Online C-arm calibration using a marked guide wire for 3D reconstruction of pulmonary arteries

被引:2
|
作者
Vachon, Etienne [1 ]
Miro, Joaquim [2 ]
Duong, Luc [1 ]
机构
[1] Ecole Technol Super, 1100 Notre Dame West, Montreal, PQ, Canada
[2] CHU St Justine, 3145 Cote St Catherine, Montreal, PQ, Canada
关键词
X-ray-Fluoroscopy; X-ray imaging applications; Cardiac imaging and image analysis;
D O I
10.1117/12.2254175
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
3D reconstruction of vessels from 2D X-ray angiography is highly relevant to improve the visualization and the assessment of vascular structures such as pulmonary arteries by interventional cardiologists. However, to ensure a robust and accurate reconstruction, C-arm gantry parameters must be properly calibrated to provide clinically acceptable results. Calibration procedures often rely on calibration objects and complex protocol which is not adapted to an intervention context. In this study, a novel calibration algorithm for C-arm gantry is presented using the instrumentation such as catheters and guide wire. This ensures the availability of a minimum set of correspondences and implies minimal changes to the clinical workflow. The method was evaluated on simulated data and on retrospective patient datasets. Experimental results on simulated datasets demonstrate a calibration that allows a 3D reconstruction of the guide wire up to a geometric transformation. Experiments with patients datasets show a significant decrease of the retro projection error to 0.17 mm 2D RMS. Consequently, such procedure might contribute to identify any calibration drift during the intervention.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Intraoperative imaging of zygomaticomaxillary complex fractures using a 3D C-arm system
    Heiland, M
    Schulze, D
    Blake, F
    Schmelzle, R
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2005, 34 (04) : 369 - 375
  • [22] C-arm angiography (DYNA-CT) for 3D Coronary Reconstruction and Myocardial Perfusion Assessment
    Rittger, Harald
    Lauritsch, Guenter
    Brachmann, Johannes
    Rieber, Johannes
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (17) : B93 - B93
  • [23] 3D Vessel Tree Reconstruction from Rotational C-arm Projections by Multi-view Stereo Reconstruction
    Zhang, Wei
    Chen, Zikuan
    Zhang, Xuefeng
    APCMBE 2008: 7TH ASIAN-PACIFIC CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING, 2008, 19 : 434 - +
  • [24] Low contrast 3D-reconstruction from C-arm data
    Zellerhoff, M
    Scholz, B
    Rührnschopf, EP
    Brunner, T
    MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 : 646 - 655
  • [25] A method for extracting 3D information of cylinder in C-arm image
    Wu, Jian
    BMEI 2008: PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS, VOL 2, 2008, : 544 - +
  • [26] Intraoperative 3D C-arm imaging. State of the art
    von Recum, J.
    Wendl, K.
    Vock, B.
    Gruetzner, P. A.
    Franke, J.
    UNFALLCHIRURG, 2012, 115 (03): : 196 - 201
  • [27] 3-D geometry calibration and markerless electromagnetic tracking with a mobile C-arm
    Cheryauka, Arvi
    Barrett, Johnny
    Wang, Zhonghua
    Litvin, Andrew
    Hamadeh, Ali
    Beaudet, Daniel
    MEDICAL IMAGING 2007: VISUALIZATION AND IMAGE-GUIDED PROCEDURES, PTS 1 AND 2, 2007, 6509
  • [28] 3D X-ray imaging of human bones using a mobile C-arm
    Lütjens, J
    Koppe, R
    Klotz, E
    Grass, M
    Rasche, V
    CARS 2002: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2002, : 1016 - 1016
  • [29] Registration of 2D C-Arm and 3D CT Images for a C-Arm Image-Assisted Navigation System for Spinal Surgery
    Chang, Chih-Ju
    Lin, Geng-Li
    Tse, Alex
    Chu, Hong-Yu
    Tseng, Ching-Shiow
    APPLIED BIONICS AND BIOMECHANICS, 2015, 2015
  • [30] Exact 3D cone-beam reconstruction from two short-scans using a C-arm imaging system
    Ramamurthi, K
    Strobel, N
    Prince, JL
    COMPUTATIONAL IMAGING III, 2005, 5674 : 87 - 98