Clinical evaluation of the CA530-VET hematology analyzer for use in veterinary practice

被引:18
|
作者
Roleff, Sabine [1 ]
Arndt, Gisela [2 ]
Bottema, Bram [4 ]
Junker, Lothar [5 ]
Grabner, Arthur [3 ]
Kohn, Barbara [1 ]
机构
[1] Free Univ Berlin, Clin Small Anim, D-14163 Berlin, Germany
[2] Free Univ Berlin, Dept Biometr, D-14163 Berlin, Germany
[3] Free Univ Berlin, Clin Equine Med, D-14163 Berlin, Germany
[4] Boule Med AB, Stockholm, Sweden
[5] Menarini Diagnost, Neuss, Germany
关键词
CA530-VET; cat; dog; hematology analyzer; horse; method comparison;
D O I
10.1111/j.1939-165X.2007.tb00202.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background: The CA530-VET is a completely automated impedance cell hematology analyzer, which yields a 16-parameter blood count including a 3-part leukocyte differential. Objectives: The aim of this study was to examine the operational potential of the CA530-VET and its value for use in veterinary practice. Methods: The analyzer was tested for blood carry-over, precision, and accuracy. Comparison methods included the CELL-DYN 3500, microhematocrit centrifugation, manual platelet (PLT) counting for feline and equine species, and a 100-cell manual WBC differential. Blood samples for comparison of the methods were obtained from 242 dogs, 166 cats, and 144 horses. Results: The carry-over ratio (K) was 0.28% for RBC, 0.59% for PLT, 0.32% for WBC, and 0.18% for hemoglobin (HGB) concentration. Coefficients of variation (CVs) for within-batch precision and duplicate measurement of blood samples were clearly within the required limits, except for duplicate platelet counts in cats (8.7%) and horses (9.5%). The WBC count was in excellent agreement for dogs and horses and RBC count was in excellent agreement for horses. The accuracy of feline WBC counts was not acceptable, with the exception of values at the high end of the range. RBC counts in dogs and cats, and HGB concentration and MCV in all 3 species were sufficiently accurate. The CA530-VET HCT results were in excellent agreement with microhematocrit results in horses but exceeded the maximum allowed inaccuracy for cats and dogs. In all species, PLT counts established mechanically and manually were not in adequate agreement. Large differences were found between the CA530-VET and the manual differential percentage for lymphocytes and "mid-sized cells'' (monocytes and basophilic granulocytes). Conclusions: The CA530-VET can be considered useful for routine canine, feline, and equine blood cell analyses. It should not be considered accurate, however, for PLT counts, feline total WBC counts in the subnormal and normal range, and leukocyte differentials, except for granulocytes.
引用
收藏
页码:155 / 166
页数:12
相关论文
共 50 条
  • [1] Evaluation of an in-house centrifugal hematology analyzer for use in veterinary practice
    Bienzle, D
    Stanton, JB
    Embry, JM
    Bush, SE
    Mahaffey, EA
    JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2000, 217 (08) : 1195 - 1200
  • [2] Evaluation of a portable clinical analyzer in a veterinary hospital setting
    Grosenbaugh, DA
    Gadawski, JE
    Muir, WW
    JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 1998, 213 (05) : 691 - 694
  • [3] Validation of the Medonic CA620/530 Vet 20-μl microcapillary sampler system for hematology testing of feline blood
    Bourges-Abella, Nathalie H.
    Reynolds, Brice S.
    Geffre, Anne
    Braun, Jean-Pierre
    Trumel, Catherine
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2009, 21 (03) : 364 - 368
  • [4] Evaluation of the Sysmex XE-2100® hematology analyzer in hospital use
    Nakul-Aquaronne, D
    Sudaka-Sammarcelli, I
    Ferrero-Vacher, C
    Starck, B
    Bayle, J
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2003, 17 (04) : 113 - 123
  • [5] Evaluation of the Mythic 18 hematology analyzer for use with canine, feline, and equine samples
    Wassmuth, Andrea K.
    Riond, Barbara
    Hofmann-Lehmann, Regina
    Lutz, Hans
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2011, 23 (03) : 436 - 453
  • [6] Clinical performance evaluation of the new hematology analyzer Mindray BC-6000
    Shen, Yimin
    Cao, Jun
    Zhou, Zhiying
    Wang, Yong
    Shen, Yili
    He, Jun
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2019, 41 (05) : 622 - 634
  • [7] Quality requirements for veterinary hematology analyzers in small animals-a survey about veterinary experts' requirements and objective evaluation of analyzer performance based on a meta-analysis of method validation studies: bench top hematology analyzer
    Cook, Andrea M.
    Moritz, Andreas
    Freeman, Kathleen P.
    Bauer, Natali
    VETERINARY CLINICAL PATHOLOGY, 2016, 45 (03) : 466 - 476
  • [8] Use of Human Intravenous Immunoglobulin in Veterinary Clinical Practice
    Spurlock, Nicole
    Prittie, Jennifer
    VETERINARY CLINICS OF NORTH AMERICA-SMALL ANIMAL PRACTICE, 2020, 50 (06) : 1371 - 1383
  • [9] Use of rivaroxaban in patients attended in a hematology unit in clinical practice
    Gavin Sebastian, Olga
    Izuzquiza Fernandez, Macarena
    Martinez Fernandez, Raquel
    Palomera Bernal, Luis
    FUTURE CARDIOLOGY, 2019, 15 (05) : 347 - 353
  • [10] A PRACTICAL EVALUATION OF THE USE OF ECG ON DOGS IN VETERINARY PRACTICE
    BOHN, FK
    TIERARZTLICHE UMSCHAU, 1993, 48 (10): : 677 - 678