Accidental embryo irradiation during barium enema examinations - An estimation of absorbed dose

被引:18
|
作者
Damilakis, J
Perisinakis, K
Grammatikakis, J
Panayiotakis, G
Gourtsoyiannis, N
机构
[1] UNIV HOSP IRAKLION,DEPT MED PHYS,IRAKLION,CRETE,GREECE
[2] UNIV HOSP IRAKLION,DEPT RADIOL,IRAKLION,CRETE,GREECE
[3] UNIV PATRAS,DEPT MED PHYS,PATRAI,GREECE
关键词
fetus; pregnancy; dosimetry; radiation;
D O I
10.1097/00004424-199604000-00009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES To investigate the possibility of an embryo to receive a dose of more than 10 cGy, the threshold of malformations induction in embryos reported by the International Commission on Radiological Protection, during barium enema examinations. METHODS. Thermoluminescent dosimeters were placed in a phantom to calculate the depth-to-skin conversion coefficient needed for dose estimation at the average embryo depth in patients, Barium enema examinations were performed in 20 women of childbearing age with diagnostic problems demanding longer fluoroscopy times, Doses at 6 cm, the average embryo depth, were determined by measurements at the patients' skin followed by dose calculation at the site of interest. RESULTS. The range of doses estimated at embryo depth for patients was 1.9 to 8.1 cGy, The dose always exceeded 5 cGy when fluoroscopy time was longer than 7 minutes. CONCLUSION. The dose at the embryo depth never exceeded 10 cGy, This study indicates that fluoroscopy time should not exceed 7 minutes in childbearing-age female patients undergoing barium enema examinations.
引用
收藏
页码:242 / 245
页数:4
相关论文
共 45 条
  • [1] In vivo measurement method of ovarian dose during barium enema examinations
    Lavoie, C
    Don, C
    BRITISH JOURNAL OF RADIOLOGY, 1997, 70 : 291 - 295
  • [2] Dose optimisation of double-contrast barium enema examinations
    Berner, K.
    Bath, M.
    Jonasson, P.
    Cappelen-Smith, J.
    Fogelstam, P.
    Soderberg, J.
    RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) : 388 - 392
  • [3] ANAPHYLAXIS ASSOCIATED WITH LATEX ALLERGY DURING BARIUM ENEMA EXAMINATIONS
    OWNBY, DR
    TOMLANOVICH, M
    SAMMONS, N
    MCCULLOUGH, J
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 156 (05) : 903 - 908
  • [4] Patient dose variation investigated in four Irish hospitals for barium meal and barium enema examinations
    Carroll, E
    Brennan, PC
    RADIATION PROTECTION DOSIMETRY, 2001, 97 (03) : 275 - 278
  • [5] Dose-area product measurements during barium enema radiograph examinations - A western cape study
    Engel-Hills, PC
    Hering, ER
    SOUTH AFRICAN MEDICAL JOURNAL, 2001, 91 (08): : 693 - 696
  • [6] Radiation dose and diagnosticity of barium enema examinations by radiographers and radiologists: a comparative study
    Crawley, MT
    Shine, B
    Booth, A
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (844): : 399 - 405
  • [7] ACCIDENTAL FILLING OF URINARY TRACT DURING BARIUM ENEMA - REPORT OF A CASE
    PERRY, SP
    HADEN, WD
    RADIOLOGY, 1948, 50 (04) : 539 - 540
  • [8] The secondary radiation grid its effect on fluoroscopic dose-area product during barium enema examinations
    Lloyd, P
    Lowe, D
    Harty, DS
    Eyes, B
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (843): : 303 - 306
  • [9] Radiation dose and diagnosticity of barium enema examinations by radiographers and radiologists: a comparative study (3)
    Hetherington, J
    Chapple, CL
    Rawlings, DJ
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (851): : 1219 - 1221
  • [10] Patient specific doses used to analyse the optimum dose delivery in barium enema examinations
    Lampinen, JS