Accuracy of real-time optoacoustic temperature determination during retinal photocoagulation

被引:4
|
作者
Baade, A. [1 ]
Schlott, K. [1 ]
Luft, S. [2 ]
Ptaszynski, L. [1 ]
Bever, M. [1 ]
Birngruber, R. [1 ,2 ]
Brinkmann, R. [1 ,2 ]
机构
[1] Med Laserzentrum Lubeck GmbH, Peter Monnik Weg 4, D-23562 Lubeck, Germany
[2] Inst Biomedizinische Optik, D-23562 Lubeck, Germany
关键词
optoacoustics; ophthalmology; retina; photocoagulation; temperature;
D O I
10.1117/12.889608
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Retinal photocoagulation is an established treatment for various retinal diseases. The temperature development during a treatment can be monitored by applying short laser pulses in addition to the treatment laser light. The laser pulses induce optoacoustic pressure waves that can be detected at the cornea. Aim of this work is the investigation of the accuracy of the determined temperatures during a treatment. To calibrate the temperature dependency of the measured pressure, whole enucleated porcine eyes were heated using an infrared laser beam, while probing the retina optoacoustically. The temperatures and the optoacoustic pressure waves were measured simultaneously using thermocouples and a piezoelectric element, respectively. From the deviation of the individual measurements an error of less than 15% in the calibration regime between 37 degrees C to 55 degrees C was found. Furthermore, the spatial and temporal temperature course was investigated. Calculations were performed to simulate the temporal and spatial temperature development during photocoagulation. A theoretical model to determine the peak temperature of the irradiated tissue from the mean temperature measured by optoacoustics was developed. The validity of the model was experimentally examined by heating the retina of porcine eyes with a laser beam diameter of 500 mu m while successively measuring the temperature optoacoustically with a probe beam diameter of 500 mu m and 100 mu m at the center of the heated area, respectively. The deviation of the theoretical model and the experimental results were found to be less than 7%.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Advances in real-time multispectral optoacoustic imaging and its applications
    Adrian Taruttis
    Vasilis Ntziachristos
    Nature Photonics, 2015, 9 : 219 - 227
  • [32] Advances in real-time multispectral optoacoustic imaging and its applications
    Taruttis, Adrian
    Ntziachristos, Vasilis
    NATURE PHOTONICS, 2015, 9 (04) : 219 - 227
  • [33] THE FABRICATION OF AN OPTOACOUSTIC TRANSDUCER FOR REAL-TIME DIAGNOSTIC-IMAGING
    SU, KY
    WADE, G
    IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1981, 28 (05): : 367 - 367
  • [34] Monitoring of temperature distribution in tissues with optoacoustic technique in real time
    Larin, K
    Larina, I
    Motamedi, M
    Esenaliev, R
    BIOMEDICAL OPTOACOUSTICS, 2000, 3916 : 311 - 321
  • [35] Real-time spectroscopy of transition states in bacteriorhodopsin during retinal isomerization
    Kobayashi, T
    Saito, T
    Ohtani, H
    NATURE, 2001, 414 (6863) : 531 - 534
  • [36] Real-time spectroscopy of transition states in bacteriorhodopsin during retinal isomerization
    Takayoshi Kobayashi
    Takashi Saito
    Hiroyuki Ohtani
    Nature, 2001, 414 : 531 - 534
  • [37] REAL-TIME ALGORITHM FOR RETINAL TRACKING
    MARKOW, MS
    RYLANDER, HG
    WELCH, AJ
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (12) : 1269 - 1281
  • [38] Modeling and parameter identification for real-time temperature controlled retinal laser therapies
    Kleyman, Viktoria
    Gernandt, Hannes
    Worthmann, Karl
    Abbas, Hossam S.
    Brinkmann, Ralf
    Mueller, Matthias A.
    AT-AUTOMATISIERUNGSTECHNIK, 2020, 68 (11) : 953 - 966
  • [39] REAL-TIME DETERMINATION OF IONOSPHERIC CURRENT PARAMETERS DURING A SUBSTORM
    HANSER, FA
    SELLERS, B
    VANCOUR, RP
    JOURNAL OF GEOMAGNETISM AND GEOELECTRICITY, 1973, 25 (04): : 339 - 348
  • [40] Accuracy of 3D real-time MRI temperature mapping in gel phantoms during microwave heating
    Dietrich, Olaf
    Lentini, Sergio
    Oecal, Osman
    Bour, Pierre
    Faller, Thibaut L.
    Ozenne, Valery
    Ricke, Jens
    Seidensticker, Max
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)