Accuracy and efficiency of an infrared based positioning and tracking system for patient set-up and monitoring in image guided radiotherapy

被引:6
|
作者
Jia, Jing [1 ,2 ,3 ]
Xu, Gongming [2 ,3 ]
Pei, Xi [2 ,3 ]
Cao, Ruifen [2 ,3 ]
Hu, Liqin [1 ,2 ,3 ]
Wu, Yican [1 ,2 ,3 ]
机构
[1] Univ Sci & Technol China, Hefei 230027, Anhui, Peoples R China
[2] Chinese Acad Sci, Inst Nucl Energy Safety Technol, Key Lab Neutron & Radiat Safety, Hefei 230031, Anhui, Peoples R China
[3] Engn Technol Res Ctr Accurate Radiotherapy Anhui, Hefei 230031, Anhui, Peoples R China
基金
中国国家自然科学基金;
关键词
Infrared positioning; Respiratory control; Patient setup; QUALITY-ASSURANCE;
D O I
10.1016/j.infrared.2015.01.001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An infrared based positioning and tracking (IPT) system was introduced and its accuracy and efficiency for patient setup and monitoring were tested for daily radiotherapy treatment. The IPT system consists of a pair of floor mounted infrared stereoscopic cameras, passive infrared markers and tools used for acquiring localization information as well as a custom controlled software which can perform the positioning and tracking functions. The evaluation of IPT system characteristics was conducted based on the AAPM 147 task report. Experiments on spatial drift and reproducibility as well as static and dynamic localization accuracy were carried out to test the efficiency of the IPT system. Measurements of known translational (up to 55.0 mm) set-up errors in three dimensions have been performed on a calibration phantom. The accuracy of positioning was evaluated on an anthropomorphic phantom with five markers attached to the surface; the precision of the tracking ability was investigated through a sinusoidal motion platform. For the monitoring of the respiration, three volunteers contributed to the breathing testing in real time. The spatial drift of the IPT system was 0.65 mm within 60 min to be stable. The reproducibility of position variations were between 0.01 and 0.04 mm. The standard deviation of static marker localization was 0.26 mm. The repositioning accuracy was 0.19 mm, 0.29 mm, and 0.53 mm in the left/right (L/R), superior/inferior (S/I) and anterior/posterior (A/P) directions, respectively. The measured dynamic accuracy was 0.57 mm and discrepancies measured for the respiratory motion tracking was better than 1 mm. The overall positioning accuracy of the IPT system was within 2 mm. In conclusion, the IPT system is an accurate and effective tool for assisting patient positioning in the treatment room. The characteristics of the IPT system can successfully meet the needs for real time external marker tracking and patient positioning as well as respiration monitoring during image guided radiotherapy treatments. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:26 / 31
页数:6
相关论文
共 50 条
  • [21] The impact of reference isocentre position on set-up errors in head-and-neck image-guided radiotherapy
    Lenko, Helena
    Peterlin, Primoz
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (01) : 104 - 113
  • [22] Detection of Patient Set-up Errors with a Portal Image - DRR Registration Software System
    Sutherland, K.
    Ishikawa, M.
    Bengua, G.
    Shirato, H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S600 - S600
  • [23] Body weight loss associates with set-up error in nasopharyngeal cancer patients undergoing image guided radiotherapy
    Wang, C.
    Chong, F.
    Wu, J.
    Lai, M.
    Cheng, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S203 - S203
  • [24] Accuracy of a patient positioning system for image-guided frameless stereotactic IMRT treatments
    Dohm, O.
    Hoenes, A.
    Paulsen, F.
    Alber, M.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S288 - S288
  • [25] Ion Radiography as a Tool for Patient Set-up & Image Guided Particle Therapy: A Monte Carlo Study
    Depauw, Nicolas
    Dias, Marta F.
    Rosenfeld, Anatoly
    Seco, Joao C.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2014, 13 (01) : 69 - 76
  • [26] EVALUATION OF THE EFFECT OF TUMOUR BED MARKER MOVEMENT UPON PATIENT SET-UP ACCURACY DURING BREAST RADIOTHERAPY
    Harris, E.
    Donovan, E.
    Coles, C.
    Eagle, S.
    Kirby, A.
    Evans, P.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S439 - S439
  • [27] TRACKING ACCURACY OF A REAL-TIME FIDUCIAL TRACKING SYSTEM FOR PATIENT POSITIONING AND MONITORING IN RADIATION THERAPY
    Shchory, Tal
    Schifter, Dan
    Lichtman, Rinat
    Neustadter, David
    Corn, Benjamin W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (04): : 1227 - 1234
  • [28] Anatomic feature-based registration for patient set-up in head and neck cancer radiotherapy
    Sharp, GC
    Kollipara, S
    Madden, T
    Jiang, SB
    Rosenthal, SJ
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (19): : 4667 - 4679
  • [29] Modeling and prediction of set-up errors in breast cancer image-guided radiotherapy using the Gaussian mixture model
    Dong, Fangfen
    Chen, Jing
    Liu, Feiyu
    Yang, Zhiyu
    Wu, Yimin
    Li, Xiaobo
    ONCOLOGY LETTERS, 2024, 28 (06)
  • [30] Accuracy of a photogrammetry-based patient positioning and monitoring system for radiation therapy
    Rogus, RD
    Stern, RL
    Kubo, HD
    MEDICAL PHYSICS, 1999, 26 (05) : 721 - 728