Detection of nucleated red blood cells using the Mindray BC-6800Plus hematology analyzer: a clinical performance evaluation

被引:5
|
作者
Wang, Nengyong [1 ]
Jiang, Junyao [1 ]
Wang, Ji'an [1 ]
He, Yao [1 ]
Wen, Ning [1 ]
Liu, Yulin [1 ]
Wang, Yashu [1 ]
Li, Haijun [1 ]
Shi, Ping [1 ]
机构
[1] Guangyuan Cent Hosp, Lab Dept, Guangyuan 628017, Peoples R China
关键词
Hematology analyzer; nucleated red blood; parameter; manual microscopic examination;
D O I
10.21037/apm-21-1772
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: With the development of instrument technology, the functions and detection methods of automatic blood cell analyzers have become more complex. To ensure optimal clinical applicability, it is crucial to select an automatic blood cell analyzer with excellent clinical detection performance. This study evaluated the latest Mindray BC-6800Plus automatic blood cell analyzer and assessed its performance in the detection of nucleated red blood cells (NRBCs). Methods: A total of 490 clinical blood samples were used to assess the performance of the instrument, including parameters such as precision, linearity, conformity rate of manual microscopic examination, carryover, and limit of quantitation. Results: The instrument showed a small carryover (<= 0.02) and excellent linearity (R-2 >= 0.9986). The reproducibility of the sample tests was satisfactory, and the coefficient of variation (CV) of the test results [0.98-1.72% and 0.62-6.97% for white blood cells (WBCs) and NRBCs, respectively] were significantly lower than that declared by the manufacturer (2.5% and 20% for WBCs and NRBCs, respectively). Thus, the BC-6800Plus satisfies the requirements of clinical testing. Two separate Mindray BC-6800Plus machines were tested and found to be in good agreement with each other and with manual microscopy methods. Furthermore, WBC and NRBC counts were highly consistent with results obtained using the XN-9100 blood analyzer. Conclusions: The Mindray BC-6800Plus is an excellent analyzer that can provide timely and accurate reports for clinical laboratory detection of NRBC.
引用
收藏
页码:8808 / 8817
页数:10
相关论文
共 50 条
  • [41] Integration of an MC-80 Digital Image Analyzer With an Automated BC-6800Plus Hematology Analyzer Enables Accurate Platelet Counting in Samples With EDTA-Induced Pseudothrombocytopenia
    So, Min-Kyung
    Huh, Jungwon
    Kim, Seunghwan
    Park, Sholhui
    ANNALS OF LABORATORY MEDICINE, 2024, 44 (06) : 478 - 486
  • [42] DISTINGUISHING NUCLEATED RED BLOOD CELLS AND BASOPHILS FROM OTHER LEUKOCYTES SYSMEX HEMATOLOGY ANALYZER SCATTERGRAM
    Kaido, Masako
    Takagi, Yuri
    Kono, Mari
    Nakazawa, Fumie
    Yamamoto, Shiori
    Wada, Atsushi
    Morikawa, Takashi
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 51 - 51
  • [43] Earlier detection of sepsis by Candida parapsilosis using three-dimensional cytographic anomalies on the Mindray BC-6800 hematological analyzer
    La Gioia, Antonio
    Bombara, Maria
    Fiorini, Fabiana
    Dell'Amico, Mariachiara
    Devito, Alessandra
    Isola, Patrizia
    Marelli, Paola
    Fiorini, Marcello
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (08) : E239 - E242
  • [44] Performance evaluation of optical platelet counting of BC-6000Plus automated hematology analyzer
    Hu, Xuehua
    Tang, Yuqun
    Li, Mingyi
    Fu, Chunyun
    Deng, Zuoxin
    Tang, Wenting
    Zhou, Huideng
    Chen, Yan
    Li, Yingfeng
    ANNALS OF TRANSLATIONAL MEDICINE, 2021, 9 (06)
  • [45] EVALUATION OF THE DETECTION OF CANCER CELLS IN BODY FLUIDS BY THE INSTRUMENTS MINDRAY 6800-PLUS AND SYSMEX X-20
    Cebreiros-Lopez, Iria
    Caparros-Guerrero, Marina
    Anton-Martinez, David
    Simon-Lopez, Ramon
    Antonio Noguera-Velasco, Jose
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 : 30 - 31
  • [46] Nucleated red blood cells in term fetuses: Reference values using an automated analyzer
    Rolfo, Alessandro
    Maconi, Mariacaterina
    Cardaropoli, Simona
    Biolcati, Marilisa
    Danise, Paolo
    Todros, Tullia
    NEONATOLOGY, 2007, 92 (03) : 205 - 208
  • [47] Performance evaluation of the measurement of complete blood count parameters between Mindray BC 6000 and BT pro 2401 hematology analyzers
    Kader, S.
    CLINICA CHIMICA ACTA, 2022, 530 : S285 - S285
  • [48] The possibility of automatic capillary blood testing in routine blood tests: an evaluation of the automatic mode of the Mindray BC-7500 CRP Auto Hematology Analyzer for capillary blood testing
    He, Jiajing
    Zhang, Ge
    Wang, Yuefang
    Yang, Hao
    Dai, Qingkai
    Guo, Siqi
    Mai, Jia
    CARDIOVASCULAR DIAGNOSIS AND THERAPY, 2023, 13 (03) : 465 - 473
  • [49] Usefulness of automatic detection of fragmented red cells using a hematology analyzer for diagnosis of thrombotic microangiopathy
    Saigo, K
    Jiang, M
    Tanaka, C
    Fujimoto, K
    Kobayashi, A
    Nozu, K
    Iijima, K
    Ryo, R
    Sugimoto, T
    Imoto, S
    Kumagai, S
    CLINICAL AND LABORATORY HAEMATOLOGY, 2002, 24 (06): : 347 - 351
  • [50] Evaluation of the nucleated red blood cell count in neonates using the Beckman Coulter UniCel DxH 800 analyzer
    Kwon, M. -J.
    Nam, M. -H.
    Kim, S. H.
    Lim, C. S.
    Lee, C. K.
    Cho, Y.
    Lee, K. N.
    Yoon, S. -Y.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2011, 33 (06) : 620 - 628