Noise characteristics of x-ray differential phase contrast CT

被引:5
|
作者
Zambelli, Joseph [1 ]
Li, Ke [1 ]
Bevins, Nicholas [1 ]
Qi, Zhihua [1 ]
Chen, Guang-Hong [1 ]
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53705 USA
关键词
CONE-BEAM CT; TOMOGRAPHY;
D O I
10.1117/12.878497
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The noise characteristics of x-ray differential phase contrast computed tomography (DPC-CT) were investigated. Both theoretical derivation and experimental results demonstrated that the dependence of noise variance on spatial resolution in DPC-CT follows an inverse linear law. This behavior distinguishes DPC-CT from conventional absorption based x-ray CT, where the noise variance varies inversely with the cube of the spatial resolution. This anomalous noise behavior in DPC-CT is due to the Hilbert filtering kernel used in the CT reconstruction algorithm, which equally weights all spatial frequency content. Additionally, we demonstrate that the noise power of DPC-CT is scaled by the inverse of spatial frequency and is highly concentrated at the low spatial frequencies, whereas conventional absorption CT increases in power at the high spatial frequencies.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] New types of X-ray computed tomography (CT) with synchrotron radiation: Fluorescent X-ray CT and phase-contrast X-ray CT using interferometer
    Takeda, T
    Momose, A
    Yu, QW
    Yuasa, T
    Dilmanian, FA
    Akatsuka, T
    Itai, Y
    CELLULAR AND MOLECULAR BIOLOGY, 2000, 46 (06) : 1077 - 1088
  • [22] Progress in biomedical application of phase-contrast x-ray imaging and fluorescent x-ray CT
    Takeda, Tohoru
    Wu, Jin
    Lwin, Thet-Thet
    Yoneyama, Akio
    Hirai, Yasuharu
    Hyodo, Kazuyuki
    Sunaguchi, Naoki
    Yuasa, Tetsuya
    Minami, Manabu
    Kose, Katsumi
    Akatsuka, Takao
    DEVELOPMENTS IN X-RAY TOMOGRAPHY V, 2006, 6318
  • [23] Differential phase contrast x-ray microimaging with scanning-imaging x-ray microscope optics
    Takeuchi, Akihisa
    Suzuki, Yoshio
    Uesugi, Kentaro
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (08):
  • [24] Line integral retrieval and reconstruction for X-ray fan-beam differential phase contrast CT
    Fu, Jian
    Liu, Zhen-zhong
    2015 FIFTH INTERNATIONAL CONFERENCE ON INSTRUMENTATION AND MEASUREMENT, COMPUTER, COMMUNICATION AND CONTROL (IMCCC), 2015, : 1348 - 1351
  • [25] On the origin of MTF reduction in grating-based x-ray differential phase contrast CT imaging
    Tan, Yuhang
    Yang, Jiecheng
    Zheng, Hairong
    Liang, Dong
    Zhu, Peiping
    Ge, Yongshuai
    MEDICAL PHYSICS, 2025, 52 (03) : 1546 - 1555
  • [26] Image Fusion in X-Ray Differential Phase-Contrast Imaging
    Haas, W.
    Polyanskaya, M.
    Bayer, F.
    Goedel, K.
    Hofmann, H.
    Rieger, J.
    Ritter, A.
    Weber, T.
    Wucherer, L.
    Durst, J.
    Michel, T.
    Anton, G.
    Hornegger, J.
    MEDICAL IMAGING 2012: IMAGE PROCESSING, 2012, 8314
  • [27] Differential x-ray phase contrast imaging using a shearing interferometer
    David, C
    Nöhammer, B
    Solak, HH
    Ziegler, E
    APPLIED PHYSICS LETTERS, 2002, 81 (17) : 3287 - 3289
  • [28] A dual detector approach for X-ray differential phase contrast imaging
    Wang, Dajiang
    Wang, Zhili
    Gao, Kun
    Ge, Xin
    Wu, Zhao
    Zhu, Peiping
    Wu, Ziyu
    RADIATION PHYSICS AND CHEMISTRY, 2014, 95 : 86 - 90
  • [29] Design and construction of an X-ray phase contrast CT system at BSRF
    Wang, Junyue
    Zhu, Peiping
    Yuan, Qingxi
    Huang, Wanxia
    Shu, Hang
    Hu, Tiandou
    Wu, Ziyu
    RADIATION PHYSICS AND CHEMISTRY, 2006, 75 (11) : 1986 - 1989
  • [30] Helical differential X-ray phase-contrast computed tomography
    Fu, Jian
    Willner, Marian
    Chen, Liyuan
    Tan, Renbo
    Achterhold, Klaus
    Bech, Martin
    Herzen, Julia
    Kunka, Danays
    Mohr, Juergen
    Pfeiffer, Franz
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (03): : 374 - 379