Validation of image registration and fusion of MV CBCT and planning CT for radiotherapy treatment planning

被引:3
|
作者
Thomas, T. Hannah Mary [1 ]
Devakumar, D. [2 ]
Balukrishna, S. [3 ]
Godson, Henry Finlay [3 ]
Ravindran, B. Paul [3 ]
机构
[1] Christian Med Coll & Hosp, Dept Bioengn, Vellore, Tamil Nadu, India
[2] Christian Med Coll & Hosp, Dept Nucl Med, Vellore, Tamil Nadu, India
[3] Christian Med Coll & Hosp, Dept Radiotherapy, Vellore, Tamil Nadu, India
关键词
Megavoltage cone beam CT; Registration; Planning CT; Fusion; Validation; CONE-BEAM CT; COMPUTED-TOMOGRAPHY; MEGAVOLTAGE CT; RADIATION-THERAPY; KILOVOLTAGE; SCANNER;
D O I
10.1007/s13246-011-0092-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In areas like adaptive therapy, multi-phase radiotherapy, and single fraction palliative treatment or in the treatment of patients with metal implants where megavoltage(MV) CT could be considered as a treatment planning modality, the reduced contrast in the MV CT images could lead to limited accuracy in localization of the structures. This would affect the precision of the treatment. In this study, as an extension our previous work on bespoke MV cone beam CT (MV CBCT), we propose to register the MV CBCT with kilovoltage (kV) CT for treatment planning. The MV CBCT images registered with kV CT would be effective for treatment planning as it would account for the inadequate soft tissue information in the MV CBCT and would allow comparison of changes in patient dimensions and assist in localization of the structures. The intensity based registration algorithm of the BrainSCAN therapy planning software was used for image registration of the MV CBCT and kV CT images. The accuracy of the registration was validated using qualitative and quantitative measures. The effect of image quality on the level of agreement between the contouring done on both the MV CBCT and kV CT was assessed by comparing the volumes of six structures delineated. To assess the level of agreement between the plans after the registration, two independent plans were generated on the MV CBCT and the planning CT using the posterior fossa of the skull as the target. The dose volume histograms and conformity indices of the plans were compared. The results of this study show that treatment planning with MV CBCT images would be effective, using additional anatomical structure information derived from registering the MV CBCT image with a standard kVCT.
引用
收藏
页码:441 / 447
页数:7
相关论文
共 50 条
  • [41] CBCT Based External Radiotherapy Planning
    Pal, S.
    Singh, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2928 - 2929
  • [42] Effect of MRI/CT Image Fusion on Radiotherapy Planning for Central Nervous System Lymphoma
    Li, D.
    Li, Y.
    Sun, Y.
    Xu, B.
    Wang, W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E474 - E474
  • [43] A strategy for multimodal deformable image registration to integrate PET/MR into radiotherapy treatment planning
    Leibfarth, Sara
    Moennich, David
    Welz, Stefan
    Siegel, Christine
    Schwenzer, Nina
    Schmidt, Holger
    Zips, Daniel
    Thorwarth, Daniela
    ACTA ONCOLOGICA, 2013, 52 (07) : 1353 - 1359
  • [44] The use of PET/CT in radiotherapy planning: contribution of deformable registration
    Soykut, Ela Delikgoz
    Ozsahin, Esat Mahmut
    Guney, Yildiz Yukselen
    Arslan, Suheyla Aytac
    Or, Ozlem Derinalp
    Altundag, Muzaffer Bedri
    Ugurluer, Gamze
    Tsoutsou, Pelagia G.
    FRONTIERS IN ONCOLOGY, 2013, 3
  • [45] CDFRegNet: A cross-domain fusion registration network for CT-to-CBCT image registration
    Cao, Yuzhu
    Fu, Tianxiao
    Duan, Luwen
    Dai, Yakang
    Gong, Lun
    Cao, Weiwei
    Liu, Desen
    Yang, Xiaodong
    Ni, Xinye
    Zheng, Jian
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2022, 224
  • [46] Virtual CT for adaptive prostate radiotherapy based on CT-CBCT deformable image registration
    Cassetta Junior, F. R.
    Ciardo, D.
    Fattori, G.
    Riboldi, M.
    Orecchia, R.
    Jereczek-Fossa, B. A.
    Baroni, G.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S902 - S903
  • [47] Accuracy in automatic image registration between MV cone beam computed tomography and planning kV computed tomography in image guided radiotherapy
    Kanakavelu, Nithya
    Samuel, E. James Jebaseelan
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2016, 21 (05) : 487 - 494
  • [48] Validation of a deformable image registration produced by a commercial treatment planning system in head and neck
    Garcia-Molla, Rafael
    de Marco-Blancas, Noelia
    Bonaque, Jorge
    Vidueira, Laura
    Lopez-Tarjuelo, Juan
    Perez-Calatayud, Jose
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (03): : 219 - 223
  • [49] A Comparison of Adaptive Radiotherapy Treatment Planning Between Planning CT and KVCBCT
    Juh, R.
    Back, G.
    Ahn, S.
    Kim, J.
    Suh, T.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [50] MRI/CT is the future of radiotherapy treatment planning
    Glide-Hurst, Carri K.
    Low, Daniel A.
    Orton, Colin G.
    MEDICAL PHYSICS, 2014, 41 (11)