Including robustness in multi-criteria optimization for intensity-modulated proton therapy

被引:146
|
作者
Chen, Wei [1 ]
Unkelbach, Jan
Trofimov, Alexei
Madden, Thomas
Kooy, Hanne
Bortfeld, Thomas
Craft, David
机构
[1] Massachusetts Gen Hosp, Dept Radiat Oncol, Boston, MA 02114 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2012年 / 57卷 / 03期
关键词
TREATMENT UNCERTAINTIES; RANGE UNCERTAINTIES; RADIOTHERAPY; SENSITIVITY; ALGORITHM; PLAN;
D O I
10.1088/0031-9155/57/3/591
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present a method to include robustness in a multi-criteria optimization (MCO) framework for intensity-modulated proton therapy (IMPT). The approach allows one to simultaneously explore the trade-off between different objectives as well as the trade-off between robustness and nominal plan quality. In MCO, a database of plans each emphasizing different treatment planning objectives, is pre-computed to approximate the Pareto surface. An IMPT treatment plan that strikes the best balance between the different objectives can be selected by navigating on the Pareto surface. In our approach, robustness is integrated into MCO by adding robustified objectives and constraints to the MCO problem. Uncertainties (or errors) of the robust problem are modeled by pre-calculated dose-influence matrices for a nominal scenario and a number of pre-defined error scenarios (shifted patient positions, proton beam undershoot and overshoot). Objectives and constraints can be defined for the nominal scenario, thus characterizing nominal plan quality. A robustified objective represents the worst objective function value that can be realized for any of the error scenarios and thus provides a measure of plan robustness. The optimization method is based on a linear projection solver and is capable of handling large problem sizes resulting from a fine dose grid resolution, many scenarios, and a large number of proton pencil beams. A base-of-skull case is used to demonstrate the robust optimization method. It is demonstrated that the robust optimization method reduces the sensitivity of the treatment plan to setup and range errors to a degree that is not achieved by a safety margin approach. A chordoma case is analyzed in more detail to demonstrate the involved trade-offs between target underdose and brainstem sparing as well as robustness and nominal plan quality. The latter illustrates the advantage of MCO in the context of robust planning. For all cases examined, the robust optimization for each Pareto optimal plan takes less than 5 min on a standard computer, making a computationally friendly interface possible to the planner. In conclusion, the uncertainty pertinent to the IMPT procedure can be reduced during treatment planning by optimizing plans that emphasize different treatment objectives, including robustness, and then interactively seeking for a most-preferred one from the solution Pareto surface.
引用
收藏
页码:591 / 608
页数:18
相关论文
共 50 条
  • [21] Comparison of intensity-modulated proton therapy and intensity-modulated photon therapy in treatment of nasopharyngeal carcinoma
    Taheri-Kadkhoda, Z.
    Nill, S.
    Wilkens, J.
    Oelfke, U.
    Bjoerk-Eriksson, T.
    Huber, P.
    Munter, M. W.
    RADIOTHERAPY AND ONCOLOGY, 2007, 82 : S52 - S53
  • [22] Improved Multi-Criteria Optimization for Intensity Modulated Proton Therapy Using Iterative Resampling of Randomly Placed Pencil-Beams
    van de Water, S.
    Kraan, A.
    Breedveld, S.
    Teguh, D.
    Madden, T.
    Kooy, H.
    Heijmen, B.
    Hoogeman, M.
    MEDICAL PHYSICS, 2012, 39 (06) : 3981 - 3982
  • [23] Region-Specific Robust Optimization for Intensity-Modulated Proton Therapy
    Liao, L.
    Chen, M.
    Adair, A.
    Zhu, X.
    Cao, W.
    Sahoo, N.
    Yang, M.
    Gunn, B.
    Frank, S.
    Zhang, X.
    MEDICAL PHYSICS, 2020, 47 (06) : E481 - E481
  • [24] Proton Therapy for Nasopharyngeal Cancer: A Matched Case-control Study of Intensity-Modulated Proton Therapy and Intensity-Modulated Photon Therapy
    Li, X.
    Lee, A.
    Lee, N.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 106 (05): : 1138 - 1139
  • [25] Optimization of the Irradiation Efficiency of intensity-modulated Proton Therapy Plans through prioritized Optimization
    Mueller, B. S.
    Wilkens, J. J.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 : 80 - 80
  • [26] Improvement in Robustness and Delivery Efficiency of Intensity-Modulated Proton Therapy Using Single-Field Optimization with Energy Absorber
    Quan, E.
    Liu, W.
    Li, Y.
    Zhang, X.
    Zhu, X.
    Mohan, R.
    Wu, R.
    MEDICAL PHYSICS, 2012, 39 (06) : 3965 - 3965
  • [27] Potential for Improvements in Robustness and Optimality of Intensity-Modulated Proton Therapy for Lung Cancer with 4-Dimensional Robust Optimization
    Ge, Shuaiping
    Wang, Xiaochun
    Liao, Zhongxing
    Zhang, Lifei
    Sahoo, Narayan
    Yang, Jinzhong
    Guan, Fada
    Mohan, Radhe
    CANCERS, 2019, 11 (01):
  • [28] Evaluation of robustness of optimization methods in breast intensity-modulated radiation therapy using TomoTherapy
    Oki, Yuya
    Akasaka, Hiroaki
    Uehara, Kazuyuki
    Mizonobe, Kazufusa
    Sawada, Masanobu
    Nagata, Junya
    Harada, Aya
    Mayahara, Hiroshi
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (02) : 465 - 475
  • [29] Fraction-variant beam orientation optimization for intensity-modulated proton therapy
    Gu, Wenbo
    O'Connor, Daniel
    Ruan, Dan
    Zou, Wei
    Dong, Lei
    Sheng, Ke
    MEDICAL PHYSICS, 2020, 47 (09) : 3826 - 3834
  • [30] Robust Optimization in Intensity-Modulated Proton Therapy for Base-Of-Skull Cancers
    Liu, W.
    Mohan, R.
    Zhu, X.
    Grosshans, D.
    MEDICAL PHYSICS, 2013, 40 (06)