Clinical evaluation of the CA530-VET hematology analyzer for use in veterinary practice

被引:18
|
作者
Roleff, Sabine [1 ]
Arndt, Gisela [2 ]
Bottema, Bram [4 ]
Junker, Lothar [5 ]
Grabner, Arthur [3 ]
Kohn, Barbara [1 ]
机构
[1] Free Univ Berlin, Clin Small Anim, D-14163 Berlin, Germany
[2] Free Univ Berlin, Dept Biometr, D-14163 Berlin, Germany
[3] Free Univ Berlin, Clin Equine Med, D-14163 Berlin, Germany
[4] Boule Med AB, Stockholm, Sweden
[5] Menarini Diagnost, Neuss, Germany
关键词
CA530-VET; cat; dog; hematology analyzer; horse; method comparison;
D O I
10.1111/j.1939-165X.2007.tb00202.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background: The CA530-VET is a completely automated impedance cell hematology analyzer, which yields a 16-parameter blood count including a 3-part leukocyte differential. Objectives: The aim of this study was to examine the operational potential of the CA530-VET and its value for use in veterinary practice. Methods: The analyzer was tested for blood carry-over, precision, and accuracy. Comparison methods included the CELL-DYN 3500, microhematocrit centrifugation, manual platelet (PLT) counting for feline and equine species, and a 100-cell manual WBC differential. Blood samples for comparison of the methods were obtained from 242 dogs, 166 cats, and 144 horses. Results: The carry-over ratio (K) was 0.28% for RBC, 0.59% for PLT, 0.32% for WBC, and 0.18% for hemoglobin (HGB) concentration. Coefficients of variation (CVs) for within-batch precision and duplicate measurement of blood samples were clearly within the required limits, except for duplicate platelet counts in cats (8.7%) and horses (9.5%). The WBC count was in excellent agreement for dogs and horses and RBC count was in excellent agreement for horses. The accuracy of feline WBC counts was not acceptable, with the exception of values at the high end of the range. RBC counts in dogs and cats, and HGB concentration and MCV in all 3 species were sufficiently accurate. The CA530-VET HCT results were in excellent agreement with microhematocrit results in horses but exceeded the maximum allowed inaccuracy for cats and dogs. In all species, PLT counts established mechanically and manually were not in adequate agreement. Large differences were found between the CA530-VET and the manual differential percentage for lymphocytes and "mid-sized cells'' (monocytes and basophilic granulocytes). Conclusions: The CA530-VET can be considered useful for routine canine, feline, and equine blood cell analyses. It should not be considered accurate, however, for PLT counts, feline total WBC counts in the subnormal and normal range, and leukocyte differentials, except for granulocytes.
引用
收藏
页码:155 / 166
页数:12
相关论文
共 50 条
  • [21] CLINICAL-EVALUATION OF THE TOA SYSMEX CC-800 8 PARAMETER HEMATOLOGY ANALYZER
    GIANDOMENICO, AR
    JOURNAL OF CLINICAL HEMATOLOGY AND ONCOLOGY, 1985, 15 (01): : 9 - 19
  • [22] Clinical evaluation of the Sysmex® CA-7000 coagulation analyzer.
    Engelhardt, W
    Gringel, E
    Hafner, G
    Kretschmer, V
    Lichte, A
    Moessmer, G
    Sinha, P
    Ziemer, S
    CLINICAL CHEMISTRY, 2002, 48 (06) : A156 - A156
  • [23] Use of cytology as a diagnostic method in veterinary practice and assessment of communication between veterinary practitioners and veterinary clinical pathologists
    Christopher, Mary M.
    Hotz, Christine S.
    Shelly, Sonjia M.
    Pion, Paul D.
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2008, 232 (05): : 747 - 754
  • [24] Optimization and evaluation of Flexicult® Vet for detection, identification and antimicrobial susceptibility testing of bacterial uropathogens in small animal veterinary practice
    Guardabassi, Luca
    Hedberg, Sandra
    Jessen, Lisbeth Rem
    Damborg, Peter
    ACTA VETERINARIA SCANDINAVICA, 2015, 57
  • [25] EVALUATION OF A HEMATOLOGICAL ANALYZER FOR ITS USE IN CANINE CLINICAL PATHOLOGY
    LAVIN, S
    MORA, FJ
    MONREAL, L
    VINAS, L
    JOURNAL OF VETERINARY MEDICINE SERIES A-PHYSIOLOGY PATHOLOGY CLINICAL MEDICINE, 1991, 38 (09): : 702 - 709
  • [26] Assembly and evaluation of an inventory of guidelines that are available to support clinical hematology laboratory practice
    Hayward, C. P. M.
    Moffat, K. A.
    George, T. I.
    Proytcheva, M.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 : 36 - 45
  • [27] Optimization and evaluation of Flexicult® Vet for detection, identification and antimicrobial susceptibility testing of bacterial uropathogens in small animal veterinary practice
    Luca Guardabassi
    Sandra Hedberg
    Lisbeth Rem Jessen
    Peter Damborg
    Acta Veterinaria Scandinavica, 57
  • [28] Evaluation of ammonia measurements in dogs with two analysers for use in veterinary practice
    Sterczer, A
    Meyer, HP
    Boswijk, HC
    Rothuizen, J
    VETERINARY RECORD, 1999, 144 (19) : 523 - 526
  • [29] CLINICAL PERFORMANCE EVALUATION OF THE SAA MODULE OF MINDRAY BC-7500CS AUTOMATED HEMATOLOGY ANALYZER
    Dai, Jincheng
    Zhang, Xiaoyu
    Zhou, Jing
    Pan, Wei
    Yu, Fei
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 : 28 - 28
  • [30] ROUND-TABLE DISCUSSION ON THE CLINICAL EDUCATION, ON THE VET-CLINICS AND ON THE PRIVATE VETERINARY PRACTICE (BUDAPEST, FEBRUARY 3, 1989)
    GAL, S
    HOLLO, F
    ZSOLDOS, I
    HARASZTI, J
    BUCSY, L
    TOTH, J
    FELKAI, F
    ZSUZSA, LS
    BALINT, T
    OKROS, I
    MAGYAR ALLATORVOSOK LAPJA, 1989, 44 (12) : 762 - 764