Measurement of 25-hydroxyvitamin D3 (25OHD3) and 25-hydroxyvitamin D2 (25OHD2) in human serum using liquid chromatography-tandem mass spectrometry and its comparison to a radioimmunoassay method

被引:100
|
作者
Chen, Huiping [1 ]
McCoy, Leslie F. [1 ,2 ]
Schleicher, Rosemary L. [1 ]
Pfeiffer, Christine M. [1 ]
机构
[1] Ctr Dis Control & Prevent, Atlanta, GA USA
[2] Battelle Mem Inst, Atlanta, GA USA
关键词
25-hydroxyvitamin D; HPLC; LC-MS/MS; radioimmunoassay; RIA; method comparison;
D O I
10.1016/j.cca.2008.01.017
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Measurement of vitamin D molecules are important in the management of patients with bone disease. We developed a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method to measure 25OHD(3) and 25OHD(2) in human serum and compared it to the traditionally used DiaSorin radioimmunoassay (RIA). Methods: Serum samples (200 mu l) were treated with acetonitrile and centrifuged to remove protein. An online solid-phase extraction procedure was used. Calibration solutions (5-100 ng/ml) of 25OHD(2) and 25OHD(3) were prepared using 4% albumin in phosphate-buffered saline. Chromatography: C 18 column, isocratic ethanol/water (83/17, v/v). Mass spectrometry system: atmospheric pressure chemical ionization in the positive ion mode. Transitions: 250HD3, m/z 401.4 -> 383.4; 25OHD(2), m/z 413.4 -> 395.4; and the internal standard hexadeuterated-25OHD(3), m/z 407.7 -> 389.7. Results: Detection limits were 0.49 ng/ml (25OHD(3)) and 1.86 ng/ml (25OHD(2)). Intra- and inter-assay coefficients of variation (CV) were <7% and <11%, respectively, for 250HD3 and <9% and <16%, respectively, for 25OHD(2). Recovery averaged (SD) 99% (2%) for 25OHD(3) and 95% (0.8%) for 25OHD(2). In a method comparison of 551 specimens from the National Health and Nutrition Examination Survey, the LC-MS/MS method gave values that were on average 13% higher (95%CI: 11-15%) than RIA results. Conclusions: This high throughput candidate reference method requires minimal sample preparation and is suitable for routine use for analysis of vitamin D status. (C) 2008 Published by Elsevier B.V.
引用
收藏
页码:6 / 12
页数:7
相关论文
共 50 条
  • [31] Routine simultaneous measurement of 25-hydroxyvitamin D2 and D3 by isotope-dilution liquid chromatography-tandem mass spectrometry assay
    Gong, Y.
    Wan, B.
    Walsh, W.
    Verjee, Z.
    Adeli, K.
    CLINICAL BIOCHEMISTRY, 2008, 41 (14-15) : 1268 - 1268
  • [32] Quantification of circulating 25-hydroxyvitamin D by liquid chromatography-tandem mass spectrometry
    Vogeser, Michael
    JOURNAL OF STEROID BIOCHEMISTRY AND MOLECULAR BIOLOGY, 2010, 121 (3-5): : 565 - 573
  • [33] SERUM 25-HYDROXYVITAMIN D (25-OHD) DEFICIENCY IN PREMATURE-INFANTS
    HILLMAN, LS
    HOFF, N
    MARTIN, LA
    HADDAD, JG
    PEDIATRIC RESEARCH, 1979, 13 (04) : 475 - 475
  • [34] Measurement of plasma 25-hydroxyvitamin D2, 25-hydroxyvitamin D3 and 3-epi-25-hydroxyvitamin D3 in population of patients with cardiovascular disease by UPLC-MS/MS method
    Abouzid, Mohamed
    Karazniewicz-Lada, Marta
    Pawlak, Kornel
    Burchardt, Pawel
    Kruszyna, Lukasz
    Glowka, Franciszek
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2020, 1159
  • [35] TREATMENT WITH 50 000 IU VITAMIN D2 EVERY OTHER WEEK AND EFFECT ON SERUM 25-HYDROXYVITAMIN D2, 25-HYDROXYVITAMIN D3, AND TOTAL 25-HYDROXYVITAMIN D IN A CLINICAL SETTING
    Demetriou, Emily T. W.
    Travison, Thomas G.
    Holick, Michael F.
    ENDOCRINE PRACTICE, 2012, 18 (03) : 399 - 402
  • [36] Fast Separation of 25-Hydroxyvitamin D3 from 3-Epi-25-Hydroxyvitamin D3 in Human Serum by Liquid Chromatography-Tandem Mass Spectrometry: Variable Prevalence of 3-Epi-25-Hydroxyvitamin D3 in Infants, Children, and Adults
    van den Ouweland, Johannes M. W.
    Beijers, Antonius M.
    van Daal, Henny
    CLINICAL CHEMISTRY, 2011, 57 (11) : 1618 - 1619
  • [37] Standardization of measurements of 25-Hydroxyvitamin D3 and D2
    Thienpont, Linda M.
    Stepman, Hedwig C. M.
    Vesper, Hubert W.
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2012, 72 : 41 - 49
  • [38] Design of experiments guided development of a liquid chromatography tandem mass spectrometry assay for 25-hydroxyvitamin D2 and D3
    Henderson, M. P. A.
    Zeidler, J.
    CLINICAL BIOCHEMISTRY, 2010, 43 (09) : 784 - 784
  • [39] Is high-throughput measurement of 25-hydroxyvitamin D3 without 25-hydroxyvitamin D2 appropriate for routine clinical use?
    Costelloe, S. J.
    Woolman, E.
    Rainbow, S.
    Stratiotis, L.
    O'Garro, G.
    Whiting, S.
    Thomas, M.
    ANNALS OF CLINICAL BIOCHEMISTRY, 2009, 46 : 86 - 87
  • [40] A simple automated solid-phase extraction procedure for measurement of 25-hydroxyvitamin D3 and D2 by liquid chromatography-tandem mass spectrometry
    Knox, Susan
    Harris, John
    Calton, Lisa
    Wallace, A. Michael
    ANNALS OF CLINICAL BIOCHEMISTRY, 2009, 46 : 226 - 230