The performance of a cryogenically cooled monochromator for an in-vacuum undulator beamline

被引:53
|
作者
Zhang, L
Lee, WK
Wulff, M
Eybert, L
机构
[1] European Synchrotron Radiat Facil, F-38043 Grenoble, France
[2] Argonne Natl Lab, Adv Photon Source, Argonne, IL 60439 USA
关键词
X-ray optics; silicon-crystal monochromators; thermal deformation; cryogenic cooling; finite-element modeling;
D O I
10.1107/S0909049503012135
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The channel-cut silicon monochromator on beamline ID09 at the European Synchrotron Radiation Facility is indirectly cooled from the sides by liquid nitrogen. The thermal slope error of the diffracting surface is calculated by finite-element analysis and the results are compared with experiments. The slope error is studied as a function of cooling coefficients, beam size, position of the footprint and power distribution. It is found that the slope error versus power curve can be divided into three regions: (i) The linear region: the thermal slope error is linearly proportional to the power. (ii) The transition region: the temperature of the Si crystal is close to 125 K; the thermal slope error is below the straight line extrapolated from the linear curve described above. (iii) The non-linear region: the temperature of the Si crystal is higher than 125 K and the thermal slope error increases much faster than the power. Heat-load tests were also performed and the measured rocking-curve widths are compared with those calculated by finite-element modeling. When the broadening from the intrinsic rocking-curve width and mounting strain are included, the calculated rocking-curve width versus heat load is in excellent agreement with experiment.
引用
收藏
页码:313 / 319
页数:7
相关论文
共 50 条
  • [41] Cryogenically cooled bent double-Laue monochromator for high-energy undulator X-rays (50-200 keV)
    Shastri, SD
    Fezzaa, K
    Mashayekhi, A
    Lee, WK
    Fernandez, PB
    Lee, PL
    JOURNAL OF SYNCHROTRON RADIATION, 2002, 9 : 317 - 322
  • [42] Fast continuous energy scan with dynamic coupling of the monochromator and undulator at the DEIMOS beamline
    Joly, L.
    Otero, E.
    Choueikani, F.
    Marteau, F.
    Chapuis, L.
    Ohresser, P.
    JOURNAL OF SYNCHROTRON RADIATION, 2014, 21 : 502 - 506
  • [43] A VARIABLE RADIUS MIRROR FOR IMAGING THE EXIT SLIT MONOCHROMATOR UNDULATOR BEAMLINE AT THE ALS
    WARWICK, T
    HOWELLS, M
    SHLEZINGER, M
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02): : 2270 - 2272
  • [44] Design and performance of a stable first crystal mount for a cryogenically cooled Si monochromator at the Advanced Photon Source
    Dufresne, EM
    Arms, DA
    Dierker, SB
    Clarke, R
    Yacoby, Y
    Pitney, J
    MacHarrie, B
    Pindak, R
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (03): : 1511 - 1513
  • [45] Performance analysis of a side-cooled vertical-focusing mirror for undulator beamline at APS
    Jaski, Y
    Meron, M
    Viccaro, PJ
    ADVANCES IN MIRROR TECHNOLOGY FOR SYNCHROTRON X-RAY AND LASER APPLICATIONS, 1998, 3447 : 62 - 71
  • [46] Fast continuous energy scan with dynamic coupling of the monochromator and undulator at the DEIMOS beamline
    Joly, L.
    Otero, E.
    Choueikani, F.
    Marteau, F.
    Chapuis, L.
    Ohresser, P.
    Journal of Synchrotron Radiation, 2014, 21 (03) : 502 - 506
  • [47] IN-VACUUM TYPE UNDULATOR FOR A VISIBLE UV REGION FEL USING A LINAC
    KEISHI, T
    KOBAYASHI, A
    TOMIMASU, T
    OKAZAKI, T
    HOSODA, Y
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 341 (1-3): : AB128 - AB129
  • [48] Magnet System Optimization for Segmented Adaptive-Gap In-Vacuum Undulator
    Kitegi, C.
    Chubar, O.
    Eng, C.
    PROCEEDINGS OF THE 12TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION (SRI2015), 2016, 1741
  • [49] Damping trapped modes in an in-vacuum undulator at a synchrotron radiation light source
    Tian, Kai
    Sebek, James J.
    Ringwall, Andrew D.
    Li, Zenghai
    PHYSICAL REVIEW ACCELERATORS AND BEAMS, 2019, 22 (05):
  • [50] X-25 cryo-ready in-vacuum undulator at the NSLS
    Tanabe, Toshiya
    Ablett, James
    Berman, Lonny
    Harder, David A.
    Hulbert, Steve
    Lehecka, Mike
    Rakowsky, George
    Skaritka, John
    Deyhim, Alex
    Johnson, Eric
    Kulesza, Joe
    Waterman, Dave
    SYNCHROTRON RADIATION INSTRUMENTATION, PTS 1 AND 2, 2007, 879 : 283 - +