, Kanako [2 ]
Asayama, Kei [3 ,4 ]
Yamamoto, Koichi [5 ]
Sugimoto, Ken [5 ]
Ohkubo, Takayoshi [3 ,4 ]
Rakugi, Hiromi [5 ]
Kamide, Kei [1 ,5 ]
机构
[1] Osaka Univ, Grad Sch Med, Div Hlth Sci, Osaka, Japan
[2] Omron Healthcare Co Ltd, Kyoto, Japan
[3] Teikyo Univ, Dept Hyg & Publ Hlth, Sch Med, Tokyo, Japan
[4] Tohoku Inst Management Blood Pressure, Sendai, Miyagi, Japan
[5] Osaka Univ, Dept Geriatr & Gen Med, Grad Sch Med, Osaka, Japan
关键词
Blood pressure; Validation; Oldest-old; Oscillometric device; ISO; 81060-2; ANTHROPOMETRIC REFERENCE DATA; INTERNATIONAL-PROTOCOL; EUROPEAN-SOCIETY; ANSI/AAMI/ISO; 81060-2/2013; HYPERTENSION; AGE; ASSOCIATION; PREVALENCE; STIFFNESS; AWARENESS;
D O I
10.1038/s41440-019-0330-7
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Despite the wide use of automated devices for the self-measurement of home blood pressure (BP), no evidence is available regarding the accuracy of such devices in oldest-old populations. The aim of this study was to validate the accuracy of the automated oscillometric upper arm-cuff BP-monitoring device according to an international protocol in oldest-old individuals. In 35 participants aged over 85 years old, BP was measured on the same arm sequentially using a mercury sphygmomanometer (by two observers) and an Omron HEM-7080IC. The difference between the test device and observer measurements and associated factors were evaluated according to the International Organization for Standardization (ISO) 81060-2:2013 protocol. A total of 105 pairs (three pairs per participant) of the test device and observer BP measurements were obtained. The mean (+/- standard deviation: SD) differences in systolic BP (SBP) and diastolic BP (DBP) between the methods were -0.7 +/- 7.1 and -1.1 +/- 4.5 mmHg, respectively, and those for each participant were -0.7 +/- 5.8 mmHg for SBP and -1.1 +/- 4.1 mmHg for DBP; the device therefore fulfilled the requirements of the ISO protocol. In the multivariate analysis with the linear mixed model, the difference was associated with the cuff size for SBP and pulse pressure for DBP. The Omron HEM-7080IC passed the ISO requirements for oldest-old individuals aged 85 years or older. This device can be recommended for clinical and self/home use in oldest-old populations.
引用
收藏
页码:30 / 35
页数:6
相关论文
共 50 条
  • [1] Validation of an automated home blood pressure measurement device in oldest-old populations
    Kayo Godai
    Mai Kabayama
    Kanako Saito
    Kei Asayama
    Koichi Yamamoto
    Ken Sugimoto
    Takayoshi Ohkubo
    Hiromi Rakugi
    Kei Kamide
    Hypertension Research, 2020, 43 : 30 - 35
  • [2] Functional status of the oldest-old in a home setting
    Krach, P
    DeVaney, S
    DeTurk, C
    Zink, MH
    JOURNAL OF ADVANCED NURSING, 1996, 24 (03) : 456 - 464
  • [3] Validation of the Chinese version of the resilience scale for the oldest-old
    Wang, Ning
    Chen, Hongyu
    Zhang, Minyi
    Wang, Yuanyuan
    Xue, Zhihan
    Hao, Xixi
    Liu, Yan
    FRONTIERS IN PSYCHOLOGY, 2023, 14
  • [4] Blood micronutrient and thyroid hormone concentrations in the oldest-old
    Ravaglia, G
    Forti, P
    Maioli, F
    Nesi, B
    Pratelli, L
    Savarino, L
    Cucinotta, D
    Cavalli, G
    JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2000, 85 (06): : 2260 - 2265
  • [5] Validation of clinical criteria for possible vascular dementia in the oldest-old
    Bacchetta, Jean-Pierre
    Koevari, Eniko
    Merlo, Marco
    Canuto, Alessandra
    Herrmann, Francois R.
    Bouras, Constantin
    Gold, Gabriel
    Hof, Patrick R.
    Giannakopoulos, Panteleimon
    NEUROBIOLOGY OF AGING, 2007, 28 (04) : 579 - 585
  • [6] VALIDATION OF A SELF BLOOD PRESSURE MEASUREMENT DEVICE
    FOUQUERAY, B
    JULIEN, J
    PAGNY, JY
    JEUNEMAITRE, X
    SASSANO, P
    BATTAGLIA, C
    PLOUIN, PF
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1988, 81 : 231 - 234
  • [7] Validation of the Microlife WatchBP Home device for self home blood pressure measurement according to the International Protocol
    Stergiou, George S.
    Giovas, Periklis P.
    Gkino, Charilaos P.
    Patouras, John D.
    BLOOD PRESSURE MONITORING, 2007, 12 (03) : 185 - 188
  • [8] VALIDATION OF OSCILLOMAT, A DEVICE FOR HOME BLOOD-PRESSURE MEASUREMENT - COMPARISON WITH MERCURY SPHYGMOMANOMETERS
    ASMAR, R
    SASSANO, P
    DEMOLIS, P
    MENARD, J
    SAFAR, M
    PRESSE MEDICALE, 1991, 20 (12): : 551 - 555
  • [9] Are oldest-old nursing home diabetic residents a distinct clinical group?
    Buttar, A
    Blaum, C
    JOURNAL OF THE AMERICAN GERIATRICS SOCIETY, 2000, 48 (08) : S61 - S61
  • [10] Blood pressure and hypertension prevalence among oldest-old in China for 16 year: based on CLHLS
    Du, Jiajun
    Zhu, Guoning
    Yue, Yanhong
    Liu, Miao
    He, Yao
    BMC GERIATRICS, 2019, 19 (01)