Thermoluminescent dosimeters (TLDs-100) calibration for dose verification in photon and proton radiation therapy

被引:8
|
作者
D'Avino, V. [1 ]
Caruso, M. [2 ]
Arrichiello, C. [3 ]
Ametrano, G. [3 ]
La Verde, G. [1 ,4 ,5 ]
Muto, P. [3 ]
Scifoni, E. [6 ]
Tommasino, F. [6 ,7 ]
Pugliese, M. [1 ,4 ]
机构
[1] Natl Inst Nucl Phys, Sect Naples, Naples, Italy
[2] Univ Naples Federico II, Dept Elect Engn & Informat Technol, Naples, Italy
[3] Fdn G Pascale, IRCCS, Radiotherapy Unit, Ist Nazl Tumori, Naples, Italy
[4] Univ Naples Federico II, Dept Phys E Pancini, Naples, Italy
[5] Univ Naples Federico II, Dept Pharm, Naples, Italy
[6] Ist Nazl Fis Nucl, Trento Inst Fdn Phys & Applicat, TIFPA INFN, Trento, Italy
[7] Univ Trento, Dept Phys, Via Sommar 14, Trento, Italy
关键词
DEPENDENCE; TOXICITY; DAMAGE;
D O I
10.1393/ncc/i2020-20142-0
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Thermoluminescent dosimeters (TLDs) are practical, accurate, and precise tools for point dosimetry in medical physics applications. TLDs are nowadays extensively used to measure dose in conformal radiation therapy in order to guarantee the safety of the treatment. Several national and international organizations recommend checking the effective dose delivered to an individual patient by means of in vivo dosimetry. Modern radiotherapy techniques employing both photon and ion beams exhibit excellent target conformation throught high steep-dose gradients between tumour and adjacent organs and tissues. In this context, catching potential dose errors and uncertainties in treatment delivering is the first step to ensure the optimization of the treatment plan. This study shows the results of the characterization of TLDs-100 at two Italian facilities devoted to advanced radiation treatments with photon and proton therapy. The individual sensitivity factor was determined, and the calibration curves were carried out in the dose range 0-20 Gy. By the analysis of the dose response curves, the linear region was identified under the dose level of 10 Gy. Characterization of the TLDs-100 has enabled their use for in vivo dosimetry especially in the dose range corresponding to the linear region of the dose response curves.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Dose verification and plan conformity with three different dosimeters for intensity-modulated radiation therapy plans
    Yassin, A.
    Elshahat, K. M.
    Attlah, E. M.
    Khaled, H.
    Kany, A.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (03): : 703 - 710
  • [22] Dose profile measurements in a Leksell Gamma Knife using TLD-100H thermoluminescent dosimeters
    Li, Y
    Woollard, J
    Gupta, N
    MEDICAL PHYSICS, 2003, 30 (06) : 1480 - 1480
  • [23] Breast specific cone-beam CT calibration for daily dose verification in proton therapy
    Omand Kirkegaard Sorensen, T.
    Muren, L.
    Elstrom, U. V.
    Kronborg, C.
    Offersen, B. V.
    Petersen, S. E.
    Stick, L. B.
    Taasti, V. T.
    Jensen, M. F.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S130 - S132
  • [24] In vivo dose verification for photon treatments of head and neck carcinomas using MOSFET dosimeters
    Tung, C. J.
    Wang, L. C.
    Wang, H. C.
    Lee, C. C.
    Chao, T. C.
    RADIATION MEASUREMENTS, 2008, 43 (2-6) : 870 - 874
  • [25] Evaluation of external dose equivalent with thermoluminescent dosimeters from residents living in radiation-contaminated buildings
    Lee, JS
    Dong, SL
    Chang, WP
    Chan, CC
    APPLIED RADIATION AND ISOTOPES, 1997, 48 (09) : 1237 - 1243
  • [26] METHODOLOGY FOR THE CALIBRATION OF PHOTON PERSONAL DOSIMETERS - CALCULATIONS OF PHANTOM BACKSCATTER AND DEPTH DOSE DISTRIBUTIONS
    BARTLETT, DT
    FRANCIS, TM
    DIMBYLOW, PJ
    RADIATION PROTECTION DOSIMETRY, 1989, 27 (04) : 231 - 244
  • [27] Photon Spectroscopy as a Tool for Range Verification In Proton Therapy
    Burbadge, C.
    Muecher, D.
    Kasanda, E.
    Hoehr, C.
    Bildstein, V.
    Mampaso, B. Olaizola
    MEDICAL PHYSICS, 2019, 46 (06) : E326 - E326
  • [29] Secondary neutron and photon dose in proton therapy
    Agosteo, S
    Birattari, C
    Caravaggio, M
    Silari, M
    Tosi, G
    RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) : 293 - 305