Limited accuracy of dose calculation for large fields at deep depths using the BrainSCAN v5.21 treatment-planning system

被引:1
|
作者
Hsi, Wen C.
Zhang, Yunkai
Kirk, Michael C.
Bernard, Damian
Chu, James C. H.
机构
[1] Rush Univ, Med Ctr, Dept Phys Med, Chicago, IL 60612 USA
[2] Rush Univ, Med Ctr, Dept Radiat Oncol, Chicago, IL 60612 USA
来源
关键词
radiation treatment-planning system; BrainSCAN; BrainLAB; quality assurance;
D O I
10.1120/jacmp.2024.25333
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Varian 120 multileaf collimator (MLC) has a leaf thickness of 5 mm projected at the isocenter plane and can deliver a radiation beam of large field size ( up to 30 cm) to be used in intensity-modulated radiotherapy (IMRT). Often the dose must be delivered to depths greater than 20 cm. Therefore, during the commissioning of the BrainSCAN v5.21 or any radiation treatment-planning (RTP) systems, extensive testing of dose and monitor unit calculations must encompass the field sizes ( 1 cm to 30 cm) and the prescription depths ( 1 cm to 20 cm). Accordingly, the central-axis percent depth doses (PDDs) and off-axis percentage profiles must be measured at several depths for various field sizes. The data for this study were acquired with a 6-MV X-ray beam from a Varian 2100EX LINAC with a water phantom at a source-to-surface distance (SSD) of 100 cm. These measurements were also used to generate a photon beam module, based on a photon pencil beam dose-calculation algorithm with a fast-Fourier transform method. To commission the photon beam module used in our BrainSCAN RTP system, we performed a quantitative comparison of measured and calculated central-axis depth doses and off-axis profiles. Utilizing the principles of dose difference and distance-to-agreement introduced by Van Dyk et al. [ Commissioning and quality assurance of treatment planning computers. Int J Radiat Oncol Biol Phys. 1993; 26: 261- - 273], agreements between calculated and measured doses are < 2% and < 2 mm for the regions of low- and high-dose gradients, respectively. However, large errors ( up to similar to 5% and similar to 7% for 20-cm and 30-cm fields, respectively, at the depth 20 cm) were observed for monitor unit calculations. For a given field size, the disagreement increased with the depth. Similarly, for a given depth the disagreement also increased with the field size. These large systematic errors were caused by using the tissue maximum ratio (TMR) in BrainSCAN v5.21 without considering increased field size as depth increased. These errors have been reported to BrainLAB.
引用
收藏
页码:12 / 18
页数:7
相关论文
共 50 条
  • [41] Small Static Fields: Real-Time Assessment of the Accuracy of a Treatment Planning System (TPS) Calculated Dose with a Step-And-Shoot Method
    Biasi, G.
    Hardcastle, N.
    Petasecca, M.
    Guatelli, S.
    Rozenfeld, A.
    Kron, T.
    MEDICAL PHYSICS, 2018, 45 (06) : E620 - E620
  • [42] Accuracy of the Small Field Dosimetry Using Acuros XB and AAA Dose Calculation Algorithms of Eclipse Treatment Planning System Within and Beyond Heterogeneous Media for Trubeam 2.0 Unit
    Codel, G.
    Serin, E.
    Pacaci, P.
    Sanli, E.
    Cebe, M.
    Mabhouti, H.
    Doyuran, M.
    Kucukmorkoc, E.
    Kucuk, N.
    Altinok, A.
    Canoglu, D.
    Acar, H.
    Ozkok, H. Caglar
    MEDICAL PHYSICS, 2016, 43 (06) : 3350 - 3350
  • [43] Delivery Accuracy of Stereotactic Radiosurgery with Tomotherapy Using Treatment Planning System Version V4.0 and the Recent Upgrade V4.2
    Thakur, V.
    Ruo, R.
    Seuntjens, J.
    Parker, W.
    Doucet, R.
    Soisson, E.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [44] Verification of the Accuracy of the Proton Range Compensator Scatter Factor and Dose within Inhomogeneity Calculations Using the Eclipse Treatment Planning System
    Pollard, J.
    Poenisch, F.
    Sahoo, N.
    Zhu, X.
    Gillin, M.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [45] Study the sensitivity of Dose Calculation in Prism Treatment Planning System Using Monte Carlo Simulation of 6 MeV Electron Beam
    Hardiansyah, D.
    Male, S.
    Haryanto, F.
    4TH INTERNATIONAL CONFERENCE ON ADVANCES IN NUCLEAR SCIENCE AND ENGINEERING (ICANSE 2013), 2014, 1615 : 213 - 223
  • [46] Accuracy of Dose Attenuation Correction for a 6D Carbon Fiber Treatment Couch Using a Virtual Couch Technique Integrated Into a Treatment Planning System
    Shioiri, K.
    Kurooka, M.
    Yoshino, S.
    Maehana, W.
    Itou, M.
    Kusano, Y.
    Ide, S.
    Onodera, M.
    Nonaka, T.
    Nakayama, Y.
    MEDICAL PHYSICS, 2012, 39 (06) : 3758 - 3758
  • [47] The accuracy of out of field dose calculations in commercial treatment planning system using GATE/GEANT4 Monte Carlo simulation
    Attalla, Ehab Marouf
    Sinousy, Dina Mohamed
    Ibrahim, Hassan Fathy
    Elhussiny, Fathi Ahmed
    Elmekawy, Ahmed Farouk
    RADIATION PHYSICS AND CHEMISTRY, 2023, 206
  • [48] Calculation of dose distribution by means of Monte Carlo and treatment planning system using identical incidence fluence maps for megavoltage photon beams
    Spirydovich, S
    Papiez, L
    Thai, V
    Langer, M
    MEDICAL PHYSICS, 2004, 31 (06) : 1867 - 1867
  • [49] Accuracy of the Small Field Dosimetry Using the Monte Carlo and Sequential Dose Calculation Algorithms of Multiplan Treatment Planning System Within and Beyond Heterogeneous Media for Cyberknife M6 Unit
    Serin, E.
    Codel, G.
    Mabhouti, H.
    Cebe, M.
    Sanli, E.
    Pacaci, P.
    Kucuk, N.
    Kucukmorkoc, E.
    Doyuran, M.
    Canoglu, D.
    Altinok, A.
    Acar, H.
    Ozkok, H. Caglar
    MEDICAL PHYSICS, 2016, 43 (06) : 3350 - 3350
  • [50] Improved dose-calculation accuracy in proton treatment planning using a simplified Monte Carlo method verified with three-dimensional measurements in an anthropomorphic phantom
    Hotta, Kenji
    Kohno, Ryosuke
    Takada, Yoshihisa
    Hara, Yousuke
    Tansho, Ryohei
    Himukai, Takeshi
    Kameoka, Satoru
    Matsuura, Taeko
    Nishio, Teiji
    Ogino, Takashi
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12): : 3545 - 3556