Monte Carlo Modeling of the Head of Varian Clinac 2100C Linear Accelerator Operating in 18 MV Utilizing GATE/GEANT4

被引:0
|
作者
Bezoubiri, Fethi [1 ]
Alem-Bezoubiri, Asma [1 ]
Khoudri, Saad [1 ,2 ]
Arib, Mehenna [3 ]
机构
[1] CRNA, Dept Phys Med, 02 Blvd Frantz Fanon,BP 399, Algiers 16000, RP, Algeria
[2] King Faisal Specialist Hosp & Res Cenfre, Ctr Reg Lutte Canc Blida, Serv Radiotherapie, Riyadh, Saudi Arabia
[3] King Faisal Specialist Hosp & Res Cenfre, Biomed Phys Dept, Riyadh, Saudi Arabia
关键词
D O I
10.1063/1.5048162
中图分类号
O59 [应用物理学];
学科分类号
摘要
In this study, the geometry of the head of VARIAN CLINAC 2100C linear accelerator lodged at Blida AntiCancer Centre in Algeria has been modeled using GATE/GEANT4 version 7.2. The model developed contains the majority of the components of the treatment head operating in 18 MV, such as: the target, the primary collimators, the flattening filter, the secondary collimators and the jaws. Using this model, associated with a water phantom of type RFA 200 with dimensions 63.8 cm x 40 cm x 44 cm positioned at a Source-to-Surface-Distance (SSD) of 100 cm, the interactions of the 18 MV radiations in water were simulated. The photon energy spectrum, the Percentage Depth Dose (PDD) and the beam quality specifiers such as D-10 (%), d(max) (cm), d(80)(cm), D-20/D-10 were calculated for field size of 10 cm x 10 cm utilizing the phase space technique. The results obtained in this study are in agreement with those obtained in our previous work using MCNP5 code.
引用
收藏
页数:4
相关论文
共 34 条
  • [31] Development and Validation of a 0.35T MR-Guided Linear Accelerator Monte Carlo Model in GEANT4
    Khan, A.
    Simiele, E.
    Yadav, P.
    MEDICAL PHYSICS, 2020, 47 (06) : E741 - E741
  • [32] Comparison of EGS4 and MCNP4b Monte Carlo codes for generation of photon phase space distributions for a Varian 2100C
    Siebers, JV
    Keall, PJ
    Libby, B
    Mohan, R
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (12): : 3009 - 3026
  • [33] A full quantitative analysis of 18 MV photon beam from 2100 C/D-Varian clinical linear accelerator with and without flattening filter
    Mahdavi, S. R.
    Ay, M. R.
    Zabihzadeh, M.
    Allahverdi, M.
    Shahriari, M.
    Hoseini-Ghahfarokhi, M.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2019, 17 (01): : 137 - 146
  • [34] Energy spectra, angular spread, fluence profiles and dose distributions of 6 and 18 MV photon beams: results of Monte Carlo simulations for a Varian 2100EX accelerator
    Ding, GX
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (07): : 1025 - 1046