Optimized RF shielding techniques for simultaneous PET/MR

被引:9
|
作者
Truhn, D. [1 ,2 ]
Kiessling, F. [1 ,2 ]
Schulz, V. [1 ,2 ]
机构
[1] Univ Aachen RWTH, Inst Expt Mol Imaging, D-52074 Aachen, Germany
[2] Philips Res Europe Aachen, D-52066 Aachen, Germany
关键词
magnetic resonance imaging (MRI); positron emission tomography (PET); magnetic shielding; eddy currents; MR-gradient;
D O I
10.1118/1.3596532
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Integration of positron emission tomography (PET) and magnetic resonance (MR) in a merged device requires shielding of the PET detector electronics. The authors demonstrate that a multiple shell setup exhibits advantages over a single shell shielding enclosure. Methods: A spherical shell model is used to derive analytical results. Numerical computations of the electromagnetic fields with a commercial software package (FEKO) and experiments are used to confirm our findings. Results: Replacing a single shell enclosure by a multilayer approach barely changes its behavior toward low frequency magnetic fields (MR gradient fields), while improving shielding at high frequencies (proton resonant frequency) by orders of magnitude. Conclusions: For a given required shielding factor at the proton resonant frequency, the eddy currents induced by the MR gradient fields may be reduced by employing a multiple shell setup. (C) 2011 American Association of Physicists in Medicine. [DOI:10.1118/1.3596532]
引用
收藏
页码:3995 / 4000
页数:6
相关论文
共 50 条
  • [31] Assessment of artifacts in simultaneous PET and MR imaging
    Slates, R
    Shao, Y
    Farahani, K
    Marsden, PK
    Cherry, SR
    Meadors, K
    Taylor, J
    Summers, PE
    Williams, S
    Beech, J
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 1357 - 1360
  • [32] Innovative technologies for simultaneous TOF PET/MR
    Schulz, V.
    Solf, T.
    Weissler, B.
    Gebhardt, P.
    Piemonte, C.
    Fischer, P.
    Ritzert, M.
    Keereman, V.
    Vandenberghe, S.
    Tsoumpas, C.
    Marsden, P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 : S219 - S220
  • [33] Simultaneous Reconstruction of Activity and Attenuation for PET/MR
    Salomon, Andre
    Goedicke, Andreas
    Schweizer, Bernd
    Aach, Til
    Schulz, Volkmar
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2011, 30 (03) : 804 - 813
  • [34] Quantitative Simultaneous PET-MR Imaging
    Ouyang, Jinsong
    Petibon, Yoann
    Huang, Chuan
    Reese, Timothy G.
    Kolnick, Aleksandra L.
    El Fakhri, Georges
    MICRO- AND NANOTECHNOLOGY SENSORS, SYSTEMS, AND APPLICATIONS VI, 2014, 9083
  • [35] Hybrid MR-guided and PET-guided motion correction of PET images in simultaneous PET/MR
    Feng, Tao
    Ahlman, Mark
    Tsui, Benjamin
    Guo, Liheng
    Guttman, Mike
    McVeigh, Elliot
    Bluemke, David
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [36] RF-Transmissive PET Detector Insert for Simultaneous PET/MRI
    Lee, Brian J.
    Grant, Alexander M.
    Chang, Chen-Ming
    Glover, Gary H.
    Levin, Craig S.
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [37] A study of artefacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner
    Slates, RB
    Farahani, K
    Shao, YP
    Marsden, PK
    Taylor, J
    Summers, PE
    Williams, S
    Beech, J
    Cherry, SR
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (08): : 2015 - 2027
  • [38] Evaluation of Two PET Motion Correction Techniques For Simultaneous Real-Time PET-MR Acquisitions Using An MR-Derived Motion Model
    Tsoumpas, Charalampos
    Agarwal, Siddharth
    Marsden, Paul K.
    King, Andrew P.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2519 - 2522
  • [39] Fast tagged MR acquisition for PET motion correction in simultaneous cardiac PET-MR
    Huang, Chuan
    Petibon, Yoann
    Ouyang, Jinsong
    El Fakhri, Georges
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [40] ML and MAP PET reconstruction with MR-voxel sizes for simultaneous PET-MR
    Belzunce, Martin A.
    Mehranian, Abolfazl
    Bland, James
    Reader, Andrew J.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,