Arc gradient coils design for unilateral magnetic resonance sensor based on Fourier series method

被引:2
|
作者
Wu, Jiamin [1 ]
He, Yucheng [1 ]
Shen, Sheng [1 ]
Guo, Pan [2 ]
Huang, Xin [1 ]
Xu, Zheng [1 ]
机构
[1] Chongqing Univ, State Key Lab Power Transmiss Equipment & Syst Se, Chongqing 400044, Peoples R China
[2] Chongqing Normal Univ, Sch Phys & Elect Engn, Chongqing 401331, Peoples R China
基金
中国国家自然科学基金;
关键词
Unilateral magnetic resonance; Fourier series method; gradient coil design; NMR-MOUSE; FIELD; SYSTEM;
D O I
10.3233/JAE-190076
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Traditional nuclear magnetic instruments are not convenient for measurement in situ due to the closed magnet structure and heavy weight. It is crucial to improve tree planting by exactly detecting moisture parameters of living trees. To investigate this problem, an open and portable magnetic resonance sensor was proposed. In this work, we introduced unilateral magnetic resonance sensor which can achieve onsite nondestructive measurement of large-volume samples due to its open structure. To fit the shape of living trees, the region of interest was set in an arc region. A magnet structure was optimized by ANSYS software. Because the magnet has a constant natural gradient along radial direction, we need gradient fields in the other two directions (Z and Y). Z/Y gradient coils for unilateral magnetic resonance sensor are studied based on the Fourier series method. Current density in gradient coils is assumed as Fourier series composed of trigonometric functions. We introduced stream function to establish relationship between the current density and the winding structure, then constructed an optimization model by combining power dissipation with the deviation between calculated field and target field. The coils are manufactured using a flexible printed circuit board and measured by gauss meter. The measurement and simulation results are consistent. Measurement results of the gradient coils showed that the magnetic fields have good linearity which is helpful to realize 3D measurement in the upcoming work.
引用
收藏
页码:447 / 459
页数:13
相关论文
共 50 条
  • [1] Gradient Coils Design with Regularization Method for Superconducting Magnetic Resonance Imaging
    Zheng, Sijie
    Li, Xia
    Yan, Wei Qi
    Xu, Wenlong
    2018 INTERNATIONAL CONFERENCE ON IMAGE AND VISION COMPUTING NEW ZEALAND (IVCNZ), 2018,
  • [2] DESIGN OF GRADIENT COILS FOR MAGNETIC-RESONANCE-IMAGING
    ADAMIAK, K
    RUTT, BK
    DABROWSKI, WJ
    IEEE TRANSACTIONS ON MAGNETICS, 1992, 28 (05) : 2403 - 2405
  • [3] Unshielded Gradient Coils Design For Magnetic Resonance Imaging
    Hidalgo, S. S.
    Gadzinski, C.
    Rutt, B.
    MEDICAL PHYSICS, 2012, 1494 : 54 - 57
  • [4] A Method for Coil Design of Superconducting Quadrupoles Based on Sector Coils and Fourier Series
    Borgnolutti, F.
    Todesco, E.
    Mailfert, A.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2010, 20 (03) : 1790 - 1793
  • [5] Equivalent Magnetization Current Method Applied to the Design of Gradient Coils for Magnetic Resonance Imaging
    Lopez, Hector Sanchez
    Liu, Feng
    Poole, Michael
    Crozier, Stuart
    IEEE TRANSACTIONS ON MAGNETICS, 2009, 45 (02) : 767 - 775
  • [6] Equivalent magnetic dipole method used to design gradient coil for unilateral magnetic resonance imaging
    Xu, Zheng
    Li, Xiang
    Guo, Pan
    Wu, Jia-Min
    CHINESE PHYSICS B, 2018, 27 (05)
  • [7] Equivalent magnetic dipole method used to design gradient coil for unilateral magnetic resonance imaging
    徐征
    李想
    郭盼
    吴嘉敏
    Chinese Physics B, 2018, (05) : 549 - 554
  • [8] Design of small-scale gradient coils in magnetic resonance imaging by using the topology optimization method
    Pan, Hui
    Jia, Feng
    Liu, Zhen-Yu
    Zaitsev, Maxim
    Hennig, Juergen
    Korvink, Jan G.
    CHINESE PHYSICS B, 2018, 27 (05)
  • [9] Design of small-scale gradient coils in magnetic resonance imaging by using the topology optimization method
    潘辉
    贾峰
    刘震宇
    Maxim Zaitsev
    Juergen Hennig
    Jan G Korvink
    Chinese Physics B, 2018, (05) : 95 - 104
  • [10] Equivalent magnetic dipole method for designing gradient coils of the Halbach magnetic resonance device
    Xu, Zheng
    Qi, Jinfeng
    Guo, Pan
    Cui, Xianjun
    Wu, Jiamin
    INTERNATIONAL JOURNAL OF APPLIED ELECTROMAGNETICS AND MECHANICS, 2018, 56 (04) : 595 - 604