Accuracy assessment for infrared camera laboratory radiometric calibration

被引:0
|
作者
Xie Xu-fen [1 ]
Zhang Wei [1 ]
Nie Hong-bin [1 ]
Cao Yi-ming [1 ]
Wang Qiang [1 ]
Wang Hong-yuan [1 ]
机构
[1] Harbin Inst Technol, Res Ctr Space Opt Engn, Harbin 150080, Peoples R China
关键词
radiometric calibration; uncertainty; imaging chain; TM;
D O I
10.1117/12.900566
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
An accuracy assessment method of infrared camera laboratory radiometric calibration was studied for the sake of validation of space infrared camera measured data. Firstly, image process of infrared camera was analyzed and modeled on laboratory radiometric calibration, a model of linear radiometric calibration coefficient synthesized impact chain was built; secondly, based on the model, a model of uncertainty of linear radiometric calibration coefficient was built; finally, An experiment verified the validity of the model. The results of experiment indicate that the difference between the assessment value and maximum experiment value is 0.08% and the difference between the assessment value and average of experiment values is 0.79% at gain uncertainty assessment, the difference between the assessment value and maximum experiment value is 0.7%, the difference between the assessment value and average of experiment values is 0.18% at offset uncertainty assessment, the uncertainty of coefficients of radiometric calibration get from experiments is basically consistent with the assessment. The method of accuracy assessment of radiometric calibration combined with chain factors uncertainty can avoid omitting and repetition in uncertainty estimation; optimization of camera used for quantification can be designed on the method.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Camera Response Estimation. Radiometric Calibration
    Negru, Mihai
    Nedevschi, Sergiu
    2009 IEEE 5TH INTERNATIONAL CONFERENCE ON INTELLIGENT COMPUTER COMMUNICATION AND PROCESSING, PROCEEDINGS, 2009, : 103 - 109
  • [22] Simplified method for radiometric calibration of an array camera
    Cano-Garcia, Angel
    Lazaro, Jose Luis
    Fernandez, Pedro R.
    2007 IEEE INTERNATIONAL SYMPOSIUM ON INTELLIGENT SIGNAL PROCESSING, CONFERENCE PROCEEDINGS BOOK, 2007, : 401 - +
  • [23] A novel radiometric calibration of an infrared thermometer
    Yamada, Yoshiro
    Harris, Subrena
    Theocharous, Evangelos
    Hayes, Michael
    METROLOGIA, 2023, 60 (06)
  • [24] Radiometric calibration of SeaWiFS in the near infrared
    Martiny, N
    Frouin, R
    Santer, R
    APPLIED OPTICS, 2005, 44 (36) : 7828 - 7844
  • [25] Radiometric Calibration Method for Infrared Camera Based on Power-weighted Least Squares Fitting
    Jin Zhanlei
    Jin Libing
    Zhang Jiushuang
    Xu Lina
    Bao Yunfei
    Li Yan
    ACTA PHOTONICA SINICA, 2024, 53 (04)
  • [26] CAMERA CALIBRATION LABORATORY AT NRC
    CARMAN, PD
    PHOTOGRAMMETRIC ENGINEERING, 1969, 35 (04): : 372 - &
  • [27] Accuracy Verification of On-Board Radiometric Calibration
    Li Xin
    Zhang Liming
    Si Xiaolong
    Chen Hongyao
    Huang Wenxin
    Song Qingjun
    ACTA OPTICA SINICA, 2020, 40 (09)
  • [28] Research on the radiometric calibration of TDI CCD remote camera
    Liu, YX
    Hao, ZH
    ICO20: ILLUMINATION, RADIATION, AND COLOR TECHNOLOGIES, 2006, 6033
  • [29] Radiometric calibration of the Scripps Earth Polychromatic Imaging Camera
    Early, EA
    Bush, BC
    Brown, SW
    Allen, DW
    Johnson, BC
    EARTH OBSERVING SYSTEMS VI, 2002, 4483 : 77 - 84
  • [30] Radiometric Calibration of Spatial⁃Spectral Coupling Multispectral Camera
    Wu, Xuan
    Yang, Bin
    Li, Xianfeng
    Zhang, Junqiang
    Lu, Tianjiao
    ACTA OPTICA SINICA, 2024, 44 (23)