Quantitative Assessment of the Effect of Cholesterol on Blood Glucose Measurement Using Near Infrared Spectroscopy and a Method for Error Reduction

被引:5
|
作者
Jiang, Jingying [1 ]
Zhang, Kai [2 ]
Qin, Jia [4 ]
Min, Xiaolin [1 ]
Zhang, Lingling [1 ]
Zou, Da [1 ]
Xu, Kexin [3 ]
机构
[1] Tianjin Univ, Tianjin Key Lab Biomed Detecting Tech & Instrumen, Coll Precis Instruments & Optoelect Engn, Tianjin 300072, Peoples R China
[2] Tianjin Hosp, Blood Transfus Dept, Tianjin 300211, Peoples R China
[3] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
[4] Univ Calif San Francisco, San Francisco, CA 94143 USA
基金
中国国家自然科学基金; 国家高技术研究发展计划(863计划);
关键词
blood glucose concentration; cholesterol; diabetes; error reduction; impaired glucose tolerance; near-infrared spectroscopy; oral glucose tolerance test; partial least square; quantitative influence; reference wavelength-based method; TEMPERATURE; TISSUE;
D O I
10.1002/lsm.22317
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Background and ObjectiveThere is a growing body of evidence suggesting that the accurate measurement of blood glucose concentration can be perturbed by many factors. Current literature is limited in describing the influence of cholesterol on non-invasive blood glucose measurements by near-infrared spectroscopy (NIRS). This study aims to investigate the influence of cholesterol on blood glucose measurement through clinical oral glucose tolerance test (OGTT) and NIRS. Further, a method to reduce the prediction errors induced by cholesterol is proposed, facilitating the clinical application of non-invasive blood glucose sensing by NIRS. Study Design/Material and MethodsWe obtained clinical data of glucose and cholesterol concentrations at specific time points (0, 0.5, 1, 2, and 3h) during OGTTs from 115 subjects. The subjects were grouped into: Norm for normal control, IGT for Impaired Glucose Tolerance, and Diabetes. In addition, spectral data between 1200 and 1800nm were collected from 130 phantom samples, which are separated into seven groups depending on glucose and cholesterol levels. Statistical methods including One Sample T-test (OSTT), Pearson Correlation Analysis(PCA), and Unary Linear Regression (ULR) were used to analyze clinical data and spectral data to determine the relationship between glucose and cholesterol concentrations with the time course of OGTT. Reference wavelength-based method (RWM) was introduced to diminish the influence of cholesterol on glucose measurement and further the prediction error induced by cholesterol was reduced when using partial least square (PLS) model. ResultsClinical results statistically show that there is a strong negative correlation between the changes of glucose and cholesterol concentrations in the diabetes group. The spectra of cholesterol exhibit similar absorbance peaks to those of glucose within NIR range. PLS modelling results demonstrate that glucose prediction is influenced by cholesterol concentrations in a calibration model. Furthermore, a model expression (Cg=0.0356Cc+1.0129R(2)=0.993) is fitted to quantitatively describe the glucose prediction increment (Cg) due to cholesterol concentration (Cc). The results show that glucose prediction accuracy can be improved up to 38.36% by using RWM when using NIRS. ConclusionsThe cholesterol has an effect on blood glucose sensing. RWM is useful to help realize non-invasive blood glucose sensing by NIRS. Lasers Surg. Med. 47:88-97, 2015. (c) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:88 / 97
页数:10
相关论文
共 50 条
  • [31] Instantaneous quantitative and qualitative assessment of pear quality using near infrared spectroscopy
    Paz, Patricia
    Sanchez, Maria-Teresa
    Perez-Marin, Dolores
    Emilio Guerrero, Jose
    Garrido-Varo, Ana
    COMPUTERS AND ELECTRONICS IN AGRICULTURE, 2009, 69 (01) : 24 - 32
  • [32] Quantitative analysis of cholesterol and cholesterol ester mixtures using near-infrared Fourier transform Raman spectroscopy
    LeCacheux, P
    Menard, G
    Quang, HN
    Weinmann, P
    Jouan, M
    Nao, NQ
    APPLIED SPECTROSCOPY, 1996, 50 (10) : 1253 - 1257
  • [33] Non-invasive blood glucose measurement by near infrared spectroscopy: Machine drift, time drift and physiological effect
    Lam, Simon C. H.
    Chung, Joanne W. Y.
    Fan, K. L.
    Wong, Thomas K. S.
    SPECTROSCOPY-AN INTERNATIONAL JOURNAL, 2010, 24 (06): : 629 - 639
  • [34] Evaluation of measurement sites for noninvasive blood glucose sensing with near-infrared transmission spectroscopy
    Burmeister, JJ
    Arnold, MA
    CLINICAL CHEMISTRY, 1999, 45 (09) : 1621 - 1627
  • [35] In vivo noninvasive measurement of blood glucose by near-infrared diffuse-reflectance spectroscopy
    Maruo, K
    Tsurugi, M
    Tamura, M
    Ozaki, Y
    APPLIED SPECTROSCOPY, 2003, 57 (10) : 1236 - 1244
  • [36] Non-invasive Blood Glucose Measurement Scheme Based on Near-infrared Spectroscopy
    Wang Shulei
    Yuan Xueguang
    Zhang Yangan
    2017 CONFERENCE ON LASERS AND ELECTRO-OPTICS PACIFIC RIM (CLEO-PR), 2017,
  • [37] Quantitative determination of serum LDL cholesterol by near-infrared spectroscopy
    Liu, KZ
    Shi, MH
    Man, A
    Dembinski, TC
    Shaw, RA
    VIBRATIONAL SPECTROSCOPY, 2005, 38 (1-2) : 203 - 208
  • [38] Using EAs for error prediction in near infrared spectroscopy
    Fonlupt, C
    Cahon, S
    Robilliard, D
    Talbi, EG
    Duponchel, L
    APPLICATIONS OF EVOLUTIONARY COMPUTING, PROCEEDINGS, 2002, 2279 : 202 - 209
  • [39] Novel method for quantitative ANA measurement using near-infrared imaging
    Peterson, Lisa K.
    Wells, Daniel
    Shaw, Laura
    Velez, Maria-Gabriela
    Harbeck, Ronald
    Dragone, Leonard L.
    JOURNAL OF IMMUNOLOGICAL METHODS, 2009, 349 (1-2) : 1 - 8
  • [40] Applying the reference-wavelength method to improve the precision of glucose measurement by near infrared spectroscopy
    陈文亮
    徐可欣
    李宇
    陈韵
    ChineseOpticsLetters, 2007, (12) : 706 - 708