Calibration of 10 nF Capacitance Standard from dc QHR using a Digital Impedance Bridge

被引:1
|
作者
Ngoc Thanh Mai Tran [1 ,3 ]
Kim, Wan-Seop [2 ]
Kim, Dan Bee [2 ]
机构
[1] Ist Nazl Ric Metrol, Str Cacce 91, I-10135 Turin, Italy
[2] Korea Res Inst Stand & Sci, 267 Gajeongno, Daejeon 34113, South Korea
[3] Politecn Torino, Corso Duca Abruzzi 24, I-10129 Turin, Italy
关键词
Capacitance measurement; Calibration; Measurement uncertainty; FARAD; RESISTANCE;
D O I
10.1109/cpem49742.2020.9191723
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes a calibration method for 10 nF capacitance standard from the primary standard of dc quantum Hall resistance using a multi-purpose digital coaxial bridge. The capacitance standard of 10 nF was calibrated by a direct comparison against the ac resistance standard of 10 k Omega- its resistance value is traced to the quantum Hall resistance- using the digital bridge in the fully digital quadrature mode. The preliminary calibration value of the 10 nF was confirmed by comparing with a reference capacitance standard within uncertainty level of 10(-6).
引用
收藏
页数:2
相关论文
共 50 条
  • [31] Preliminary Study for the Establishment of DC Charge Calibration Facility at CSIR-NPL Using Charged Capacitance Source
    Babita
    Narula, L. K.
    John, Thomas
    Saleem, M.
    Saxena, A. K.
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2014, 29 (04): : 269 - 272
  • [32] Calibration of Rogowski Coils at Frequencies up to 10 kHz Using Digital Sampling
    Djokic, Branislav
    I2MTC: 2009 IEEE INSTRUMENTATION & MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-3, 2009, : 418 - 422
  • [33] Calibration of Rogowski Coils at Frequencies up to 10 kHz Using Digital Sampling
    Djokic, Branislav
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2010, 59 (05) : 1303 - 1308
  • [34] Calibration of AC Shunts at Frequencies up to 10 kHz Using Digital Sampling
    Djokic, Branislav
    Parks, Harold
    2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [35] Standard Resistor Calibration Method in Measurement of Noise Source Impedance Using Dual-probe Approach
    Zhao Bo
    Zhao Min
    Feng Zhiming
    Yao Min
    CEEM: 2009 5TH ASIA-PACIFIC CONFERENCE ON ENVIRONMENTAL ELECTROMAGNETICS, 2009, : 408 - +
  • [36] Comparison of a 100-pF Capacitor With a 12 906-Ω Resistor Using a Digital Impedance Bridge
    Feige, Mona
    Schlamminger, Stephan
    Koffman, Andrew D.
    Jarrett, Dean G.
    Payagala, Shamith
    Panna, Alireza
    Waltrip, Bryan C.
    Berilla, Michael
    Seifert, Frank
    Wang, Yicheng
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2022, 71
  • [37] Calibration of reference torque wrenches using a 10 N m deadweight torque standard machine
    Nishino, Atsuhiro
    Ogushi, Koji
    Ueda, Kazunaga
    MEASUREMENT, 2015, 61 : 1 - 8
  • [38] Capacitance measurement method using sinusoidal voltage injection in isolating phase-shifted full-bridge DC-DC converter output stage
    Hannonen, Janne
    Honkanen, Jari
    Strom, Juha-Pekka
    Korhonen, Juhamatti
    Silventoinen, Pertti
    Raisanen, Samuli
    IET POWER ELECTRONICS, 2016, 9 (13) : 2543 - 2550
  • [39] Suppression of the Peak Harmonics from Loads by Using a Variable Capacitance Filter in Low-Voltage DC/DC Converters
    Huang, Jin
    Shi, Haixia
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2016, 58 (04) : 1217 - 1227
  • [40] Quantum Calibration System for Digital Voltmeters at Voltages from 10 nV to 1 kV
    Georgakopoulos, Dimitrios
    Budovsky, Ilya
    Grady, Stephen
    Hagen, Thomas
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2013, 62 (06) : 1581 - 1586